Re: 2465B auto measurment error


 

¿ªÔÆÌåÓý

Did you try to?re-do the vertical calibration? Can you look at the waveform during the measurement process to see if it is loading the input down?
?
?
Strange one. Hope you find it.
?
Tom
?

----- Original Message -----
Sent: Wednesday, January 02, 2013 6:58 PM
Subject: [TekScopes] Re: 2465B auto measurment error

?


Any ideas?

B
--- In TekScopes@..., "andersen_bill@..." wrote:
>
> I posted a few months ago on this scope.
>
> I was not able to get this resolved at the time and got busy on other projects.
>
> I want to take a fresh look at it now.
>
> Here is an example of what I am seeing.
>
> Using a function generator with 50 ohm termination, frequency set to .5 MHz, volts 4.0 peak to peak- out of the function generator.
>
> If sending a square wave the scope will auto measure the voltages corectly.
>
> Example;
>
> peak to peak 2.02 volts
> pos peak 1.01
> neg peak 1.01
>
> If I change to a triang. wave. (frequency and voltage the same) the scope will now not measure the voltages correctly.
>
> Eample;
>
> peak to peak 1.1 volts
> pos peak .97 volts
> neg peak .14 volts
>
> If I measure manually using the cursers, they are correct.
>
> I verified the trigger cal. and did a DC balance.
>
> Please help.
>
> Thanks, B
>