¿ªÔÆÌåÓý

ctrl + shift + ? for shortcuts
© 2025 Groups.io

Re: The case of the missing elements

 

¿ªÔÆÌåÓý

Attached is an example of how radionuclides can be distributed within a mineral specimen, showing the variability of the depth of these included sources. There is an excellent picture of my knee on one of the slides.

Steve


From: [email protected] <[email protected]> on behalf of GEOelectronics@... <GEOelectronics@...>
Sent: Friday, February 7, 2020 5:55 PM
To: [email protected] <[email protected]>
Subject: Re: [XRF] The case of the missing elements
?
Hi Charles, cool observation. If I may guess at it:

We know from experience that an atom will make XRF if excited by energy from alpha, beta (or electron stream) Gamma, X-Rays etc. Probably many other ways too.

So if you rely on internal energy to do this, it will do so in response to the available radiation from the rock. Some have LOTS, some have very little, some almost none.

One advantage to that way is all the atoms, even those deep within a big rock are being excited. If those XRF rays can get out depends only on the energy of the ray and the overburden.

Rays above 50 keV have little problem penetrating silicon-calcium-oxygen atoms.
This goes for adding external energy or gathering internally generated XRF.

So your method might show up something that an external exciter misses.

But a good exciter will work on a rock that has no discernable radioactivity, the thing is how radioactive is the rock and how radioactive is does the external exciter have to be to equal or better it.....

The pro units have a need for speed so they have X-Ray tubes, blasting the sample with as much as the law allows. We don't actually have a need for speed so we can use tiny little exciters, but as far as the elements at the surface and a little further inside are concerned, it is much more "exciting" than the stuff coming from deep inside and spread over the entire surface of the rock.

In other words it may be small, but it is concentrated where it counts, we substitute long run times to get the results we need.

Say on another note, do you have enough Trinitite to do some Gamma Spec runs on it with your Si-PIN? You could precisely calibrate the detector with Am, the see if the X-Rays coming from Trinitite are from Np or U atoms, or both.

Geo


----- Original Message -----
From: Charles David Young <charlesdavidyoung@...>
To: [email protected], Mike Loughlin <loughlin3@...>
Sent: Fri, 07 Feb 2020 17:12:15 -0500 (EST)
Subject: [XRF] The case of the missing elements

I recently picked up a specimen of melanocerite-Ce at the mineral show and did a scan without Am241 by depending on the internal radiation.? I was not surprised to see Th La Ce and Nd in the blue plot.

Then I wanted to see if I got the same results using Am241 as an additional exciter.? So I carefully moved the specimen just far enough from the detector to allow me to slide in the Am241 jig.? I was careful to keep the same spot on the rock even though it looked quite uniform.

Curiously I am now getting additional strong Fe and Y peaks in the red plot.? What could account for this?? I am not surprised that the La Ce Nd are lit up as they are.

I have also attached both .mca files in case one wishes to analyze them.? However, I am not really asking for that.? I would just like to understand why the Am241 lights up the Fe and Y so strongly compared to the internal radiation.? It is kind of a pain to weed through the Am241 background peaks but it appears that I may not get a complete picture with the internal radiation alone.

Charles




Re: The case of the missing elements

 

Thanks for that Steve, I didn't realize about the varying U/Th ratios, that makes sense.

Until Charles started mentioning then showing the self-excited aspect it never occurred to me to XRF radioactive minerals too. Usually all I did was gamma spec on radioactive rocks, and the XRF was for stable elements only. This mix of both is certainly challenging. Have you seen it referred to in the literature? I can't remember anyone ever mentioning it before.


Geo

----- Original Message -----
From: WILLIAM S Dubyk <sdubyk@...>
To: [email protected]
Sent: Fri, 07 Feb 2020 20:57:12 -0500 (EST)
Subject: Re: [XRF] The case of the missing elements


Geo, great explanation but it gets even crazier than that. Different mineral districts have different U/Th ratios, and they are different ages. So you have an unknown ratio of these elements and their decay products, and x-ray emissions from these elements
and daughter products will vary not only from one district to another, but even one mineral to another. It will be very inconsistent. An external source such as Am241 will be very consistent.


Another variable will be, as you mentioned, refraction within the target mineral, versus reflection from the exciter used. Just my thoughts.


Steve




From: [email protected] <[email protected]> on behalf of GEOelectronics@... <GEOelectronics@...>
Sent: Friday, February 7, 2020 5:55 PM
To: [email protected] <[email protected]>
Subject: Re: [XRF] The case of the missing elements
?
Hi Charles, cool observation. If I may guess at it:

We know from experience that an atom will make XRF if excited by energy from alpha, beta (or electron stream) Gamma, X-Rays etc. Probably many other ways too.

So if you rely on internal energy to do this, it will do so in response to the available radiation from the rock. Some have LOTS, some have very little, some almost none.

One advantage to that way is all the atoms, even those deep within a big rock are being excited. If those XRF rays can get out depends only on the energy of the ray and the overburden.

Rays above 50 keV have little problem penetrating silicon-calcium-oxygen atoms.
This goes for adding external energy or gathering internally generated XRF.

So your method might show up something that an external exciter misses.

