开云体育

ctrl + shift + ? for shortcuts
© 2025 开云体育

Re: Measuring Phase with Agilent E5062A ENA


 

开云体育

Hi Alwyn, thanks for that suggestion. Yes I am pretty sure the 5062A has that capability. I will give it a try and see how it correlates with the expanded phase option I think will also do exactly what I was hoping for. Thank you for chiming in.

Gedas, W8BYA

Gallery at 
Light travels faster than sound....
This is why some people appear bright until you hear them speak.
On 11/18/2018 3:56 AM, alwyn.seeds1 wrote:

Dear Gedas,

There is a way round this you might find useful- reference line extension function; I used this on an 8753B to take out the delay in a long optical delay line.

Essentially, you sweep over a frequency range that the DUT is phase linear over.?

Then adjust reference line extension until you get flat phase.?

Then read off the extension in ns.?

Then calculate the phase shift from frequency and delay.

I have n’t checked whether the same function is available for E series analysers; it would be a pity, if not.

Regards,

Alwyn
??
_____________________________________________________

Alwyn Seeds, Director
SynOptika Ltd.,
114 Beaufort Street,
London,
SW3 6BU,
England.

Tel.: +44 (0) 20 7376 4110


SynOptika Ltd., Registered in England and Wales: No. 04606737
Registered Office: 114 Beaufort Street, London, SW3 6BU, United Kingdom.
_____________________________________________________

Join [email protected] to automatically receive all group messages.