Re: Readings from 436A power meter using 82357B USB-GPIB?
Hello,
Thanks for the suggestions. I tried to figure out how to do low-level communication, but don't seem to find any way.? I hope I'm just missing something trivial here. If anyone has an example of communicating with the 436 (or similar - old HP-IB) using 82357 USB-GPIB, preferably using the VISA library, that would be the best - that would give a good starting point for digging further. Regards, ? Staffan |
||||||||||
Re: A homemade diode power sensor for HP meters
开云体育-------- Weitergeleitete Nachricht --------
Mouser also shoes the wrong picture.
I use these: Lumberg Series 03 - Circular connectors with threaded joint M16 acc. to IEC 61076-2-106, IP40/IP67/IP68 But not cheap either. 73, Rainer Am 18.11.2018 um 10:04 schrieb rmc321 via Groups.Io: The 12-pin circular connector used on the power sensors is an Amphenol.? The old part number was 91-T-3638. |
||||||||||
Re: Measuring Phase with Agilent E5062A ENA
开云体育So I tried a quick test on my test cable and when
put in the smith chart mode it functions in a similar manner to
a Cartesian phase plot where the data is displayed from -180
degrees to +180 degrees, in a single continuous arc around the
outer left edge of the smith chart. Even though I had enabled
Expanded phase selected it seems that function is not applicable
when the smith chart display function is enabled. Interesting
test and TU agn for the idea. Gedas, W8BYA Gallery at Light travels faster than sound.... This is why some people appear bright until you hear them speak. On 11/18/2018 1:44 AM, bownes wrote:
|
||||||||||
Re: Readings from 436A power meter using 82357B USB-GPIB?
And if you're into Linux, there is a GPIB driver package available with bindings for python and perl. It's at
toggle quoted message
Show quoted text
and supports USB adapters as well as older boards. John ---- On 11/18/18 11:14 AM, Leo Bodnar wrote:
Staffan, |
||||||||||
Re: Readings from 436A power meter using 82357B USB-GPIB?
Staffan,
toggle quoted message
Show quoted text
Have a look at python? I am from the generation that used to use FORTRAN IV and punchcards but you have to move up with times at some point. Python has a few?idiosyncrasies that drive me mad but it's here to stay I am using python as a goto choice for all my GPIB needs now. Leo On 18 Nov 2018, at 15:55, Staffan wrote: Hello, |
||||||||||
Re: Measuring Phase with Agilent E5062A ENA
开云体育Hi Dave. The machine does actually have quite a bit of capability and I find it vastly superior to the older HP VNA's I have had in the past.? The problem is it is so poorly documented or not documented at all (at least in the operators manual) it makes using it or learning about it frustrating and very time consuming. Like in the case of Expanded Phase......turns out
that does exactly what I want it to do but you will find nothing
useful describing the function or how to use it or any of it's
limitations etc. Hard to believe that Agilent would put out such
a powerful "VNA" and not tell people how to use many of it's
common commands for making all sorts of neat measurements. I
would have assumed there was some kind of app note of other
documentation that detailed such things. Anyway, thanks for the ideas but for now at least
in this case I have found what I need. Gedas, W8BYA Gallery at Light travels faster than sound.... This is why some people appear bright until you hear them speak. On 11/18/2018 10:00 AM, Dr. David
Kirkby from Kirkby Microwave Ltd wrote:
|
||||||||||
Re: Readings from 436A power meter using 82357B USB-GPIB?
