Re: 2465B auto measurment error
Any ideas?
B
toggle quoted message
Show quoted text
--- In TekScopes@..., "andersen_bill@..." <andersen_bill@...> wrote: I posted a few months ago on this scope.
I was not able to get this resolved at the time and got busy on other projects.
I want to take a fresh look at it now.
Here is an example of what I am seeing.
Using a function generator with 50 ohm termination, frequency set to .5 MHz, volts 4.0 peak to peak- out of the function generator.
If sending a square wave the scope will auto measure the voltages corectly.
Example;
peak to peak 2.02 volts pos peak 1.01 neg peak 1.01
If I change to a triang. wave. (frequency and voltage the same) the scope will now not measure the voltages correctly.
Eample;
peak to peak 1.1 volts pos peak .97 volts neg peak .14 volts
If I measure manually using the cursers, they are correct.
I verified the trigger cal. and did a DC balance.
Please help.
Thanks, B
|