开云体育

ctrl + shift + ? for shortcuts
© 2025 Groups.io
Date

Re: 564 retrace blanking, CRT fault? Self-healing?

Rob
 

In TV work.... way back when.... I saw a couple of shorts clear via
movement....never a reconnect.

However I have seen a light bulb filament fortunately fall together on a
shake. So I think it is not outside the realm of possible...

Unfortunately the bulb was extremely short lived after this.

If the CRT did 'selfheal' I would suggest you treat it very carefully
*smile *. You think a dirty connection perhaps or are you sure it was
internal? I know your skill set so it is indeed a rhetorical question....

Anyway I would be interested in its life span from here. You may have a
Guinness record in the making....

Obviously being a bit silly but congratz none the less. Whatever the case.
It is good when the Oscope gods smile....
Rob

-----Original Message-----
From: TekScopes@... [mailto:TekScopes@...] On Behalf
Of Albert
Sent: Friday, July 13, 2012 3:41 PM
To: TekScopes@...
Subject: [TekScopes] Re: 564 retrace blanking, CRT fault? Self-healing?

I swapped the CRT of this 564 with the CRT of a known to be good 564. Lo and
behold, both 'scopes worked fine now! Then I moved the CRTs back to their
original positions and again both 'scopes worked fine.
Spontaneous self-healing of a loose contact inside the CRT?

Albert

--- In TekScopes@..., "Steve" <ditter2@...> wrote:


--- In TekScopes@..., "Albert" <aodiversen@> wrote:

Thanks Bob,

I still don't understand it. Today I measured DC developed between the
blanking plates (across the 570k resistor) with no plugins installed. There
is DC, rising to about 0.4 V after some warming up. So there seems to be at
least "some" connection to pins 5 and 7 internally. But there was no
response in intensity whatsoever when I supplied 20 V DC between those
plates, using a floating DC supply.
In a good 564 there was also some 0.3-0.4 V DC between the plates, and
20 V was sufficient to make the dot disappear.

Albert
I would be careful about making assumptions on potential difference
between the plates, generated by a floating supply. While it is true that
the field difference between the plates bends the beam, there is a component
that needs to be referenced to the relative charge of the electrons at the
point in time they pass the plates - in other words a DC bias on both plates
relative to ground.
What I don't understand is why the 56x series uses BOTH deflection
unblanking AND a DC coupled grid control. I have seen deflection blanking
used when the grid is AC coupled to the intensity blanking / control, and DC
biased directly from the cathode. In these cases, deflection blanking is
needed for very long blanking intervals, as the time constant in an
capacitor coupled system can only remain off for the time constant. But the
56x uses the more complicated DC system, which uses dedicated HV windings
for both the cathode and grid bias, eliminating any time constants and the
need for another blanking system.
So why did Tek design deflection based blanking into the 56x series, and
why do the time bases use it, with the DC grid control only used for
intensify in dual time bases?
- Steve



------------------------------------

Yahoo! Groups Links


Re: 7603 Help for jumping readouts

 

Jumping readouts resolved.

After tracking signals from readout board to V and H amp boards everything
looked OK, however when looking for XY signal on vert interface it was missing.

Turns out interconnect cable between H amp and main interface board had came
loose from main interface connection. After reinstalling everything is fine with
readouts.

Thanks to those who responded.

Larry

--- In TekScopes@..., "photo692002" <photo692002@...> wrote:

My R7603 has developed a problem where the readouts jump around a lot based on the input signal frequency and amplitude, signal trace is fine. With no input, readouts are steady. This readout problem occurs with several different vertical and horizontal plugins, problem appears to be in the 7603 frame.

Power supplies measure well and look good on scope, dont think they are involved. I have been unable to determine if the readout board or the vertical or horizontal amp board is involved as I do not have spares.

Has anyone seen this problem and resolved it?

Larry


Re: Polish out scratches in blue filter

John Griessen
 

On 07/13/2012 05:11 PM, KenS wrote:
a sure-fire way to polish out scratches from these blue filters?
A product called gel gloss is worth trying.


Re: Polish out scratches in blue filter

Tom Jobe
 

开云体育

Hi,
I don't know if the 2467 uses the same size CRT filter as the 2465 and the similar scopes, but if it does, I think Victor Silva has those blue filters.
Maybe Victor will see your filter request. If not, you could email him at daejon1@...
Victor is?very knowledgeable?resource for 24xx scopes, and he posts on Tekscopes when he can.
tom jobe...
PS I have a small?assortment of 24x5 scopes, but not a 2467, or I could answer my own question about the filter size.
?
?
?