But a good exciter will work on a rock that has no discernable radioactivity, the thing is how radioactive is the rock and how radioactive is does the external exciter have to be to equal or better it.....

The pro units have a need for speed so they have X-Ray tubes, blasting the sample with as much as the law allows. We don't actually have a need for speed so we can use tiny little exciters, but as far as the elements at the surface and a little further
inside are concerned, it is much more "exciting" than the stuff coming from deep inside and spread over the entire surface of the rock.

In other words it may be small, but it is concentrated where it counts, we substitute long run times to get the results we need.

Say on another note, do you have enough Trinitite to do some Gamma Spec runs on it with your Si-PIN? You could precisely calibrate the detector with Am, the see if the X-Rays coming from Trinitite are from Np or U atoms, or both.

Geo


----- Original Message -----

From: Charles David Young <charlesdavidyoung@...>

To: [email protected], Mike Loughlin <loughlin3@...>

Sent: Fri, 07 Feb 2020 17:12:15 -0500 (EST)

Subject: [XRF] The case of the missing elements

I recently picked up a specimen of melanocerite-Ce at the mineral show and did a scan without Am241 by depending on the internal radiation.? I was not surprised to see Th La Ce and Nd in the blue plot.

Then I wanted to see if I got the same results using Am241 as an additional exciter.? So I carefully moved the specimen just far enough from the detector to allow me to slide in the Am241 jig.? I was careful to keep the same spot on the rock even though
it looked quite uniform.

Curiously I am now getting additional strong Fe and Y peaks in the red plot.? What could account for this?? I am not surprised that the La Ce Nd are lit up as they are.

I have also attached both .mca files in case one wishes to analyze them.? However, I am not really asking for that.? I would just like to understand why the Am241 lights up the Fe and Y so strongly compared to the internal radiation.? It is kind of a pain
to weed through the Am241 background peaks but it appears that I may not get a complete picture with the internal radiation alone.

Charles







Re: The case of the missing elements

 

¿ªÔÆÌåÓý

Geo, great explanation but it gets even crazier than that. Different mineral districts have different U/Th ratios, and they are different ages. So you have an unknown ratio of these elements and their decay products, and x-ray emissions from these elements and daughter products will vary not only from one district to another, but even one mineral to another. It will be very inconsistent. An external source such as Am241 will be very consistent.

Another variable will be, as you mentioned, refraction within the target mineral, versus reflection from the exciter used. Just my thoughts.

Steve



From: [email protected] <[email protected]> on behalf of GEOelectronics@... <GEOelectronics@...>
Sent: Friday, February 7, 2020 5:55 PM
To: [email protected] <[email protected]>
Subject: Re: [XRF] The case of the missing elements
?
Hi Charles, cool observation. If I may guess at it:

We know from experience that an atom will make XRF if excited by energy from alpha, beta (or electron stream) Gamma, X-Rays etc. Probably many other ways too.

So if you rely on internal energy to do this, it will do so in response to the available radiation from the rock. Some have LOTS, some have very little, some almost none.

One advantage to that way is all the atoms, even those deep within a big rock are being excited. If those XRF rays can get out depends only on the energy of the ray and the overburden.

Rays above 50 keV have little problem penetrating silicon-calcium-oxygen atoms.
This goes for adding external energy or gathering internally generated XRF.

So your method might show up something that an external exciter misses.

But a good exciter will work on a rock that has no discernable radioactivity, the thing is how radioactive is the rock and how radioactive is does the external exciter have to be to equal or better it.....

The pro units have a need for speed so they have X-Ray tubes, blasting the sample with as much as the law allows. We don't actually have a need for speed so we can use tiny little exciters, but as far as the elements at the surface and a little further inside are concerned, it is much more "exciting" than the stuff coming from deep inside and spread over the entire surface of the rock.

In other words it may be small, but it is concentrated where it counts, we substitute long run times to get the results we need.

Say on another note, do you have enough Trinitite to do some Gamma Spec runs on it with your Si-PIN? You could precisely calibrate the detector with Am, the see if the X-Rays coming from Trinitite are from Np or U atoms, or both.

Geo


----- Original Message -----
From: Charles David Young <charlesdavidyoung@...>
To: [email protected], Mike Loughlin <loughlin3@...>
Sent: Fri, 07 Feb 2020 17:12:15 -0500 (EST)
Subject: [XRF] The case of the missing elements

I recently picked up a specimen of melanocerite-Ce at the mineral show and did a scan without Am241 by depending on the internal radiation.? I was not surprised to see Th La Ce and Nd in the blue plot.

Then I wanted to see if I got the same results using Am241 as an additional exciter.? So I carefully moved the specimen just far enough from the detector to allow me to slide in the Am241 jig.? I was careful to keep the same spot on the rock even though it looked quite uniform.

Curiously I am now getting additional strong Fe and Y peaks in the red plot.? What could account for this?? I am not surprised that the La Ce Nd are lit up as they are.

I have also attached both .mca files in case one wishes to analyze them.? However, I am not really asking for that.? I would just like to understand why the Am241 lights up the Fe and Y so strongly compared to the internal radiation.? It is kind of a pain to weed through the Am241 background peaks but it appears that I may not get a complete picture with the internal radiation alone.

