开云体育

ctrl + shift + ? for shortcuts
© 2025 开云体育
Date   
How to recover the video of my graphic 54810-66505 card 2
Hello, I have picked up an oscilloscope 54810 I have some problems on it, but it’s settled. Meanwhile, I tried to put the graphics card 54810-66505 on another motherboard (which also had its video output built in), for tests. Since the video card no longer provides output. The screen is black. If I put an old video card trident, the screen works. I think my manipulation must have "written" information in the chip 65550. (Some register?) What manipulation could reactivate it? NB: the card is visible in the bios pnp/pci parameters Your opinion. cdt
Started by Gerard @ · Most recent @
A homemade diode power sensor for HP meters 81
Hi guys, as you all know, the HP 435A, 436A, 437A and so on are quite nice power meters and the meter itself is availabe at quite low prices. However, the power sensors are quite expensive. I own a 8481A and a 8484A, both in nice and working condition, but I don't want to damage anything so I treat them like raw eggs and handle them extremely careful. So I always thought whether it would be possible to make a own diode power sensor which is compatible to the HP 436A. This would have the advantage that the 436A could be used (and therefore its HPIB capabilities), and if it works with the 436A, it will do so also with the other meters (and possibly others, I don't know). Further, the homebrew power sensor would be cheap and if it breaks, no matter, it can be repaired easily. So today I spent a few hours on making my own power sensor. I checked the schematic of the 8481A power sensor, and I did basically the very same. I came up with this schematic: (I don't want to waste valueable web space for the group, so I uploaded the high resolution pictures to my webspace) https://hb9fsx.ch/wordpress/wp-content/uploads/2018/11/20181117_173101.jpg The letters on the right side are the connector pins for the HP 436A, of course. Then I made a quick 'dead bug' style prototype, as follows.... (please don't laugh at me :-) ). https://hb9fsx.ch/wordpress/wp-content/uploads/2018/11/20181117_170657.jpg I connected to the 436A and I also installed a SMA 30dB attenuator at the input and connected my power sensor to the power reference of the 436A. At first, the readings were quite off a bit, but using the little trim pot on my PCB, I was able to precisely adjust the gain of my sensor such that the reading on the power meter was exactly 0.00 dBm. I then connected a signal generator to the power sensor and tested various power levels between -10dBm and +20dBm. Incredible, but my power sensor was accurate to 0.1dB! I also tested different frequencies between 1MHz and 2.6GHz. The flatness was not extremely good, but between +/-1dB. So, basically, the concept seems to work. I also tested different ranges for the power sensor - there is a 'mount resistor' inside the HP power sensors (8481, 8482, 9494, ...) which tells the power meter which range the sensor uses; the meter then internally adjusts the gain according to the power sensor range. It appears that it should be possible to home brew such power sensors. However, at the moment, I have two issues: a) the auto zero function does not work - even if I run the autozero feature, the power meter afterwards displays -13dBm anyway, so I cannot properly zero my power sensor. At the moment, I do not understand why this is so. Has anyone on the list an idea? b) I don't understand exactly how the amplifier using the NPN transistor works. When I was adjusting the gain of my sensor, I tried to change the collector resistor (330 Ohms), but this had almost no effect on the gain. The emitter resistor (1k Ohms) also had only little effect on the gain. Only when I placed the pot where it is now I was able to adjust the gain in a wide range. So it basically looks to me like this amplifier is a common emitter amplifier, with the gain being RC / RE, is that true? why then is a feedback from the emitter required to the meter? Thanks for any hints, tips and so on, Tobias HB9FSX
Started by Tobias Pluess @ · Most recent @
looking for service manual for HP4934A 5
Hello, Any one have a service manual for a HP 4934A? An electronic copy? Any good ideas where to find either a paper or electronic copy? It is for a TIMS. Transmission impairment measuring set. thank you for any help Brian
Started by w7fdd @ · Most recent @
Agilent/Keysigh ESG series CLIP 45
Hi Fellows, Last offer before retirement... CLIP P/N is E4400-90336 Model covered: E4400B, E4420B, E4421B, E4422B, E4423B, E4424B, E4425B, E4426B, E4430B, E4431B, E4432B, E4433B, E4434B, E4435B, E4436B I’m sure that also E44xxA series are covered because i have also ESG-D3000A = E4432A and boards/modules of this guy are also covered. Keysight seems didn’t put E44xxA series in the Find-A-Part database because is older than B series. Was a lot of work to scan over 1000pages...I'm asking 35USD or 30EUR When we did a scan, not optimal scanner parameters were used... scanner ignored inductors (spring like lines) but references and values are there. Once we saw this, was to late to do job again. Zip file size is about 122MB I told, not perfect scan quality but totally useful, please check attachment. Please contact me on my private email: milan.vasic@... BR, Milan