----- Original Message -----
From: KenS
Sent: Friday, July 13, 2012 3:11 PM
Subject: [TekScopes] Polish out scratches in blue filter

?

Hi,

I have a 2467 with some minor scratches in the blue filter. I found out the hard way with another Tek scope that you can easily make problems worse if you try to remove scratches with a plastic polish, so I am cautious. There appear no sources of NOS blue filters out there. Anybody know of any that wouldn't come up in Google? Or, a sure-fire way to polish out scratches from these blue filters?

Thanks!

- Ken


Re: 7704A repair attempt, progress

 

Albert <aodiversen@...> ha scritto:


The time base sends the unblanking pulse to the main frame *only* during the sweep. Since you have no sweep the dot is blanked and you shouldn't see any effect of the Intensity control.
I suppose the time base doesn't run in AUTO mode either?
Faulty Q2577 and Q2579 can produce a Sweep Lockout condition at contact B8 of the main interface connectors and block the time base.
Thank you Albert, more details:
I'm testing with the TB in A Hor., B hor. is empty,
I have the trigger and the gate pulse out on the back.
The TB is in auto.
A sweep lockout (J3-B8) is low enough, I think (about 0V),
B sweep lockout (J4-B8) is about 3.8V.

Tried also swapping U2530 with U2540, so I can select B Hor.
With TB in B slot still no sweep, in this configuration I have no
working trigger.

I extracted the main interface board, I'll check the diodes and
transistors, and clean it a bit.

Fabio.

----------------------------------------------------------------
This message was sent using IMP, the Internet Messaging Program.


Polish out scratches in blue filter

 

Hi,

I have a 2467 with some minor scratches in the blue filter. I found out the hard way with another Tek scope that you can easily make problems worse if you try to remove scratches with a plastic polish, so I am cautious. There appear no sources of NOS blue filters out there. Anybody know of any that wouldn't come up in Google? Or, a sure-fire way to polish out scratches from these blue filters?

Thanks!

- Ken


Re: For Sale AS IS

 

Hi, There is a chance to get the crt nearby, and that means that the shipping costs are cheaper.
So I cannot make a decission yet.
Rens


From: jerry massengale To: TekScopes@...
Sent: Friday, July 13, 2012 3:20 PM
Subject: Re: [TekScopes] For Sale AS IS

?
Hi,

the shields means the blue filter and clear implosion shield. the interface includes
1. main 670-1623-01
2. vert intf? 670-1625-06
3. horz intf 670-1626-07
4. trig intf 670-1627-??
5. logic 670-1624-01
6. cables

crt @ 5#14 boxed is $50 plus $50=$100 total
add intf 7#8 boxed is $50 plus $40 plus $57 = $147

Jerry Massengale



-----Original Message-----
From: Rens Tel
To: TekScopes
Sent: Thu, Jul 12, 2012 5:43 pm
Subject: Re: [TekScopes] For Sale AS IS

?
Hi,any idea what the shipping costs would be of the?CRT 154-0644-05 with bezel to the Netherlands?
what do you mean by the shields?
are the interface cards of the 7903 the vertical-horizontal- and trigger interfaces?
kind regards,
Rens Tel



From: jerry massengale <j_massengale@...>
To: tekscopes@...; TekScopes2@...
Sent: Thursday, July 12, 2012 7:27 PM
Subject: [TekScopes] For Sale AS IS

?
Hi,

As parts of moving I have found more Items I no longer need. They no known problems but I no longer have ways to test them. They are offered as is.

1. 7903 Main interface with logic and interface cards $40 plus shipping.
2. 7704A Main Interface with logic and interface cards $40 plus shipping.
3. 7704A/7904 crt 154-0644-05 with bezel,shields, and metal shield $50 plus shipping.
4. 670-1853-01 Horizontal Amp $15 plus shipping.
5. 670-1854-00 Z-axis card $15 plus shipping.
6. 670-1854-00 Z-Axis card $15 plus shipping.
7. 670-1636-03 Z-axis card $20 plus shipping.
8. R7603 with $80 plus shipping. will negoiate plugins.
8a. 7B70
8b. 7A12 Dual tracer amplifier, push button
8c. 7A15A Single trace amplifier
8d. 367-0138-00 handles or 367-0022-00 handles

don't know with handle is which is which, I have 2 of each. One folds and the other is fixed. The R7603 buyer gets choice. spare handles go for $5 per pair.