Charles




Re: The case of the missing elements

 

Hi Charles, cool observation. If I may guess at it:

We know from experience that an atom will make XRF if excited by energy from alpha, beta (or electron stream) Gamma, X-Rays etc. Probably many other ways too.

So if you rely on internal energy to do this, it will do so in response to the available radiation from the rock. Some have LOTS, some have very little, some almost none.

One advantage to that way is all the atoms, even those deep within a big rock are being excited. If those XRF rays can get out depends only on the energy of the ray and the overburden.

Rays above 50 keV have little problem penetrating silicon-calcium-oxygen atoms.
This goes for adding external energy or gathering internally generated XRF.

So your method might show up something that an external exciter misses.

But a good exciter will work on a rock that has no discernable radioactivity, the thing is how radioactive is the rock and how radioactive is does the external exciter have to be to equal or better it.....

The pro units have a need for speed so they have X-Ray tubes, blasting the sample with as much as the law allows. We don't actually have a need for speed so we can use tiny little exciters, but as far as the elements at the surface and a little further inside are concerned, it is much more "exciting" than the stuff coming from deep inside and spread over the entire surface of the rock.

In other words it may be small, but it is concentrated where it counts, we substitute long run times to get the results we need.

Say on another note, do you have enough Trinitite to do some Gamma Spec runs on it with your Si-PIN? You could precisely calibrate the detector with Am, the see if the X-Rays coming from Trinitite are from Np or U atoms, or both.

Geo


----- Original Message -----
From: Charles David Young <charlesdavidyoung@...>
To: [email protected], Mike Loughlin <loughlin3@...>
Sent: Fri, 07 Feb 2020 17:12:15 -0500 (EST)
Subject: [XRF] The case of the missing elements

I recently picked up a specimen of melanocerite-Ce at the mineral show and did a scan without Am241 by depending on the internal radiation.? I was not surprised to see Th La Ce and Nd in the blue plot.

Then I wanted to see if I got the same results using Am241 as an additional exciter.? So I carefully moved the specimen just far enough from the detector to allow me to slide in the Am241 jig.? I was careful to keep the same spot on the rock even though it looked quite uniform.

Curiously I am now getting additional strong Fe and Y peaks in the red plot.? What could account for this?? I am not surprised that the La Ce Nd are lit up as they are.

I have also attached both .mca files in case one wishes to analyze them.? However, I am not really asking for that.? I would just like to understand why the Am241 lights up the Fe and Y so strongly compared to the internal radiation.? It is kind of a pain to weed through the Am241 background peaks but it appears that I may not get a complete picture with the internal radiation alone.

Charles




Re: Mystery wire

 

"although I don¡¯t recall that he quantified percentages of each of the metals in the mix."

Ken I don't have the program to do that. It is very costly. All I can do? is (to try to) identify elements.

The Mystery Wire insulation is very Teflon like, I haven't tried its melting point yet. We use all Teflon wire in the shop, some has a really high melting point. Because if the tiny little silver plated wires inside our hookup wire, we use a heat-stripper. It has adjustable temp and can be read with an IR thermometer.

Geo

----- Original Message -----
From: Ken Sejkora <kjsejkora@...>
To: [email protected]
Sent: Fri, 07 Feb 2020 18:37:56 -0500 (EST)
Subject: Re: [XRF] Mystery wire

Hi Dud,

?

Thanks for the very detailed analysis of the mystery wire.? I see in the description that you call it ¡®insulated wire¡¯.? Was the insulation still on the wire when you ran the analysis, or had the insulation been removed?? I don¡¯t know for certain, but suspect the insulation might possibly be Teflon, or something of similar nature.? If the analysis was performed with the insulation still on the wire, could that account for some of the low-energy (LE) response, potentially from the Carbon, Chlorine, and Fluorine, etc. if the insulation was indeed Teflon?

?

I¡¯ll have to pull out Geo¡¯s results and see how they compare, although I don¡¯t recall that he quantified percentages of each of the metals in the mix. Again, many thanks for your analysis of the wire.?

?

Have a great weekend.

?

73s? --? Ken

?

?

From: Dude
Sent: Friday, February 7, 2020 04:05 PM
To: [email protected]
Subject: [XRF] Mystery wire

?

Ken,

Attached are the results of your Mystery Wire.

The data were taken with an Olympus DP4050 pXRF. As I don¡¯t have an ALLOY calibration for this gun I used the Mining Mode to get the percentage of elements present

The set up used a 50kV 11 uA x-ray beam with a live time of 30 secs obtaining account rate of 9505 cps from the SDD detector.

?

Mn was identified but it has an interference from Cr and Fe, probably there but not confirmed

V is not present

Co is not present Fe interference

Ti is not there, interference from Si escape peak

Zn is not confirmed

LE are the Low Energy elements

?

?

?

?


?





Re: The case of the missing elements

 

¿ªÔÆÌåÓý

I think you are comparing two different files. One is Allanite the other is Melanocerite, one from

Canada one from Norway.

Send single spectra plots no overlays.

Dud

From: [email protected] [mailto:[email protected]] On Behalf Of Charles David Young
Sent: Friday, February 7, 2020 2:12 PM
To: [email protected]; Mike Loughlin
Subject: [XRF] The case of the missing elements

?