Started by Milan Vasic @ · Most recent @
5041-8820 Rear Strap Handle Cap — 3D print files? 5
Does anybody have a 3D model of Agilent 5041-8820 Rear Strap Handle Cap? Mine has shattered and I'd like to replace it. I have the metal insert so only need plastic outside piece. Cheers Leo
Started by Leo Bodnar @ · Most recent @
HP 8673D synthesized generator - need help identifying the two PC boards attached to A1A11 power amplifier 4
I'm on (the other) HP 8673D repair project now. This one never worked in the time I have owned it, and I'd been storing it as a parts mule, but decided (now that I have some experience with disassembly and repair) that I'd like see if I can bring it back to life. Primary symptom: Very low output (rarely approaching 0 dBm, at some frequency ranges much less), unleveled indicator (as the available RF would never reach the setpoint of the ALC circuit), yet never an "unlocked" indication. Fault isolation so far: The service manual vol 2 08673-90022 (in section BD5) walks through several tests; it turns out RF level (so-called "baseband") 2.0-6.6 GHz out of YTO (then through a bias tee, modulator for pulse modulation, and a high pass filter) measures with adequate RF on all YTO frequencies (spec is >= 10 dBm; this is satisfied with about +18 dBm at bottom end of range and staying above +10 dB to the end of the range.) This signal is sent to the YTM assembly; output from YTM is very low and fluctuates widely across frequencies. The service manual details RF level tests at the high pass filter output (what goes to the power amplifier which is the start of the YTM system) and RF level tests at output of the YTM (this output is very low and uneven across frequencies, resembling output of the 8673D overall, so fault is in this area. Further attempt at fault isolation: The YTM assembly includes a power amplifier (A1A11, part number in my (preliminary) manual is 08673-67004. The popwer amplifier in this 8673D part number is 08673-67011. The manual provides zero detail (that I can find) detailing expected output of the power amplifier that feeds the YTM itself. As that connection was SMA, I connected A1A11 power amplifier output (via attenuation) to the power meter and found ~0 dBm output (varying widely across frequencies), and after a few minutes in this configuration, A1A11 amplifier output went to zero and never returned. Unsuccessful search for details in service manual: Attached are photos of the A1A11 amplifier and two associated PC boards, part numbers 08673-60140 (provides DC to the A1A11 power amplifier via pins E1 through E12) and 08673-60139. The DC test points on 08673-60140 (upper/larger board in photo) expect +5, -3.5, -6.5 and -9 VDC; I measured these (in system while running) at +5, -6.5, -6.5 and -9 (+/- some mV), so the -3.5V test point's voltage was incorrect. I'm not able find schematics of either board in the (large/heavy) service manual, hence my posting here - does this ring a bell with anyone that may have encountered these modules in another context, has worked on an 8673D in this area, or has seen these mentioned in the service manual volume 2 (document number 08673-90022)? In modern times, getting more gain and power at 2.0-6.6 GHz would not be at all difficult, however, the YTM is sensitive to the level of RF input, and it appears visually that the circuitry around the A1A11 amplifier is at least in part allowing the CPU to control output from A1A11 to keep drive within some (undocumented) acceptable level window. Thanks for any clues! Dave
Started by David Feldman @ · Most recent @
Threads for screws on 8656 Siggen 28
Hi can someone tell me what thread the screws are that hold the back panel on HP8656. I am assuming they are 5/32 as 3mm and 4mm metric are either too small or too big. or wrong thread Are the UNF , UNC or something else? imperial thread screws are not common in Australia, and I have lost all teh screws out of the back of my sig gen. I think the ones i ordered were bsw and of course don't fit Wayne VK2WDE
Started by Wayne Eckert @ · Most recent @
HP Series 200/300 and Plotter Manuals
HP series 200/300 and Plotter Manuals PM direct for offer… 7470A Plotter, Statistics, Graphic Presentation, Project Management, Basic 4.0… Interface Techniques, Utilities, Waveform Analysis, Graphic Techniques, Asynchronous Terminal Emulator Thanks, Alan
Started by alan victor @
File /All HP, Agilent and Keysight instruments in folders by part numbers/3000 to 3999/33xx Level detectors and synthesizer level generators/3336 A-B-C Section VIII - Service.pdf uploaded #file-notice
The following items have been added to the Files area of the [email protected] group. /All HP, Agilent and Keysight instruments in folders by part numbers/3000 to 3999/33xx Level detectors and synthesizer level generators/3336 A-B-C Section VIII - Service.pdf By: Harold Foster <halfoster@...> Description: Combined schematics and diagrams for the "Section VIII - Service" portion of the 3336A-900000 manual
Started by Group Notification @
Added photo album Microscope #photo-notice
Radu Bogdan Dicher <vondicher@...> added the photo album Microscope
Started by Group Notification @
where to find HP 8511A,B and Antenna Measurement System Performance,Verification Software. 3
I might be able to get an HP 8511A...
