Re: NanoVna H-4 stucks


 

On Mon, Jan 25, 2021 at 12:55 PM, <ossi.vei@...> wrote:


My new analyzer stucks when I turn it on. Touch Screen do not work.
It generates error: si5351-init failed.
What is wrong?
Version 0.5.0
edy 555
Kernel: 5.1.0
You probably got a device from the first production, the fault of which was that the cover was pressed against the touch screen and due to constant pressure it could not log in either.
What to do: Unscrew the cover and place a plastic washer (0.5mm) on the screw towers and screw the cover back on.