I recently picked up a specimen of melanocerite-Ce at the mineral show and did a scan without Am241 by depending on the internal radiation.? I was not surprised to see Th La Ce and Nd in the blue plot.

?

Then I wanted to see if I got the same results using Am241 as an additional exciter.? So I carefully moved the specimen just far enough from the detector to allow me to slide in the Am241 jig.? I was careful to keep the same spot on the rock even though it looked quite uniform.

?

Curiously I am now getting additional strong Fe and Y peaks in the red plot.? What could account for this?? I am not surprised that the La Ce Nd are lit up as they are.

?

I have also attached both .mca files in case one wishes to analyze them.? However, I am not really asking for that.? I would just like to understand why the Am241 lights up the Fe and Y so strongly compared to the internal radiation.? It is kind of a pain to weed through the Am241 background peaks but it appears that I may not get a complete picture with the internal radiation alone.

?

Charles


Re: Mystery wire

 

¿ªÔÆÌåÓý

Hi Dud,

?

Thanks for the very detailed analysis of the mystery wire.? I see in the description that you call it ¡®insulated wire¡¯.? Was the insulation still on the wire when you ran the analysis, or had the insulation been removed?? I don¡¯t know for certain, but suspect the insulation might possibly be Teflon, or something of similar nature.? If the analysis was performed with the insulation still on the wire, could that account for some of the low-energy (LE) response, potentially from the Carbon, Chlorine, and Fluorine, etc. if the insulation was indeed Teflon?

?

I¡¯ll have to pull out Geo¡¯s results and see how they compare, although I don¡¯t recall that he quantified percentages of each of the metals in the mix. Again, many thanks for your analysis of the wire.?

?

Have a great weekend.

?

73s? --? Ken

?

?

From: Dude
Sent: Friday, February 7, 2020 04:05 PM
To: [email protected]
Subject: [XRF] Mystery wire

?

Ken,

Attached are the results of your Mystery Wire.

The data were taken with an Olympus DP4050 pXRF. As I don¡¯t have an ALLOY calibration for this gun I used the Mining Mode to get the percentage of elements present

The set up used a 50kV 11 uA x-ray beam with a live time of 30 secs obtaining account rate of 9505 cps from the SDD detector.

?

Mn was identified but it has an interference from Cr and Fe, probably there but not confirmed

V is not present

Co is not present Fe interference

Ti is not there, interference from Si escape peak

Zn is not confirmed

LE are the Low Energy elements

?

?

?

?

?


The case of the missing elements

 

I recently picked up a specimen of melanocerite-Ce at the mineral show and did a scan without Am241 by depending on the internal radiation.? I was not surprised to see Th La Ce and Nd in the blue plot.

Then I wanted to see if I got the same results using Am241 as an additional exciter.? So I carefully moved the specimen just far enough from the detector to allow me to slide in the Am241 jig.? I was careful to keep the same spot on the rock even though it looked quite uniform.

Curiously I am now getting additional strong Fe and Y peaks in the red plot.? What could account for this?? I am not surprised that the La Ce Nd are lit up as they are.

I have also attached both .mca files in case one wishes to analyze them.? However, I am not really asking for that.? I would just like to understand why the Am241 lights up the Fe and Y so strongly compared to the internal radiation.? It is kind of a pain to weed through the Am241 background peaks but it appears that I may not get a complete picture with the internal radiation alone.

Charles


Re: Mystery wire

 

Thanks for that Dudley, when I get my own SDD going I can compare- it may be a while tho.

Geo

----- Original Message -----
From: Dude <dfemer@...>
To: [email protected]
Sent: Fri, 07 Feb 2020 15:59:56 -0500 (EST)
Subject: [XRF] Mystery wire

Ken,

Attached are the results of your Mystery Wire.

The data were taken with an Olympus DP4050 pXRF. As I don¡¯t
have an ALLOY calibration for this gun I used the Mining Mode to get the
percentage of elements present

The set up used a 50kV 11 uA x-ray beam with a live time of
30 secs obtaining account rate of 9505 cps from the SDD detector.

?

Mn was identified but it has an interference from Cr and Fe,
probably there but not confirmed

V is not present

Co is not present Fe interference

Ti is not there, interference from Si escape peak

Zn is not confirmed

LE are the Low Energy elements

?

?

?

?





Mystery wire

 

¿ªÔÆÌåÓý

Ken,

Attached are the results of your Mystery Wire.

The data were taken with an Olympus DP4050 pXRF. As I don¡¯t have an ALLOY calibration for this gun I used the Mining Mode to get the percentage of elements present

The set up used a 50kV 11 uA x-ray beam with a live time of 30 secs obtaining account rate of 9505 cps from the SDD detector.

?

Mn was identified but it has an interference from Cr and Fe, probably there but not confirmed

V is not present

Co is not present Fe interference

Ti is not there, interference from Si escape peak

Zn is not confirmed

LE are the Low Energy elements

?Mystery wire assay.bmp

?

mystery wire.bmp

?

?


Re: 2020 Portable X-Ray Tube XRF Outfit.Sensor Port Upgrades

 

Sounds good.? These two big side holes are from the two inch PMT units.? ?I also have the thick shield tubes from them.? It's heavy like lead but adds you can see looks like copper/bronze, probably a passivation layer??