Started by John Griessen @ · Most recent @
hp8970a recallibration attenuator help 6
hello all I have a hp8970A meter that has been turned off for a long time. I turned it back on and I know that the internal attenuator needs to be recalibrated with the special command function SP 33.0. I re-entered the ENR table of the head but I don't remember if the temperature also needs to be entered or not just for the recalibration of the attenuator does anyone remember? thank iw5bsf Roberto
Started by Roberto @ · Most recent @
HP8510C TRL Calibration 14
Went to great lengths to create PyVISA program to read using the GPIB the x4 parameter S-parameter data from the HP8510C. This eventually would read consecutive 1701 point sweeps in a round 5 seconds using the HP binary data format; ASCII data formate took much longer. That was for uncalibrated data. However, when i tried to read TRL calibrated data it would take over two minutes to read the data. The obvious solution then would be to just read the uncalibrated data and perform the TRL calibration in the PC. I could trawl through endless papers to find a suitable mathematical equations to code up in Python. From what i gather, mathematically its quite complicated, solving something like 12 simultaneous equation, clearly doable, but i've also got a massive list of other stuff to do. Might anyone have an algorithm to perfor the TRL calibration, or know where i might find this? many thanks, N
Started by neil @ · Most recent @
3" disks for 8566A/B, 8568A/B on HP 9000 Series 200 Computers - Any interest in images? 10
Hello Forum, I think I posted this before but no harm in bumping... Together with an HP8568B, I got a number of 3?-inch disks: 85863-10027 Rev C : 85863A Opt 630 Basic Software Library 08568-10004 Rev A: HP 8568B Spectrum Analyzer Operation Verification 08568-10004 Rev A: BASIC Operation Verification for 8568B 85863-10027 Rev B: 85863A Basic Software Library for 8566A and 8568A Opt 630. All disks (C) 1984 and two of them marked "For HP 9000 Series 200 Computers with Basic 2.0 or 3.0 O/S Any interest in images? I do not know if any data remains on them. If so, How to read and make images on a PC running XP or W7? Cheers Ulf Kylenfall SM6GXV
Started by Ulf Kylenfall @ · Most recent @
8643A failure 4
My 8643 failure message indicated 5V supply failure. Voltage at test point measured 4.9 volts. I replaced all the capacitors in both switching supplies of the 8643A. Both supplies checked out OK before capacitor changes, and OK after capacitor changes. But I'm happy it has new capacitors. The battery voltage was OK, but being old and not interested in leaks, it has a new set of 3xAAA in little plastic holder. After reading about tinned Molex connectors developing voltage drops I sprayed DeOxit on pins of the power supply connector, and both input and output connectors of the power filter board. And of course the cable pins that plugged into those connectors. Power supply and filter board input connectors were tinned pins. Pins on the output end of the filter board were gold. Did the best I could, but it wasn't as good a job as I would like to have done. NONE of the connector bodies were discolored. There was maybe 1/4 volt drop, and not that many hours of use, before I took it apart. Probably not enough time and heat to discolor the Nylon. Applied power, and test point now measures 5.10 volts. So looks like it was voltage drop in the tinned Molex connectors. No error message. While I haven't checked exhaustively to verify everything is working, but I expect it's OK again. THANK YOU GUYS FOR POINTING TO THE CONNECTORS Regards - Bill W7AAZ
Started by Bill Carver @ · Most recent @
HP 8643A and Capacitors 16
My 8643A started doing a "recalibrate" every time I turned it on. Once the 8643A completes that "recalibrate" its RF output and frequency are OK, although I have not tested every aspect of modulation, sweep, etc. I assumed it was failure of the memory backup battery, but it was original and OK. But old and leakage would be a disaster, so replaced the battery with three AAA NiCad. Still "recalibrated", as expected. The fault message at the end reported the 5.25V supply was not within spec. I checked a test point at the output of the power supply filter board and found less than 5.0V. Having experience with tired computer grade capacitors I assumed that would be the case here. Surprise: no transformer supply with big caps, just two switchers. I pulled both switching supplies and found that both the oven switcher and the main 