Hello,
I can control the instrument so I guess termination is correct, but reading isn't working. I'd really like to do the low level byte by byte communication, but anyone knows hos to do this with 82357? Any hints would be highly appreciated! Regards, ? Staffan |
||||||||||
Re: Measuring Phase with Agilent E5062A ENA
Dr. David Kirkby from Kirkby Microwave Ltd
On Sun, 18 Nov 2018, 01:45 Gedas <w8bya@... wrote:
You could save the data as a Touchstone file the import it into the software used for the VNWA. and display as "c. phase" which is the phase unwrapped - I assume the c is for continuous. Does your ENA support any sort of macro/programming? If so, you might be able to program something in there to unwrap phase. There's no doubt the ENAs are pretty poor compared to the PNAs. The ENAs were developed in Japan and the PNAs in the USA. Support on the older ENAs on the Keysight forums is almost nonexistent.? The latest ENAs run the same firmware as the PNAs, so things are different. I have the firmware of an ENA running on a PC (actually a virtual machine) as an emulator, but the ENA software is quite poor. Dave, G8WRB. |
||||||||||
Re: Measuring Phase with Agilent E5062A ENA
开云体育Hi Alwyn, thanks for that suggestion. Yes I am
pretty sure the 5062A has that capability. I will give it a try
and see how it correlates with the expanded phase option I think
will also do exactly what I was hoping for. Thank you for
chiming in. Gedas, W8BYA Gallery at Light travels faster than sound.... This is why some people appear bright until you hear them speak. On 11/18/2018 3:56 AM, alwyn.seeds1
wrote:
Dear Gedas, |
||||||||||
Re: Measuring Phase with Agilent E5062A ENA
开云体育I don't know your name but that's a great idea,
thank you. Gedas, W8BYA Gallery at Light travels faster than sound.... This is why some people appear bright until you hear them speak. On 11/18/2018 1:44 AM, bownes wrote:
|
||||||||||
Re: A homemade diode power sensor for HP meters
Tobias
toggle quoted message
Show quoted text
Interesting project. I am surprised that you had trouble finding 848X schematics. They are in every service manual from the 1980's? ( and we of course have them available for modest purchase price in PDF :-).. 8481A, 8481B, 8481D, 8481H, 8482A, 8483A, 8481B, 8484A, 8485A, 8487A an a few more) Any preliminary data on frequency range and how level it is over the range? What frequency are you doing the initial testing at ..50MHz calibrator? looking forward to seeing the design files ! Dave NR1DX manuals@... On 11/18/2018 9:05 AM, Tobias Pluess wrote:
Hi Guys, --
Dave Manuals@... www.ArtekManuals.com |
||||||||||
Re: A homemade diode power sensor for HP meters
Hi Guys,
sorry for my late reply. Today I was able to make some further progress. Robert was right in his post (/g/HP-Agilent-Keysight-equipment/message/92641) - the autozero problem I encountered was indeed related to the position where I connected the autozero signal. In principle, I have now the same schemativ as the 8484A diode power sensor. After moving the autozero signal, I still had the problem that the autozero did not work properly, so I further analyzed the circuit of the 8484A and realised that it has less gain than my circuit, so I changed the base resistor of Q1 (https://hb9fsx.ch/wordpress/wp-content/uploads/2018/11/8481schem.png) to 1k Ohms and the autozero problem was fixed - pressing the autozero button perfectly allows the meter to zero the sensor. However, with this modification, my sensor had a different gain, and therefore the readings on the meter had an offset. I was able to re-adjust the gain with the trim pot and when I put a 10dB attenuator at the sensor input, I was able to have the same readings on the meter as on the signal generator I used to test the sensor. However, as I said, it is not terribly accurate, at very low power levels, it was around -1dB off, and at high power levels, it seems like the diode transitions from square law region to linear region and therefore the readings were also off a bit. But this is no big deal; one can simply add attenuators in front of the sensor such that it is ensured the diode is in its square law region. Thanks at Orin for the 8484A sensor schematic. I have never found it, even though I was of course looking for it. Joarez (/g/HP-Agilent-Keysight-equipment/message/92646) says that the capacitors need to be there to prevent the circuit from generating switching abberations. This is probably true because the 8484A manual tells that the capacitors were factory selected individually for each sensor, so they are probably related somehow to the JFET switches. In my sensor, I currently don't have these capacitors, and this is probably another reason why the readings are not 100% accurate. It could also be related to the JFETs I use - I chose the first ones I could get, but somewhere in my junk boxes I should have a pile of 2N4856, which are probably better. However, I would prefer SMD devices since I will make a PCB in SMD design. I got a bunch of MMBFJ310, but I have not enough of them to make a lot of experiments :-/ Thanks also to rmc321 and Rainer for the link to the connectors. I was once told that HP used their own connectors, so I didn't even try too hard to find the connectors, but very good you found them! I could have saved my valuable power sensor cable if I had known it earlier :-) I think I am soon going to make another dead bug prototype and in parallel design a PCB and probably my own housing. I will share the design files as soon as I am ready. Tobias HB9FSX |
||||||||||
Re: Readings from 436A power meter using 82357B USB-GPIB?
Many older instruments don't use the EOI line, but also "Must" have the correct command terminator byte (if one is specified) no more, no less.. It'll all be documented in the individual instruments manual.? The fun bit these days is getting "modern" bus controllers to play nice with older non IEE488.2 devices. Don't rely on the more modern high level function calls, start with the low level byte by byte stuff, and run a bus traffic capture too, guessing Agilent has an equivalent to NI's 'I/O Trace' (was NI Spy) program. Regards Dave B (G0WBX) |
||||||||||
Re: A homemade diode power sensor for HP meters
开云体育Mouser also shoes the wrong picture.