Jerry Massengale






Re: 564 retrace blanking, CRT fault? Self-healing?

Albert
 

The version saying "... than can be observed through a magnifier glass" seems to fit perfectly here.

Albert

--- In TekScopes@..., "raymonddompfrank" <r.domp.frank@...> wrote:

"There are more things in heaven and earth, Horatio,
Than are dreamt of in your philosophy"

Hamlet

--- In TekScopes@..., "Albert" <aodiversen@> wrote:

I swapped the CRT of this 564 with the CRT of a known to be good 564. Lo and behold, both 'scopes worked fine now! Then I moved the CRTs back to their original positions and again both 'scopes worked fine.
Spontaneous self-healing of a loose contact inside the CRT?

Albert


Re: Tube(s) needed

 

Pulled from old tek?

I search 8416.

Regards
Matt

--- In TekScopes@..., "Ron" <mercedesmann@...> wrote:


I have one. 20 dollars + shipping? = 25 total?


--- In TekScopes@..., Michael Dunn <mdunn@> wrote:

Hi. Does anyone have any 8608 tubes? (med power pentodes with
plate cap, used for vertical deflection on some Tek scopes (eg, 556,
549))

I'm willing to pay a reasonable amount for 1 or several, but not
an exorbitant price.

-----------------------------------------------------------------
Michael Dunn | Surround Sound Decoder & Stereo Enhancement
Cantares | Self-Amplified Speakers, Test Equipment
74 George St. | Ambisonic Surround Sound CDs and Recording
Waterloo, Ont. | (519) 744-9395 (fax: 744-7129)
N2J 1K7 | mdunn@
Canada |
-----------------------------------------------------------------


Re: 564 retrace blanking, CRT fault? Self-healing?

raymonddompfrank
 

"There are more things in heaven and earth, Horatio,
Than are dreamt of in your philosophy"

Hamlet

--- In TekScopes@..., "Albert" <aodiversen@...> wrote:

I swapped the CRT of this 564 with the CRT of a known to be good 564. Lo and behold, both 'scopes worked fine now! Then I moved the CRTs back to their original positions and again both 'scopes worked fine.
Spontaneous self-healing of a loose contact inside the CRT?

Albert

--- In TekScopes@..., "Steve" <ditter2@> wrote:


--- In TekScopes@..., "Albert" <aodiversen@> wrote:

Thanks Bob,

I still don't understand it. Today I measured DC developed between the blanking plates (across the 570k resistor) with no plugins installed. There is DC, rising to about 0.4 V after some warming up. So there seems to be at least "some" connection to pins 5 and 7 internally. But there was no response in intensity whatsoever when I supplied 20 V DC between those plates, using a floating DC supply.
In a good 564 there was also some 0.3-0.4 V DC between the plates, and 20 V was sufficient to make the dot disappear.

Albert
I would be careful about making assumptions on potential difference between the plates, generated by a floating supply. While it is true that the field difference between the plates bends the beam, there is a component that needs to be referenced to the relative charge of the electrons at the point in time they pass the plates – in other words a DC bias on both plates relative to ground.
What I don't understand is why the 56x series uses BOTH deflection unblanking AND a DC coupled grid control. I have seen deflection blanking used when the grid is AC coupled to the intensity blanking / control, and DC biased directly from the cathode. In these cases, deflection blanking is needed for very long blanking intervals, as the time constant in an capacitor coupled system can only remain off for the time constant. But the 56x uses the more complicated DC system, which uses dedicated HV windings for both the cathode and grid bias, eliminating any time constants and the need for another blanking system.
So why did Tek design deflection based blanking into the 56x series, and why do the time bases use it, with the DC grid control only used for intensify in dual time bases?
- Steve


Re: 564 retrace blanking, CRT fault? Self-healing?

Albert
 

I swapped the CRT of this 564 with the CRT of a known to be good 564. Lo and behold, both 'scopes worked fine now! Then I moved the CRTs back to their original positions and again both 'scopes worked fine.
Spontaneous self-healing of a loose contact inside the CRT?

