Re: TDS520 in Brittany: FAIL++Acquisition FAIL++ Attn/Acq interface
Hi,
Sorry for this old story but there's some news about my TDS520. Due to real life it waited since december 2023 for my attention. Last week only took it from its shelf, opened it and I removed the nearly 50 chemical caps on the acq board (A10). Then A10 went in a friend's dishwasher. I then recapped with Nichicon and put it back. Remember it booted in 2023 and only failed on Acquisition & Attn/Acq interface. Now no boot. I measured on the new caps and found one supposed +5V to be -0.6V. To make it short there's a through hole just near C1584 and of course it juiced there and now no more continuity. I soldered a small piece of wire and got back that 5V. But still no boot and no 1kHz on the probe comp signal. I followed the troubleshooting procedure which told me to "Change A12". Well, thanks but no. That procedure had me disconnect some connectors though and once reconnected the scope booted. -Fail processor (pass in 2023) -Pass acquisition (fail in 2023) -Fail Attn/acq (as in 2023) There I got myself a little snack away from the darn thing. Came back two hours later and rebooted Now Fail processor and all the rest pass. And : "Consult service manual for fault isolation. Press CLEAR MENU to continue" which I did but not with much hope. I was wrong. I put a probe on CH1 and got the probe comp signal on screen. I did not proceed recapping A11 (processor) because it booted and I wanted to test A10 first. Now I will. Also I should "Consult service manual for fault isolation" ! There's still that Fail processor. Here I am and, as usual, all criticism, comments or advices welcome. Best Renaud |
Re: Anyone selling a PG506 or 506A?
SG 503, often needs main frequency range C'AM switches fixed.
Requires special Tek cable for cal into 50 ohms SG504 needs à leveling head PG 506 rare and costly, see epay. Often needs repair or cal PG506A can be $700..2000 see epay The TM 5000 mainframe takes a 3-4 wide CG5110 that combines the functions of all the TM 500 plugins for cal Bon chance J |
Re: Tektronix 2247A problem. Keep or return?
I did some more measurements and there is indeed some ripple on the LV PSU. The worst of it is on the 130V and the 15V rails. The 5V rail which powers the area of the circuit that 'appears' to be problematic is marginal. I'm contemplating replacing all of the caps in the LV while I have it apart.
|
Re: 2230 A4 timing board weird connection
I am still waiting for some advice.
In the mean time I checked for poor connection solder joints. Clean the A4 timing board. On Sun, Mar 30, 2025, 8:42 AM Saroj Pradhan via groups.io <sarojman.Pradhan= [email protected]> wrote: I was working on 2230 which has power supply failure. |
Re: 2465B low serial - checksum & recapping
On Wed, Apr 9, 2025 at 11:53 AM, Giuseppe Marullo[iw2jww] wrote:
I'm very interested in the topic of 'recapping.' I know there was a problem with some brands of electrolytic capacitors, coming from China, circa 1999... and that has an effect on those things manufactured that used them... mostly from China. Yet...IME... what I have seen is more damage to the PCB, then good. What's the motivation for ordering a capacitor 'kit'... and doing 'shotgun' replacement of capacitors? IMO... if one does not take the time to test... for a bad component... and then to identify, and specify the appropriate component.... Well how do you know, you are not replacing something, with something that is worse? Also... but importantly, using inappropriate de-soldering, and soldering tools, and supplies, risks damage to the pcb... which might not be trivial. -- Roy Thistle |
Re: 2465B low serial - checksum & recapping
If you have an A or B series, you can get the calibration values back in the NVRAM with this utility (or with GPIB, if you have the equipment):
toggle quoted message
Show quoted text
