Hi Ed, i did manage to get the NI GPIB / USB to laptop interface working. In terms of running the 8510C from the laptop it seems to work quite well but only with uncalibrated data. From the Operation and Programming Manual i tried all fives forms of data transfer: FORM1 to Form5. FORM1 was a little tricky and the manual doesnt really explain for format well. After trial and error all forms were giving the same numerical values, which gave some confidence that i had the programming correct. FORM5 was the fastest, collecting data for a 801 point sweep in about 6 seconds. Slow by today's standards, but not bad for vintage 1991.
?
However, when i ask it to do calibrated data, the machine peeps indicating there's an error and still continues to show data transfer increase from 6 seconds to 100 seconds, so clearly that's not much good.
?
How do i find which software version i have running. On the display i see the information: HP 8510C Monitor C.02.16 March 9 1991, so is C.02.16 the software version number?