5V/15V supplies voltages were OK. Particularly the 5.25 I tested with a 2A load current and it was perfect. All capacitors in both supplies with Nichicon 105C and operating at a fraction of rated voltage. Nonetheless, considering the age and the disassembled state of the 8643A I have all new capacitors (again Nichicon 105C which were also my choice when I was working). I have replaced caps and retested the 24V oven switcher. The last capacitor for the main switcher is supposed to arrive today and I expect the main supply will test OK by tonight. Out of curiosity I tested the removed capacitors and all but one exceeded the rated capacitance and had tiny leakage current. It appears I am losing about 300mV between a supply at 5.25 volts and the connector at the output of the power-filter board. Something drawing too much current? It would have to be a seriously large amount of current. That seems unlikely given that it seems to be working. All DC supplies pass through a passive filter board with series inductors, shunt capacitors. With the supplies still out and the filter output connector disconnected, I will measure the shunt impedance of that filter board. Anyone with 8643A experience have any other ideas? Bill On 3/7/2025 7:51 AM, AArnaud via groups.io wrote:
Started by Bill Carver @ · Most recent @
HP859x Power Supply Schematics 23
Hello all, This is my first post since I joined the list. I have two 8591E Spectrum Analysers, one of which has a completely dead power supply. The other was working, until I powered it up a few weeks ago, when it gave the ADC-2V error. Once again it has a PSU fault, this time the -15V had failed. Investigation revealed a short circuit to GND on the -15V Output from the PSU. I took the unit apart and traced backwards. I soon found that C263, a 22uF Tantalum was to blame. Unfortunately this didn't fix the problem fully, there is still only -0.6 V on the output. Like many others I failed in finding much information about the circuits, so needed to investigate further. I don't have an isolation transformer, so I can't safely work on the unit live so I decided to trace the circuits to understand how the supply works. I have completed the first pass trace of all three boards and understand enough now to know that the fault is either in the -15V output regulator, or the reference passed to it from the control board. Once I have completed the circuit trace I should be able to power up the boards independently to test the circuits. I have checked the Output Regulator board and created an electronic schematic and component layout which are attached to this post. I am going to do the same with the other two boards. Hopefully this will help get some of these PSUs repaired, as they seem to be the weakest part of an otherwise generally reliable design. Chris
Started by Chris Weir @ · Most recent @
HP8510C CalKit file .CK_ format 3
Hi there, I'm trying to generate the calkit file CK_ programmatically instead of punching them in various GUI based CalKit manager. Basically, I'm doing my homemade open and slot standards and characterize them to use as poor man's Calkit for non-critical measurement to reduce the wear and tear on my expensive Cal Kit. Manually entering the polynomials and class assignments in the VNA CalKit manager is very slow and painful. I'm trying to generate it a C program so that I can send it to my 8510 either by GPIB or floppy disk but I can't find the format of the 8510 CK_ file. Does anyone know? Regards, Calvin
Started by Calvin Guan @ · Most recent @
Artek manuals 6
Hello, Is Artek manuals still alive? I've sent two requests for manuals by email, one in February and another in March and have yet to receive a reply (or a failure to deliver notification) to either one. Regards, Waz
Started by Warren @ · Most recent @
HP 4192A Sweep Utility / Michael Kelzenberg 6
Hi, Does anyone know what happened to the website for the HP 4192a sweep utility written by Michael Kelzenberg? I didn't have it bookmarked so don't recall the URL, I think it was on a university website. It's completely disappeared from search results online. I luckily have a copy of the installer I downloaded in November, but someone else asked me where I got it and I can't find it anymore. It would also be nice to have a copy of the text of the site for reference. Thanks! BTW, the file name of the installer is "Install-4192A-V1-full.exe". Even that doesn't show in any search results. Thanks
Started by pac1085 @ · Most recent @
Current Image
Image Name
Sat 8:39am