Re: 2247A PSU Troubleshooting
Variacs are tapped autotransformers. No isolation. You can get an assembly that combines a variac with an isolation transformer. Some companies make a device specifically for checking leakage and
By
Harvey White
·
#155350
·
|
Re: 2247A PSU Troubleshooting
In general, no. A Variac is just a primary with an adjustable tap. Variacs with isolation exist, having their adjustable tap on a secondary winding. Raymond
By
Raymond Domp Frank
·
#155349
·
|
Re: 577 Curve tracer ringing CRT and noisy step generator
/g/TekScopes/album?id=86536&p=Name,,,20,1,0,0 Updates to the photo gallery, probing the base of the 177 test fixtures reveals small parasitic spikes, I believe this is what is causing
By
DW
·
#155348
·
|
Re: 2247A PSU Troubleshooting
Interesting information Brian. I have often wondered if a Variac would provide isolation also. I have never checked mine I probably should. Regards John
By
John Williams
·
#155347
·
|
Re: 2247A PSU Troubleshooting
I was only answering to the comment about isolation transformers. If any user gets one then I hope they can check them out to be sure that they are safe instead of expecting it to work & getting
By
Brian Symons
·
#155346
·
|
Re: Test input for an SD-32 sampling head
And if you do get a 2ps rise pulse generator how do you check that that is up to its spec? Roger
By
Roger Evans
·
#155345
·
|
Re: Tektronix 015-0580-00 Pocket Signal Generator
Thank you for the additional information. I was curious if it would be of any interesting use when I came across it by chance on ebay. I do not own any TDS series scope and upon better understanding
By
Alex
·
#155344
·
|
Re: 2247A PSU Troubleshooting
Brian, Let's not confuse the OP. He has a 'scope with a HV (mains switching power supply) part that is galvanically connected to the mains and a LV part that is under normal conditions separated from
By
Raymond Domp Frank
·
#155343
·
|
Re: Test input for an SD-32 sampling head
With difficulty. I believe there's a method involving three sampling heads looking at each other's kickout pulses and some maths which can be used to evaluate their performance. The SD-24's pulse
By
cmjones01
·
#155342
·
|
Re: Not to be missed by 7854 fans...
John, Great find! Thanks for sharing. I¡¯m looking forward to seeing this! https://youtu.be/3Co8Z8BQgWc A hoard of never-developed film from the Apollo 11 mission was unearthed, some of it in 70mm.
By
Glydeck
·
#155341
·
|
Not to be missed by 7854 fans...
Tek content begins around 8:58. Specifically, this is part 7 of CuriousMarc's series on restoring an Apollo Guidance Computer, but the whole video (and whole series) is great. You really feel like
By
John Miles
·
#155340
·
|
Re: Looking for Sony-Tektronix 318 Logic Analyser Operator's Guide
Debugging I2C data streams, FPGA outputs, video data outputs from graphics chips. Have a CPLD standalone board that takes apart I2C communications and drives an HP logic analyzer. Goes from time to
By
Harvey White
·
#155339
·
|
Re: Looking for Sony-Tektronix 318 Logic Analyser Operator's Guide
Logic analyzers are very useful, but somewhat limited in what they do. They really were needed in the times of 8 and 16 bit microprocessors, and are less useful for microcontrollers. However, since
By
Harvey White
·
#155338
·
|
Re: 2247A PSU Troubleshooting
*Re: Isolation Transformers* When using an isolation transformer for troubleshooting you do not want an earth being connected through. The whole idea is to break any reference between earth & the
By
Brian Symons
·
#155337
·
|
Re: 2467B geometry
Hi Tom, That occurred to me earlier, it is -970V, which seems a little low, but there is no information on tolerance. I measured it with both a 20M DVM, and a 200M 410C. It would be nice to know what
By
Chuck Harris <cfharris@...>
·
#155336
·
|
Re: 2467B geometry
How about the -1kV supply to the PDD electrode? It is regulated mostly by four series 200 volt zeners. If one of those were shorted, it would lower that supply and affect the geometry adjustment. You
By
Tom Miller
·
#155335
·
|
Re: 2467B geometry
I tried pulling the plug on trace rotation, and on y-axis rotation, and the only effect was the trace rotated one line width clockwise. I guess that eliminates them, and any fringing effects from the
By
Chuck Harris <cfharris@...>
·
#155334
·
|
Re: 2246 no display/odd behavior
Hi Matt, Did you check the various low voltage supplies before you took the power supply out of your 2246? Knowing what the individual voltages are, and knowing how much ripple is present on them will
By
tom jobe <tomjobe@...>
·
#155333
·
|
WTB 012-0648-00 6 pin LEMO M-M 30" cable for Tracking Generator
I got my TR502 Tracking Generator, DC508 Option 7, and TM503 Option 7 all set up and I then discovered I need the 6 pin LEMO cable (3 male pins and 3 female pins) that goes from the Spectrum Analyzer
By
Dennis Tillman W7pF
·
#155332
·
|
Re: Looking for Sony-Tektronix 318 Logic Analyser Operator's Guide
Since I only have the 308, I can't say about the tektronix ones. However, HP makes things in the 5000 series (such as a 5005B) that are signature analyzers, and peak reading voltmeters... Harvey
By
Harvey White
·
#155331
·
|