Re: 11801 diagnostic help
Albert, Thanks. The exit and debug buttons are the only ones that are not active when it stops. Setting "stop on error" off has no effect. With NVRAM replacements on the way and no service data yet, I
By
Reginald Beardsley
·
#155390
·
|
Re: vintageTEK museum releases Replaceable Parts Registry (RPR)
152- Diodes are now uploaded to tekwiki.org
By
Dave Brown
·
#155389
·
|
Re: 11801 diagnostic help
Hi Reg, IIRC you can exit the Extended Diagnostics screen by pressing the Exit touch-screen field two times. Another possibility to have the instrument at least doing something is to bypass the Self
By
Albert Otten
·
#155388
·
|
Re: 11801 diagnostic help
It's the 11801C manual despite what it says on the website. It's enough different I can't even Identify the boards using it. My original problem is solved and I have NVRAM replacements en route from
By
Reginald Beardsley
·
#155387
·
|
Re: 11801 diagnostic help
At KO4BB there is the factory diagnosis manual. Pretty cryptic, but has lots of information. Service manuals are available, but no schematics..
By
Bob Koller <testtech@...>
·
#155386
·
|
Re: 11801 diagnostic help
Albert, The reseating resolved it not getting past the power on test. I've got NVRAM replacements on the way from the UK courtesy of another member of the group who got the last two that RS had in
By
Reginald Beardsley
·
#155385
·
|
Re: 11801 diagnostic help
Hi Reg, The first thing I look at is the total Ontime and power-ups count. I have seen11801x/CSA803xx with useful display but Ontime of over 140 khrs. That must be due to some out-of-range condition;
By
Albert Otten
·
#155384
·
|
Re: 11801 diagnostic help
Be careful what you wish for. Now I need to build a pulse generator with a fast enough edge to test it. I am still a bit amazed I was so lucky. though for a couple of hours I was going, " What have I
By
Reginald Beardsley
·
#155383
·
|
Re: 11801 diagnostic help
I'm envious of your scores!? Some day I will get back into owning and operating 11000 series gear....Jim Ford?Sent from my Verizon, Samsung Galaxy smartphone
By
Jim Ford
·
#155382
·
|
Re: 11801 diagnostic help
No one in the US will have the ST part in stock until June! But someone in the UK has offered to get a couple for me there. I bought the scope for $100 plus $85 shipping because I knew what the E5622
By
Reginald Beardsley
·
#155381
·
|
Tektronix 2465B Test 05 Fail 40 repair.
This seems to be a common problem so I thought I would report on my repair. Summary: Apparently fixed by replacing the U500 hybrid with a known good one. Details: This is my main scope. In 2015 it
By
MaxFrister
·
#155380
·
|
Re: 11801 diagnostic help
Tek 11801 Timebase Errors Tektronix 11801, 11802, and CSA803 sampling scopes eventually fail with an E5622 timebase error. If you remove the bottom panel, you'll see a cluster of sram chips... Two of
By
Tom Miller
·
#155379
·
|
Re: 11801 diagnostic help
I'm not sure I can identify the board. There are a couple of boards with no markings. I reseated the 4 boards on the left side while checking to see what they are. the left two are a memory board and
By
Reginald Beardsley
·
#155378
·
|
Re: 11801 diagnostic help
What is the voltage on pin 28 of the two socketed rams on the timing and control board in the bottom of the unit?
By
Tom Miller
·
#155377
·
|
11801 diagnostic help
My 11801 arrived today. On power up it makes two beeps and puts up two lines of text on the screen: Diagnostics in progress Comm Test in progress But does not go past that. The seller sent a picture
By
Reginald Beardsley
·
#155376
·
|
200/300 series scopes at stuff day
I have had several requests to list all the tek portables that will be at stuff day, so here's the info. We will be checking these next week, so status on every unit is still unknown as of today
By
walter shawlee <walter2@...>
·
#155375
·
|
Re: 577 Curve tracer ringing CRT and noisy step generator
/g/TekScopes/photo/86536/11?p=Name,,,20,1,0,0 Probing pins 2 and 6 of U220 reveals square waves that are not symmetrical, should I be concerned and is there a adjustment
By
DW
·
#155374
·
|
Re: M48Z35-70PC1 NVRAM for CSA803 & 11801 E5622 error
I've done exactly this to my 2430A. I've replaced both of the dead Dallas NVRAM chips with Cirrus FM18W08 FRAMs in off-the-shelf SOIC to DIP converters (Aries LCQT-SOIC28), all bought from Farnell in
By
cmjones01
·
#155373
·
|
Re: 2247A PSU Troubleshooting
If they are marked ZM or ZS they are well known to get leaky. Replace these diodes with MUR160 600V 1Amp from Mouser/DK.
By
Alex
·
#155372
·
|
Re: Test input for an SD-32 sampling head
Hi, Reg. Regarding fast pulses and such, there is a wealth of information at Jim Andrews' page here: https://kh6htv.com/pspl-app-notes/ He founded Picosecond Pulse Labs in 1980 and sold it to
By
Jim Ford
·