/g/TekScopes/files/24xxAB-writecal-1.1.zip The only way to get them back into a 2445/2465 (plain) that I know of at the moment is via the optional GPIB interface (OPT 10). A similar utility to the above for A/B versions could be developed for 2445/2465 scopes, but ER1400 failures are far and few between (unlike dead batteries), and virtually no one has backed up their 2445/2465 cal data so it can be used with such a utility. If you have a backup of your cal data and are trying to get it back in the ER1400, please speak up! The above zip archive has demonstration code in ./related/earom.c on how the Tek "spiral add" checksum algorithm works on the 2445/2465, plus how parity is calculated for the range of locations that include parity. So, you might be able to determine the offending entry by putting your EXER 02 data in the code. If you want to post your 2465 EXER 02 data, we could also look at it together in this thread. Theoretically, you could force the scope to recalculate the checksum by performing a calibration step, such as CAL 04, where you don't actually modify anything and then exit the cal procedure. I haven't tried this, but keep in mind if you do you're ignoring bad cal data that could affect the scope's operation. From a reliable source, I am told there is an undocumented screen utility similar to EXER 02 for *entering* the data, but the person I spoke to can't recall how it was done. More work with a better logic analyzer (which I have now), and/or analysis with Siggi's 2465 Ghidra image and 2465 MAME simulator may find the answer. It's on my to-do list. -mark On Thu, Apr 10, 2025 at 02:52 AM, Mr Lurpak's Electronics Rescue wrote:
|
Re: 2465B low serial - checksum & recapping
Jon,
many thanks but: 1) I have hopefully a good capture video of all the CAL 02 values. 2) I found a lab that claims they have all the required equipment to fully calibrate it I was wondering: a) if there is a method to be certain that the values I captured in video can be checked against the checksum(no GPIB card). I have found a source code for the spiral checksum but not sure how to apply to the data. Not much of a cryptologist, but still can code something. b) assuming I will make recalibrate it in the lab (still deciding about it) I would like to recap the scope prior to recalibrate it. So I am looking for a BOM for Digikey or Mouser. a reasonably priced kit (price+customs and so on, shipped to Italy) can be a viable option. Hope it is more clear. -- Giuseppe Marullo IW2JWW - JN45RQ |
tektronix 2712 sa
Dear tektronix fans
To introduce myself Hubert ON6JZ living in Belgium and an electronics repairer for 45 years I now have a tektronix 2712 on the table that does not work. It has been idle for over a year. At startup it gave all kinds of error messages , after replacing both batteries on the digital board and on the digital option board it seems to start up normally , But I get no response from any button and the encoder does nothing either All supply voltages are correct and viewed cleanly with a scope I replaced all the smd electrolytics on the sweep board and on the digital storage board No result On the screen I get an image of out of memory. Does anyone have any tips on where I can look for this error? Thanks in advance Hubert |
Re: 2465B low serial - checksum & recapping
What do you do with the values read from CAL/EXEC function. is there a way to enter/change them?
I am asking as I have a 2465 as well that just had 04 10 which is describes by Tektronix as just out of cal (guessing old data or checksum fails. ER1400 can hold data for 10+ years) but TBH the cost of calibration if you dont have the very specialised TMG is not worth it for a scope this age (my opinion as you can by newer versions for 3-500GBP) or go crazy old skool and aim for 4xx series IF BW is the issue. I am for fun experimenting with reading the data out of the ER1400 (potentially replacing is if it is bad - I got like 10 spare ones). I am not too fuzzed about accuracy as I have digital never scope as well if needed). Just curious to understand the logic behind what is stored in ER1400, checksums, etc and if it can be trigged into thinking it is calibrated and don't fail on selfcheck. Cheers. |
2465B low serial - checksum & recapping
CHECKSUM
I would kindly like to know, assuming I dumped correctly the CAL 02 values, if there is a sound method to verify the checksum for my scope before fiddling with the battery. Any pointer before I get mad with X-Way forensics editor? I found a guy that could re calibrate it for 300EUR (maybe 250, we have two scopes) so I could just avoid the hassle to change the battery myself and/or switch to FRAM. It seems it could even be a reasonable price, I don't know. RECAPPING Of course, if I have someone to recalibrate it, I may well think of recalibrating it AFTER recapping. Is there any current BOM for Digikey/Mouser available? TIA. -- Giuseppe Marullo IW2JWW - JN45RQ |