I'm in San Antonio so LN2 shouldn't be an issue,? but haven't checked prices yet.?


On Fri, Feb 7, 2020, 10:52 AM <GEOelectronics@...> wrote:
"I think the shield from a liquid scint counter I parted out will work for this with some work. "

No doubt, we did learn the lining (passivation layer) is nearly as important as the safety shielding.

Nothing wrong w LN2 if you lie in/near a city and can get it delivered. I hear Carl W. transports it in the trunk of his car (not recommended).

Great on the Kevex, now look at the THERMO power supplies, some as low as $30, and will handle an SDD. If it's a Si-PIN, then the DP-5 should work with it. My SDD is another brand, and won't work with DP-5 (yet- it should when the DP5 is set up as a stand alone MCA). SDD uses high voltage of opposite polarity from Si-PIN.

DP-5 stacks marked Si-PIN will drive the SI-PIN with its own preamp nothing else needed. For use with other detectors, the PC5 isn't used, just the DP5 which must be configured as a stand alone MCA, but there is no preamp in in the DP5.

Geo

----- Original Message -----
From: Nick Andrews <nickjandrews@...>
To: [email protected]
Sent: Fri, 07 Feb 2020 10:48:29 -0500 (EST)
Subject: Re: [XRF] 2020 Portable X-Ray Tube XRF Outfit.Sensor Port Upgrades

Cool.? ?I think the shield from a liquid scint counter I parted out will work for this with some work. While I can foresee moving to a more modern (no LN2 needed) detector than the HPGe, I did buy the thing so I might as well play with it. If it works.?

I did pick up that Kevex fan cooled detector for $200?and that should be interesting to try. It arrives Monday.?

On the Amptek board stacks.? ?What's the consensus on value??

On Thu, Feb 6, 2020, 1:55 PM <GEOelectronics@...> wrote:
The inner aluminum layer is held in place inside the lead layer by a threaded pipe. The length of that pipe limits the insertion depth when that hole is used for the sensor port:

Looking_down-small-note.jpg

New machined threaded coupling pipe allows for greater insertion and can be trimmed another 1/4" if needed. The diameter was also increased to allow other probes to be inserted into this port.

External-View-New-Coupler-small-notes.jpg

However, the design allows the Exciter Port to be located anywhere on the floor of the chamber:
Beam-Port-Adjustable-Location-small-notes.jpg

When the beam leaves the Exciter Module, its is round and 10mm diameter:
10mm-at-tube-port-opening-small.jpg


?By the time it leaves the floor of the chamber it is slightly larger and at 3" from the Exciter Module brass fitting it is 25mm diameter.

Beam diameter test jig (Shop-Built Industrial Spinthariscope):
Beam_Test_Jig-Pub1-small.jpg










Re: 2020 Portable X-Ray Tube XRF Outfit.Sensor Port Upgrades

 

"I think the shield from a liquid scint counter I parted out will work for this with some work. "

No doubt, we did learn the lining (passivation layer) is nearly as important as the safety shielding.

Nothing wrong w LN2 if you lie in/near a city and can get it delivered. I hear Carl W. transports it in the trunk of his car (not recommended).

Great on the Kevex, now look at the THERMO power supplies, some as low as $30, and will handle an SDD. If it's a Si-PIN, then the DP-5 should work with it. My SDD is another brand, and won't work with DP-5 (yet- it should when the DP5 is set up as a stand alone MCA). SDD uses high voltage of opposite polarity from Si-PIN.

DP-5 stacks marked Si-PIN will drive the SI-PIN with its own preamp nothing else needed. For use with other detectors, the PC5 isn't used, just the DP5 which must be configured as a stand alone MCA, but there is no preamp in in the DP5.

Geo

----- Original Message -----
From: Nick Andrews <nickjandrews@...>
To: [email protected]
Sent: Fri, 07 Feb 2020 10:48:29 -0500 (EST)
Subject: Re: [XRF] 2020 Portable X-Ray Tube XRF Outfit.Sensor Port Upgrades

Cool.? ?I think the shield from a liquid scint counter I parted out will work for this with some work. While I can foresee moving to a more modern (no LN2 needed) detector than the HPGe, I did buy the thing so I might as well play with it. If it works.?

I did pick up that Kevex fan cooled detector for $200?and that should be interesting to try. It arrives Monday.?

On the Amptek board stacks.? ?What's the consensus on value??

On Thu, Feb 6, 2020, 1:55 PM <GEOelectronics@...> wrote:
The inner aluminum layer is held in place inside the lead layer by a threaded pipe. The length of that pipe limits the insertion depth when that hole is used for the sensor port:



New machined threaded coupling pipe allows for greater insertion and can be trimmed another 1/4" if needed. The diameter was also increased to allow other probes to be inserted into this port.



However, the design allows the Exciter Port to be located anywhere on the floor of the chamber:


When the beam leaves the Exciter Module, its is round and 10mm diameter:



?By the time it leaves the floor of the chamber it is slightly larger and at 3" from the Exciter Module brass fitting it is 25mm diameter.