I use these: Lumberg Series 03 - Circular connectors with threaded joint M16 acc. to IEC 61076-2-106, IP40/IP67/IP68 But not cheap either. 73, Rainer Am 18.11.2018 um 10:04 schrieb rmc321 via Groups.Io: The 12-pin circular connector used on the power sensors is an Amphenol.? The old part number was 91-T-3638. |
||||||||||
Re: Readings from 436A power meter using 82357B USB-GPIB?
开云体育Hi re? HPIB ? I am not a computer programmer but I either use? HP basic 4.1 with a HP computer that has a HPIB port ?or ?national instruments lab view ?on windows xp ??and the hardware is either a NI card for the PC or a NI ???USB converter Paul B ??south cost UK ? From:
[email protected]
[mailto:[email protected]] On Behalf Of Staffan ? Hello, ? Thanks for the suggestion. I have tried this, but it doesn't seem?to work. Any recommendations to how I should configure Interactive IO? I will use Visual Basic in Excel for later programming - hopefully using Agilent VISA libraries... ? In general, anyone with experience of communicating over USB-GPIB 82357 with _very_ old instruments (single character commands etc...)? Is there a fundamental limitation? ? Is there any way of performing low level communication/control of the 82357, i.e., controlling the separate digital lines? ? Regards, ? Staffan? ? On Sun, Nov 18, 2018 at 2:45 AM Daun Yeagley <daun@...> wrote:
No virus found in this message. |
||||||||||
Re: Readings from 436A power meter using 82357B USB-GPIB?
Hello, Thanks for the suggestion. I have tried this, but it doesn't seem?to work. Any recommendations to how I should configure Interactive IO? I will use Visual Basic in Excel for later programming - hopefully using Agilent VISA libraries... ? In general, anyone with experience of communicating over USB-GPIB 82357 with _very_ old instruments (single character commands etc...)? Is there a fundamental limitation? Is there any way of performing low level communication/control of the 82357, i.e., controlling the separate digital lines? Regards, ? Staffan? On Sun, Nov 18, 2018 at 2:45 AM Daun Yeagley <daun@...> wrote:
|
||||||||||
Re: A homemade diode power sensor for HP meters
In my connector search above, I didn't pay any attention to how the connector mounts onto the device.? For a new homebrew design, it probably doesn't matter.? But, if you are specifically trying to replace the connector in an existing HP power sensor unit, you might have to check if you needed a front-threaded or a rear-threaded panel mount connector.?
|
||||||||||
Re: A homemade diode power sensor for HP meters
The 12-pin circular connector used on the power sensors is an Amphenol.? The old part number was 91-T-3638.
Now, it's part of their "C 091 A/B/D" series circular connectors.? Several variations are listed in the Amphenol catalog: https://www.amphenol-sine.com/pdf/catalog/C091Amphenol.pdf Mouser has several slightly expensive options (a few of which appear in this link -- note that widening the Mouser search filter will supply more results): It appears that an eBay seller is selling them at a lower price.? His text says they are the 12-pin versions, although his photo shows the 14-pin version.? It's hard to tell which of the two connectors he is selling.? However, it appears the two additional pins shouldn't matter since it's the female part of the connector -- you would leave those two extra pins unconnected. |
||||||||||
Re: Measuring Phase with Agilent E5062A ENA
开云体育Dear Gedas,There is a way round this you might find useful- reference line extension function; I used this on an 8753B to take out the delay in a long optical delay line. Essentially, you sweep over a frequency range that the DUT is phase linear over.? Then adjust reference line extension until you get flat phase.? Then read off the extension in ns.? Then calculate the phase shift from frequency and delay. I have n’t checked whether the same function is available for E series analysers; it would be a pity, if not. Regards, Alwyn ?? _____________________________________________________ Alwyn Seeds, Director SynOptika Ltd., 114 Beaufort Street, London, SW3 6BU, England. Tel.: +44 (0) 20 7376 4110 SynOptika Ltd., Registered in England and Wales: No. 04606737 Registered Office: 114 Beaufort Street, London, SW3 6BU, United Kingdom. _____________________________________________________
|
||||||||||
Re: Measuring Phase with Agilent E5062A ENA
开云体育Have you tried displaying your results on a Smith chart? Check out Dr Joel Dunsmore’s book, “Microwave Component Measurements” for more than you ever wanted to know on the subject.? On Nov 18, 2018, at 01:00, Gedas <w8bya@...> wrote:
|