Albert

--- In TekScopes@..., "Steve" <ditter2@...> wrote:


--- In TekScopes@..., "Albert" <aodiversen@> wrote:

Thanks Bob,

I still don't understand it. Today I measured DC developed between the blanking plates (across the 570k resistor) with no plugins installed. There is DC, rising to about 0.4 V after some warming up. So there seems to be at least "some" connection to pins 5 and 7 internally. But there was no response in intensity whatsoever when I supplied 20 V DC between those plates, using a floating DC supply.
In a good 564 there was also some 0.3-0.4 V DC between the plates, and 20 V was sufficient to make the dot disappear.

Albert
I would be careful about making assumptions on potential difference between the plates, generated by a floating supply. While it is true that the field difference between the plates bends the beam, there is a component that needs to be referenced to the relative charge of the electrons at the point in time they pass the plates – in other words a DC bias on both plates relative to ground.
What I don't understand is why the 56x series uses BOTH deflection unblanking AND a DC coupled grid control. I have seen deflection blanking used when the grid is AC coupled to the intensity blanking / control, and DC biased directly from the cathode. In these cases, deflection blanking is needed for very long blanking intervals, as the time constant in an capacitor coupled system can only remain off for the time constant. But the 56x uses the more complicated DC system, which uses dedicated HV windings for both the cathode and grid bias, eliminating any time constants and the need for another blanking system.
So why did Tek design deflection based blanking into the 56x series, and why do the time bases use it, with the DC grid control only used for intensify in dual time bases?
- Steve


Re: 7704A repair attempt, progress

Albert
 

I'm still messing around with the 7704A.

Faulty parts:
Till now I found 3 shorted transistors in the
Z-axis board, for now I swapped them with normal
transistors, I will search better replacement
in future, the parts are Q4159 Q4183 Q4185.
I changed also Q4163 because it had a very low beta.
Also U2530 isnt working ok, it doesnt select B horizontal
timebase with the correct input combination.
In the diagram and measuring it, seem
that the output is high when pin 6 is at +1V,
and low when is slightly negative,am I correct?
I think so. The 7704 manual shows the ICs while pin 6 is either +1.0 V or -0.7 V. But a saturated collector voltage (a few tenths of a volt above ground) must be LO as well since some pins 8 (and 1) are driven that way.

The other faulty component is Q2577 with a beta of zero.
Q2577 and 2579 are selected parts, anyone knows what
are the selection criteria? Can I try to put in
normal npn transistors for now?
As long as you don't use CHOP or ALT any high beta NPN could do the job, I think.


There is a little progress:
After replacing Q4159 Q4183 Q4185
I have the readout working on the screen,
also readout intensity regulation is ok.
The trace intensity regulation is still non
working. It remains very low, almost like
if the z-axis is blanking the screen.
Also I still dont have the sweep from the timebase.
When I use the TB in vertical bays it works, when I insert
into one of the 2 horizontal bays it doesnt generate the
sweep, but the horizontal position pot on the TB moves
the dot on the screen, so it's like the TB sweep is
disabled.
The time base sends the unblanking pulse to the main frame *only* during the sweep. Since you have no sweep the dot is blanked and you shouldn't see any effect of the Intensity control.
I suppose the time base doesn't run in AUTO mode either?
Faulty Q2577 and Q2579 can produce a Sweep Lockout condition at contact B8 of the main interface connectors and block the time base.

Albert



Fabio.


Re: 7000 series on screen graphics question 7704A versus 7633

 

Astigmatism adjustments are on the Z-Axis board (inside, on top of the HT box). You see four pots in a square together. All locations based on looking at the PCB from the right hand side of the scope. The right hand two as you face the board from the side of the 'scope are the astigmatism adjustments - the rear one for full scan, the front one for reduced scan.


Regards,
David Partridge

-----Original Message-----
From: TekScopes@... [mailto:TekScopes@...] On Behalf Of Chris Wilson
Sent: 13 July 2012 11:31
To: TekScopes@...
Subject: RE: [TekScopes] Re: 7000 series on screen graphics question 7704A versus 7633



13/07/2012 11:27

Is the astigmatism adjustment internal on these David? If I look closely it almost appears to be trying to show two numerals / letters at the same time, possibly the figure 1 (one)? As it's otherwise 100% and new old stock I'd quite like to get this facet of its operation fixed. Thanks.

--
Best Regards,
Chris Wilson.
mailto: chris@...


Re: I have a type 585 oscope how much is it worth

Rob
 

Beautifully said and indeed what lead to my diving head long into the 7000
series I have chosen for now as my particular expression of the thoughts
below.

Tis why I enjoy this forum so much. Regardless of the specific cup of tea
people choose to quench the thirst (ok corny metaphor but what the heck I am
an Chem E after all not a poet).

