¿ªÔÆÌåÓý

Date
Re: Looking for a feedthrough capacitor
You probably have to enter it on the german site, I don't know. Its a guy who sells 5 PCM testers for 370 EUR. He bought them for the sole purpose of having parts for his NFAs... cheers Martin
By Martin · #133320 ·
Re: Looking for a feedthrough capacitor
Well, yes, just one instrument for both of us would be fine.... But no, thanks. Cheers....
By Harke Smits · #133319 ·
Re: Looking for a feedthrough capacitor
That item doesn't show up as anything. What was it? -- Paul Amaranth, GCIH | Manchester MI, USA Aurora Group of Michigan, LLC | Security, Systems & Software paul@... | Unix/Linux - We don't
By Paul Amaranth · #133318 ·
Re: Looking for a feedthrough capacitor
Hi Harke, item #294954912663 on a well known auction site would probably yield a good stock of these. Not a bargain, unfortunately. cheers Martin
By Martin · #133317 ·
Re: Looking for a feedthrough capacitor
Hello Martin, I immediately recognized this feed-through capacitor. Seems typical for WG as my Rauschpegelsender RS100 has the same. One showed a short and as I considered it as unobtanium, I just
By Harke Smits · #133316 ·
Looking for a feedthrough capacitor
Hi all, I'm looking for a replacement for a shorted feedthrough capacitor for my SNA1 Wandel & Goltermann analyzer. It seems special in that it is very thin on the interiours side, where the bended
By Martin · #133315 ·
Re: HP 8595E SA DB-25 RS232 is not functioning.
Hi again. I pulled a manual from:- https://www.manualslib.com/download/529136/Hp-8590-E-Series.html After a lot of scrolling and reading, sadly, the only references I can see for the RS232 port, is to
By Dave_G0WBX · #133314 ·
Re: HP 8341A how to install attenuator
Thank's for the information,? finally we have succeed? ?comparing one 8341A with another having an attenuator, now the unit control the attenuator nicely, obviously the explanation in service
By Guy DAIPRA · #133313 ·
Re: HP 8595E SA DB-25 RS232 is not functioning.
Thank your for valuable notes. What I did try yesterday was to clear DTR (PC) <-> CTS(SA) and set RTS (PC) <-> DCD (SA) but I didn't observed the L letter on bottom left but nothing. I'll try the COPY
By candrian · #133312 ·
Re: HP 8595E SA DB-25 RS232 is not functioning.
Correction - to the lines in the message below To get an L, SA CTS must not be asserted, because T will obscure the L. Then SA DCD must be asserted (+). If both CTS and DCD on the SA are asserted,
By k1ggi · #133311 ·
Re: HP 8595E SA DB-25 RS232 is not functioning.
Ok, getting the T is reassuring. To get an L, SA RTS must not be asserted, because T will obscure the L. Then SA DCD must be asserted (+). If both RTS and DCD on the SA are asserted, i.e. when the pc
By k1ggi · #133310 ·
Re: Firmware for HP 54200D
One more discovery on this - it appears that the D-version (54200D) and A-version (54200A) used the same EPROM set. I have found the same labels in pictures online of motherboards from both versions,
By medasaro · #133309 ·
Re: 5086-7337 *Yig Oscillator issue*
Hi All Ok about the acronyms We had RPG Rotary Pulse Generator apparently this is the HMI control knob Human machine interface (a Knob ) or as I know it RPE rotary pulse encoder RUD new one for me but
By Paul Bicknell · #133308 ·
Re: HP 8595E SA DB-25 RS232 is not functioning.
Just a correction here. The connections I have made are: SA? ? ? ? ? ? ? PC (USB RS232 adapter) P1. GND? ? ?Unconnected P2. TxD? ? ? RxD P3. RxD? ? ? TxD P4. RTS? ? ? DCD P5. CTS?
By candrian · #133307 ·
Re: HP 8595E SA DB-25 RS232 is not functioning.
I just run a test regarding your notes. You are correct when I set the DTR (+VE) from RealTerm I can see a T visible on the bottom right of the SA screen which I didn't notice in the past. When i
By candrian · #133306 ·
Re: 5086-7337 *Yig Oscillator issue*
Or rocket-propelled grenade, in another profession. In certain circumstances both can undergo rapid uncontrolled disassembly (RUD). Tom Sent from my iThing; please forgive the typos and brevity
By Tom Lee · #133305 ·
Re: 5086-7337 *Yig Oscillator issue*
Dear, RPG stand for "Rotary Pulse Generator". This is the endless knob on front panel for changing parameters. Regards.
By Roberto Vegliach · #133304 ·
Re: 5086-7337 *Yig Oscillator issue*
Don - Thanks - I have not played much with the 8350 plug YIGS (mostly 834X stuff) - good information to know. I do find it a bit surprising though. Reasoning is as follows, correct me if I'm wrong: 1)
By Bruce · #133303 ·
Re: 5086-7337 *Yig Oscillator issue*
Ruben - The RPE is likely to be an easy fix - there is an incandescent lamp that illuminates the rotating defraction grid. It can be replaced by either another lamp (available on E-bay and other
By Bruce · #133302 ·
Re: HP 8595E SA DB-25 RS232 is not functioning.
Thank you for response. The translator is an old Chinese one blocked by Windows but there is a workaround to install older drivers and make it work. That is what I had available at the time. I did
By candrian · #133301 ·