Re: HP 8595E SA DB-25 RS232 is not functioning.


 

¿ªÔÆÌåÓý

Correction ¨C to the lines in the message below

?

To get an L, SA CTS must not be asserted, because T will obscure the L.

Then SA DCD must be asserted (+).

?

If both CTS and DCD on the SA are asserted, i.e. when the pc port is open normally, only T will appear

?


From: [email protected] [mailto:[email protected]] On Behalf Of k1ggi
Sent: Tuesday, March 14, 2023 10:24 PM
To: [email protected]
Subject: Re: [HP-Agilent-Keysight-equipment] HP 8595E SA DB-25 RS232 is not functioning.

?

Ok, getting the T is reassuring.

?

To get an L, SA RTS must not be asserted, because T will obscure the L.

Then SA DCD must be asserted (+).

?

If both RTS and DCD on the SA are asserted, i.e. when the pc port is open normally, only T will appear.

?

Please look at the voltage.

?

Do you see a positive voltage on SA DCD (DB25 pin 8)? (produced when pc RTS is asserted on pc DE9 pin 7)

?

That is necessary for commands to be recognized by the SA.

?

I have option 043, your option 023 may behave differently, but if you can configure your COPY button to plot to rs232 in CONFIG, then the SA should be able to transmit rs232 data when you press COPY, regardless of whether it can receive rs232 commands.

So then if you see the T, with your RealTerm port open, pressing COPY should transmit a lengthy burst of ascii plot data that should be displayed by RealTerm. That may be worth trying.

?

I¡¯m still not sure if your pc usb-rs232 adapter needs its DSR (DE9 pin6) asserted in order to transmit. I¡¯m suspicious that the chip may need that control signal asserted in hardware even if the RealTerm software application is configured for no hardware flow control. My FTDI adapter does not. Try tying the adapter pin 6 to pin 8 (both driven by SA DTR pin 20).

?

Ed


From: [email protected] [mailto:[email protected]] On Behalf Of candrian
Sent: Tuesday, March 14, 2023 4:05 PM
To: [email protected]
Subject: Re: [HP-Agilent-Keysight-equipment] HP 8595E SA DB-25 RS232 is not functioning.

?

Just a correction here.

The connections I have made are:

SA? ? ? ? ? ? ? PC (USB RS232 adapter)

P1. GND? ? ?Unconnected

P2. TxD? ? ? RxD

P3. RxD? ? ? TxD

P4. RTS? ? ? DCD

P5. CTS? ? ? DTR? ?(measured -6V coming from the adapter)

P6. unconnected

P7. GND? ? ?GND

P8. DCD? ? ?RTS

P20. DTR? ?CTS

So when I said that I set DTR I meant DTR on PC side which is connected to CTS on SA. So when I set SA DTR thus the CTS on SA side I see the T letter on bottom left of the SA screen.
When I set or clear RTS on PC thus DCD on SA side oI can't notice any difference or the L letter appearing.?

In both options and all possible combinations of DTR & RTS in PC side I couldn't manage to send any commands and receive a response.?

I tried the following commands:
IP;
ID;
SNGLS;
*IDN?