Bravo Kurt.

-----Original Message-----
From: TekScopes@... [mailto:TekScopes@...] On Behalf
Of sipespresso
Sent: Friday, July 13, 2012 3:23 AM
To: TekScopes@...
Subject: [TekScopes] Re: I have a type 585 oscope how much is it worth

There are many reasons a person might value a scope. When a person cares
about the beauty of an instrument and what it represents historically, its
utility as an oscilloscope for everyday use becomes a secondary factor, if a
factor at all.

Also, a lot of interesting concepts are embodied in old scopes. For
example, distributed amplifiers, equivalent-time analog sampling, tunnel
diode triggering, and storage CRTs. These things tickle my mind. I own
some scopes just so I can marvel at them. Basically, they are Faberge eggs
for an electronics enthusiast.

That said, I do like to bring my old scopes into proper calibration.
Sometimes a decision has to be made, whether to functionally restore or just
clean up a scope. Historic integrity is likely to be lost in a functional
restoration of an early Tek scope. It is a tough decision. I have no rigid
position on the issue.

-Kurt


For example, I recently got a 514. I'm pretty sure that I'd need to do
extensive replacement of internal parts to get it working properly. It is
never clear what's the right thing to do in this situation and every person
has to make their own decision. For the 514, I'm going to leave it alone.
Other times, I couldn't resist the urge to do functional restoration. My
567 was just too interesting not to repair. So I replaced the electrolytics
and got it fully operational. I wouldn't blame somebody if they chose, in
the same situation, to leave the scope unmolested.




--- In TekScopes@..., "Steve" <ditter2@...> wrote:


--- In TekScopes@..., "mattko87" <matt8@> wrote:

That ist right.
585, 581 are a epic technologiy.
high bandwith, distributed deflection and tunneldiode -base triggering
(only with A-version)
IMO, Tunnel diode trigging in the A models is a VERY IMPORTANT difference.
The non-A's only triggered at a tiny fraction of their BW rating.

Not much of a fan of 4 division vertical as well. The 545 and 545A shared
this same trait. Lots of BW for the time, but at an expense.

The design is interesting for its day, but I don't find them too useful as
working scopes.

- Steve



------------------------------------

Yahoo! Groups Links


Re: TDS 540 scope needs help

 

--- In TekScopes@..., "clboy168" <clboy168@...> wrote:

Sir:
Thank you very much. How kind of you!
Yes,i made some mistakes.You are perfectly right.
Now,i have some new questions,
1: the ic(u502) got 250M(ACQCLK) input and DLYCLK input,but the output was not a 250M sign,just was a 10khz disturbing sign.Could
i say it's dead? u501 was the same.

2: how can i get the scope of the 250M(acqclk)?
i got a tek 520a to catch the sign,but failed.
So i used a frequency counter to catch the sign finally.

Regards
------------
1: There are two more inputs that are common to both chips:
pin 9 (CLKMODE2) which should be TTL High
pin 24 (VSUB) which should be -10 VDC
2: With a TDS520A and a suitable probe you should be able to see
this signal.
/H?kan


Re: I need CRT for 305DMM

Bob Koller
 

Martin,

I looked today, and I do not have a CRT for your 305. The tube I had in mind is for the 221.

Bob


From: Martin Mehlhose
To: tekscopes@...
Sent: Thursday, July 12, 2012 3:13 PM
Subject: [TekScopes] I need CRT for 305DMM

?
I need a CRT for my Tek 305DMM

greetings from germany
Martin



Re: 7000 series on screen graphics question 7704A versus 7633

Jim Popwell Jr
 

HI CHRIS,
I would do the astig and refocus….astig…and refocus for sharpest beam definition. use the calibrator and adjust for best looking sq wave…then kill the vert deflection and the horz deflection…turn the intensity down and adjust both again for the sharpest spot….increase the intensity and readjust both again…then go back to normal sweep…….

NOW DO THE CAL procedure on the display bd.

there are a couple of adjustments on the bd that will straighten out the "jumpy double" characters.
won't take much time but will make a big difference.
if i remember right these characters were created by different currents at different times… so a little too much or too little current misshaped the characters.
i often find the right controls, then adjust them by eye for best display, by finding the sweet spot of the adjustment. i.e. find the ends of the range where the adjustment makes a total mess with the adjustment one way then going back thru the best look…until you get a mess again……...then splitting the difference.
jim

On Jul 13, 2012, at 3:30 AM, Chris Wilson wrote:



13/07/2012 11:27

Is the astigmatism adjustment internal on these David? If I look
closely it almost appears to be trying to show two numerals / letters
at the same time, possibly the figure 1 (one)? As it's otherwise 100%
and new old stock I'd quite like to get this facet of its operation
fixed. Thanks.