Beam diameter test jig (Shop-Built Industrial Spinthariscope):











Re: 2020 Portable X-Ray Tube XRF Outfit.Sensor Port Upgrades

 

Cool.? ?I think the shield from a liquid scint counter I parted out will work for this with some work. While I can foresee moving to a more modern (no LN2 needed) detector than the HPGe, I did buy the thing so I might as well play with it. If it works.?

I did pick up that Kevex fan cooled detector for $200?and that should be interesting to try. It arrives Monday.?

On the Amptek board stacks.? ?What's the consensus on value??

On Thu, Feb 6, 2020, 1:55 PM <GEOelectronics@...> wrote:
The inner aluminum layer is held in place inside the lead layer by a threaded pipe. The length of that pipe limits the insertion depth when that hole is used for the sensor port:

Looking_down-small-note.jpg

New machined threaded coupling pipe allows for greater insertion and can be trimmed another 1/4" if needed. The diameter was also increased to allow other probes to be inserted into this port.

External-View-New-Coupler-small-notes.jpg

However, the design allows the Exciter Port to be located anywhere on the floor of the chamber:
Beam-Port-Adjustable-Location-small-notes.jpg

When the beam leaves the Exciter Module, its is round and 10mm diameter:
10mm-at-tube-port-opening-small.jpg


?By the time it leaves the floor of the chamber it is slightly larger and at 3" from the Exciter Module brass fitting it is 25mm diameter.

Beam diameter test jig (Shop-Built Industrial Spinthariscope):
Beam_Test_Jig-Pub1-small.jpg







Re: 2020 Portable X-Ray Tube XRF Outfit.Sensor Port Upgrades

 

¿ªÔÆÌåÓý

Looks good let¡¯s see some data

Dud

?

From: [email protected] [mailto:[email protected]] On Behalf Of GEOelectronics@...
Sent: Thursday, February 6, 2020 11:56 AM
To: [email protected]
Subject: Re: [XRF] 2020 Portable X-Ray Tube XRF Outfit.Sensor Port Upgrades

?

The inner aluminum layer is held in place inside the lead layer by a threaded pipe. The length of that pipe limits the insertion depth when that hole is used for the sensor port:

Looking_down-small-note.jpg

New machined threaded coupling pipe allows for greater insertion and can be trimmed another 1/4" if needed. The diameter was also increased to allow other probes to be inserted into this port.

External-View-New-Coupler-small-notes.jpg

However, the design allows the Exciter Port to be located anywhere on the floor of the chamber:
Beam-Port-Adjustable-Location-small-notes.jpg

When the beam leaves the Exciter Module, its is round and 10mm diameter:
10mm-at-tube-port-opening-small.jpg


?By the time it leaves the floor of the chamber it is slightly larger and at 3" from the Exciter Module brass fitting it is 25mm diameter.

Beam diameter test jig (Shop-Built Industrial Spinthariscope):
Beam_Test_Jig-Pub1-small.jpg






Re: 2020 Portable X-Ray Tube XRF Outfit.Sensor Port Upgrades

 
Edited

The inner aluminum layer is held in place inside the lead layer by a threaded pipe. The length of that pipe limits the insertion depth when that hole is used for the sensor port:

Looking_down-small-note.jpg

New machined threaded coupling pipe allows for greater insertion and can be trimmed another 1/4" if needed. The diameter was also increased to allow other probes to be inserted into this port.

External-View-New-Coupler-small-notes.jpg

However, the design allows the Exciter Port to be located anywhere on the floor of the chamber:
Beam-Port-Adjustable-Location-small-notes.jpg

When the beam leaves the Exciter Module, its is round and 10mm diameter:
10mm-at-tube-port-opening-small.jpg


?By the time it leaves the floor of the chamber it is slightly larger and at 2.25" from the Exciter Module brass fitting it is 25mm diameter.

Beam diameter test jig (Shop-Built Industrial Spinthariscope):
Beam_Test_Jig-Pub1-small.jpg





Re: 2020 Portable X-Ray Tube XRF Outfit.- Fine Tuning the Chamber Lid

 


Aluminum scrap from the shop, for some unknown past project:

Scrap_Bucket_Al_find-small.jpg

Top view
Scrap_Bucket_Al_find_TOP-small.jpg


The thickness is fine, works as expected to eliminate unwanted XRF, but the diameter is way to large.
Initial testing shows scattering from exposed edges, so new final version of the lid will have all the internal aluminum XRF passivation layer shielded on the outside by 1/4" of Pb. The aluminum interior is to provide clean spectrum scans, reducing or eliminating Pb XRF from the personnel safety shielding.

Lathe rough cut to dimensions for fit testing:

Scrap_Bucket_Al_find_Rough-cut-test-fit-small.jpg

As expected, the plug section is 1mm oversized, and the retainer lip on top is way to thick. More measurements were taken, then back to the shop for the final machining cuts.

More test fitting, then back to lathe for touch-up and polishing.
?The finished piece:

Final_Machining-small.jpg

and the fit test:
First-Aluminum-LId-Test-small.jpg

Now all that's left to do is craft the final layer of the top from 1/4" Pb and a handle.

Geo










Re: Stainless Steel XRF

 

¿ªÔÆÌåÓý

Nice tight beam good for crystal work

Dud

?

