I am not a computer programmer but I either
use? HP basic 4.1 with a HP computer that has a HPIB port
?or ?national instruments lab view
?on windows xp ??and the hardware is either a NI card for the PC
or a NI ???USB converter
Paul B ??south cost UK
?
From:[email protected]
[mailto:[email protected]] On Behalf Of Staffan Sent: 18 November 2018 10:16 To:[email protected] Subject: Re:
[HP-Agilent-Keysight-equipment] Readings
from 436A power meter using 82357B USB-GPIB?
?
Hello,
?
Thanks for the suggestion. I have tried this, but it doesn't
seem?to work. Any recommendations to how I should configure Interactive
IO? I will use Visual Basic in Excel for later programming - hopefully using
Agilent VISA libraries...
?
In general, anyone with experience of communicating over USB-GPIB 82357
with _very_ old instruments (single character commands etc...)? Is there a
fundamental limitation?
?
Is there any way of performing low level communication/control of the
82357, i.e., controlling the separate digital lines?
On Sun, Nov 18, 2018 at 2:45 AM Daun Yeagley <daun@...> wrote:
Hi Staffan
Yes, you need to meter to be in addressable mode (or as you say, "normal").
If you use the interactive IO, you would use the "Send and
Read"..? Type the command to fetch the reading in the text box, and
use the "Send and Read" button, which will set the computer as the
talker, send the command, and then it sets it up so the meter is the
talker.? You should see the results show up in the history below.? It
puts the actual results there, with no additional formatting so you can see
exactly what it sends back.? Of course you can also do it individually
using the separate "Send Command" and "Read Response" as
well. The "Send and Read" just combines the two actions.
What language do you plan to use for your program?
Daun
On 11/17/2018 4:52 PM, Staffan wrote:
Hello,
Just got hold of an old HP436A power meter and would like to do some automated
measurements. Controlling the instrument is simple using the 82357B USB-GPIB
module (I can set range etc), but reading data? As a first step I'd like to
test using Agilent Interactive IO from Connection Expert. If that works, I'm
fairly sure I can get it working in a program also.?
The 436 is set to NORMAL,
i.e., not TALK ONLY. It does have the HP-IB option 022 installed. My guess is
that I need to address the instrument to Talk in some other way than the
"Read Response" used in Interactive IO.
Anyone with experience of this?
Regards,
? Staffan
?
--