--
Best Regards,
Chris Wilson.
mailto: chris@...



------------------------------------

Yahoo! Groups Links



Re: 7000 series on screen graphics question 7704A versus 7633

 

The astigmatism adjustment is internal on the Z-Axis board (pages 5-35
and 5-36 in the service manual) and there are two adjustments: One is
for full scan operation and the other is for reduced scan operation.

There is also a character scan adjustment (page 5-53 in the service
manual) on the readout board which control character overscan. I
would fiddle with that first. The "m" and any character with a closed
loop make good calibration targets.

If there are issues with the characters wiggling as the trace position
changes, then vertical thermal balance and horizontal transient
response calibration may help.

The diagrams showing the locations of the internal adjustments are at
the end of section 8 in the service manual.

After calibration, my 7834 readout was sharper than my 7603 with the
worn CRT and not significantly worse than my 7905.

On Fri, 13 Jul 2012 11:30:34 +0100, Chris Wilson
<chris@...> wrote:

13/07/2012 11:27

Is the astigmatism adjustment internal on these David? If I look
closely it almost appears to be trying to show two numerals / letters
at the same time, possibly the figure 1 (one)? As it's otherwise 100%
and new old stock I'd quite like to get this facet of its operation
fixed. Thanks.


Re: TDS 684a keeps turning ON and OFF

 

开云体育

You may have shorted power diodes, shorted power transistors and shorted capacitors in the power supply. As a result the supply keeps trying to start-up but senses a short and shuts down and repeats continuously, hence the voltage on the bulk capacitor jumps up and down.

On 2012-07-13 1:02 AM, mickeyaslam wrote:

Hi Jay,
   Thanks for your suggestion but it didn't work. I am pretty sure power supply is the problem. Maybe something has to do with start up circuit, since power supply tries to start and then shuts off. One thing I noticed is that the voltage on the input bulk cap (470uF, 450V) is not stable and jumps ups and down(should have been stable DC). Maybe something has to do with power factor correction circuit. not sure need to do further investigation. Maybe something not working in the OV and OC protection circuit.


mickey




Re: Help with R7903 needed, some success...

 

I did not bother with a photo of the heater current since it matched
yours.

I will keep watching the discussion to see where I can help but let me
know if you want any specific measurements.

On Fri, 13 Jul 2012 12:51:45 -0000, "keithostertag"
<keitho@...> wrote:

Thanks David for taking the time to check. As Magnus says, your confirmation allows me to tick this off my check list and move forward. I'm relieved too, because needing to replace an expensive and hard-to-get part like the CRT or a transformer would put a serious crimp in my day...

Thanks,
Keith Ostertag

--- In TekScopes@..., David <davidwhess@...> wrote:

This is almost exactly what I saw when I measured the heater current
on my 7904 with a P6021 set at 10mA/mV and 5mV/div. Peak to peak was
23mV on mine so 230mA peak to peak.

On Thu, 12 Jul 2012 18:02:12 -0000, "keithostertag"
<keitho@...> wrote:

Still trying to confirm whether the CRT is good or not, or why there's no activity from it.

Here is a photo of the current through the conductor to pin 1 to the CRT, which is one side of the (6.3V) heater:



Does this seem reasonable? Running about 22KHz, but I didn't realize it would be a square wave... with the current probe set to 10mA/mV, this calculates to about 215mA AC p-p, right? Is this normal/reasonable for a CRT heater? I don't see anywhere in the manual that indicates what it should be. One of you guys will have the experience to know if it is within the ball-park of normal operations.

I also looked at pins 2,3, and 4 of the CRT. None of those show any current using my current probe (with no scope input). Shouldn't pin 2 (the 3KV cathode) show some current? (measure -2950V with my Fluke and HV probe).

Basically at this point I am struggling to find out why there is no CRT activity (no BF, no readout, no trace, no dot). So far, the tests points I've looked at on the z-axis board seem within tolerances, though I can't go very far into the calibration check list without a trace. Same with the vertical and horizontal deflection inputs.

Any suggestions as to how to troubleshoot this would be greatly appreciated.

Thanks,
Keith Ostertag