From: [email protected] [mailto:[email protected]] On Behalf Of GEOelectronics@...
Sent: Wednesday, February 5, 2020 6:59 PM
To: [email protected]
Subject: Re: [XRF] Stainless Steel XRF

?

Dud, I'll look at the link tomorrow, thanks. Meantime, per your request,? my? module gives a 10mm round beam at the surface of the brass port that is part of the module. 3 inches away the spot it still circular and has spread to 25mm.

?

?

Geo

?

----- Original Message -----
From: Dude <dfemer@...>
To: [email protected]
Sent: Wed, 05 Feb 2020 19:19:29 -0500 (EST)
Subject: Re: [XRF] Stainless Steel XRF

?



Geo,

You use XRF filters to remove tube artifacts, Brem, ?and
fine tune low energy response. The filters are used in conjunction with the
tube voltage and the filter material is tailored for the absorption edge for
the analytes you are interested in. ?Putting a 4.8mm block of Al will reduce
the 50 kV intensity by a factor 0.6, at 30 kV the beam is attenuated down to
0.2. At 20 kV its gone at 0.012.? The only thing ¾±³Ù¡¯²õ doing is killing your
beam intensity (count time) and shape resulting in excitation only by the highest
energies.? For looking at a broad band range of elements don¡¯t use a filter.
When looking to reduce noise select a filter and beam kV tailored to that
material of interest.? Filters are used to maximize the sensitivity to
different groups of elements.? They are not thick, they are thin, a couple of um
to mils, and are single or layered composite of Al , Ti, Cu, C, Pd, In, etc.?
Filters are used to tailor the excitation ?energy range. The filter is designed
to absorb x-rays with energies immediately above the absorption edge of the filter
material.? Lower energy x-rays and the much higher energy x-rays will make it ?through.?
This reduces the Brem just above the absorption edge.? For an Al filter (not a 4.8
mm! but a mil or so), the absorption edge of Al is 1.56 keV so the Ka of the lower
Z elements Mg (1.2 keV) and Na (1.0 keV) will be absorbed by filter while the
region just above that will have a very low background allowing better
sensitivity to elements from Si (1.7 keV) to Ca (3.7) due to the lower brem.
Above that the higher energies are unaffected.? The filter is used to help pull
up the sensitivity in energy ranges defined by the element¡¯s Z.

The ?only thing you are doing with the Al block is
reducing the total excitation flux rate, lowering the low energy flux rate and killing
the low energy element sensitivity while increasing the count time.

Take the Al filter out. ?If you¡¯re getting Sum peaks then
attenuate the beam but use the beam current first.

?

Dud

?

?

From: [email protected]
[mailto:[email protected]] On Behalf Of GEOelectronics@...
Sent: Wednesday, February 5, 2020 10:24 AM
To: [email protected]
Subject: Re: [XRF] Stainless Steel XRF

?

"
Geo,Why is
there ?4.8mm of Al attenuator in the beam path? "

?

Two
reasons, the first is to selectively block the lowest energies from the tube
(to help the detector see less noise at the area of interest below 15keV, then
the top layer is to control beam spreading and provide a clear path (1mm hole)
for the beam I want to strike the target.

?

Some
of the same thought went into the Am_X8 to reduce the 13.9 +17.74 but leave the
59.5 pretty much unaffected flux-wise.

?

Geo

?

-----
Original Message -----

From: Dude <dfemer@...>

To: [email protected]

Sent: Wed, 05 Feb 2020 00:25:05 -0500 (EST)

Subject: Re: [XRF] Stainless Steel XRF

?

Geo,

Why is there ?4.8mm of Al attenuator in the beam path?
?What is the

no collimated diameter of the micro-focus beam on the target? Is it well

aligned with the 1mm collimator, if ¾±³Ù¡¯²õ not you¡¯ll have ?a lot of attenuation.?

Collimation usually uses 2 ?separated collimators and sometimes one on the

detector. ?Start simple, no collimator, no attenuators. Look at the beam

diameter at the target plane. Set up the collimation to get the beam diameter

you want at the target plane. ?Then move the Detector close to the target
say

about 1 ¨C 2cm. ?Take a shot using the lowest beam current and look at Dead
Time.

You should be able to count at 40Kcps and have good resolution and count
time.?

Play with the beam current. Play with Detector distance to target.

Your comment ¡°¡­allows the sensor to be considerably close to the center of the

chamber. Not that? it maters very much¡±
?What is considerably

closer? It matters very much. The beam will expose a small circle on the target

that will fluoresce over a solid angle. The flux in is proportional to the flux

out. The percentage of that signal your detector will get is directly dependent

on the distance from the target and that solid angle. Close up its 1/r, at a

distance it falls as 1/r2.? Remember the time distance shielding thing

?Do a sensitivity analysis. What gets the best signal? What
changes

things the most? ¨C where do I optimize my effort? Is it Distance, geometric
angle

between beam and detector, collimation, need for attenuator, what size beam

spot do I want on the target, what s that get me in count time etc.

The problem here is not the difference between a SDD and a SI-PIN
¾±³Ù¡¯²õ

the set up geometry and the shield system limiting you to very poor performance.

Dud

KK7IF

?

?

From:
[email protected]

[mailto:[email protected]] On Behalf Of GEOelectronics@...
Sent: Tuesday, February 4, 2020 7:32 PM
To: [email protected]
Subject: Re: [XRF] Stainless Steel XRF

?

