From:[email protected]
[mailto:[email protected]]
On Behalf Of Pierre-Fran?ois
(f5bqp_pfm) Sent: 22 April 2020 09:40 To:[email protected] Subject: Re:
[HP-Agilent-Keysight-equipment] Problem with HP MMS 70000, USERLOCK ON
?
I then decided to open my 70900B, have not done this since many years
when I upgraded the firmware with the latest from my previous 70952B.
At that time I uploaded this latest firmware on KO4BB web site.
So I tested the battery and found 3.074v for a 3v battery, so my problem isn't
this.
The battery model not replaced by me is a Panasonic BR-2/3A .
So the problem seems to be more complex than expected, may be a faulty memory
??
Question: Is there somewhere a memory test that could be run to test the
memories of the 70900B memory CPU card?
Any suggestion to the person having experience with the 70900B logic ?
The problem is just when I load an HP DLP (70875A) within the analyzer I got
this problem, otherwise in spectrum analyzer mode the MMS is working fine.
pf
--
Pierre-Fran?ois, F5BQP
JN18AN
No virus found in this message.
Checked by AVG -
Version: 2016.0.8048 / Virus Database: 4793/15886 - Release Date: 08/14/18
Internal Virus Database is out of date.