?

Yes,

the exciter flux is very low. Only 20 uA and in the beam port right now are 3

ea. 1.6mm Aluminum plates (solid) and then a 4.75mm plate with a 1mm hole in

the center. I'm please with the clean patterns, even tho slow. This is a

Si-PIN, not an SDD.?

?

I

did machine a new part the other day that connects the Al and Pb?

cylinders together, that allows the sensor to be considerably close to the

center of the chamber. Not that? it maters very much, as the beam can be

positioned anywhere on the floor of the chamber.

?

Also

today, a trip to the garage yielded a good thick aluminum part that can be
turned

on the lather to make the lid-liner,

With
luck, the lid will be finished tomorrow and work on the

permanent collimator started.

Geo

?

?

?


Re: Stainless Steel XRF

 
Edited

Dud, I'll look at the link tomorrow, thanks. Meantime, per your request,? my? module gives a 10mm round beam at the surface of the brass port that is part of the module.2.25 inches away the spot it still circular and has spread to 25mm.
?
?
Geo
?

----- Original Message -----
From: Dude <dfemer@...>
To: [email protected]
Sent: Wed, 05 Feb 2020 19:19:29 -0500 (EST)
Subject: Re: [XRF] Stainless Steel XRF
?
?

Geo,

You use XRF filters to remove tube artifacts, Brem, ?and
fine tune low energy response. The filters are used in conjunction with the
tube voltage and the filter material is tailored for the absorption edge for
the analytes you are interested in. ?Putting a 4.8mm block of Al will reduce
the 50 kV intensity by a factor 0.6, at 30 kV the beam is attenuated down to
0.2. At 20 kV its gone at 0.012.? The only thing ¾±³Ù¡¯²õ doing is killing your
beam intensity (count time) and shape resulting in excitation only by the highest
energies.? For looking at a broad band range of elements don¡¯t use a filter.
When looking to reduce noise select a filter and beam kV tailored to that
material of interest.? Filters are used to maximize the sensitivity to
different groups of elements.? They are not thick, they are thin, a couple of um
to mils, and are single or layered composite of Al , Ti, Cu, C, Pd, In, etc.?
Filters are used to tailor the excitation ?energy range. The filter is designed
to absorb x-rays with energies immediately above the absorption edge of the filter
material.? Lower energy x-rays and the much higher energy x-rays will make it ?through.?
This reduces the Brem just above the absorption edge.? For an Al filter (not a 4.8
mm! but a mil or so), the absorption edge of Al is 1.56 keV so the Ka of the lower
Z elements Mg (1.2 keV) and Na (1.0 keV) will be absorbed by filter while the
region just above that will have a very low background allowing better
sensitivity to elements from Si (1.7 keV) to Ca (3.7) due to the lower brem.
Above that the higher energies are unaffected.? The filter is used to help pull
up the sensitivity in energy ranges defined by the element¡¯s Z.

The ?only thing you are doing with the Al block is
reducing the total excitation flux rate, lowering the low energy flux rate and killing
the low energy element sensitivity while increasing the count time.

Take the Al filter out. ?If you¡¯re getting Sum peaks then
attenuate the beam but use the beam current first.

?

Dud

?

?

From: [email protected]
[mailto:[email protected]] On Behalf Of GEOelectronics@...
Sent: Wednesday, February 5, 2020 10:24 AM
To: [email protected]
Subject: Re: [XRF] Stainless Steel XRF

?

"
Geo,Why is
there ?4.8mm of Al attenuator in the beam path?
"

?

Two
reasons, the first is to selectively block the lowest energies from the tube
(to help the detector see less noise at the area of interest below 15keV, then
the top layer is to control beam spreading and provide a clear path (1mm hole)
for the beam I want to strike the target.

?

Some
of the same thought went into the Am_X8 to reduce the 13.9 +17.74 but leave the
59.5 pretty much unaffected flux-wise.

?

Geo

?

-----
Original Message -----

From: Dude <dfemer@...>

To: [email protected]

Sent: Wed, 05 Feb 2020 00:25:05 -0500 (EST)

Subject: Re: [XRF] Stainless Steel XRF

?

Geo,

Why is there ?4.8mm of Al attenuator in the beam path?
?What is the

no collimated diameter of the micro-focus beam on the target? Is it well

aligned with the 1mm collimator, if ¾±³Ù¡¯²õ not you¡¯ll have ?a lot of attenuation.?

Collimation usually uses 2 ?separated collimators and sometimes one on the

detector. ?Start simple, no collimator, no attenuators. Look at the beam

diameter at the target plane. Set up the collimation to get the beam diameter

you want at the target plane. ?Then move the Detector close to the target
say

about 1 ¨C 2cm. ?Take a shot using the lowest beam current and look at Dead
Time.

You should be able to count at 40Kcps and have good resolution and count
time.?

Play with the beam current. Play with Detector distance to target.

Your comment ¡°¡­allows the sensor to be considerably close to the center of the

chamber. Not that? it maters very much¡±
?What is considerably

closer? It matters very much. The beam will expose a small circle on the target

that will fluoresce over a solid angle. The flux in is proportional to the flux

out. The percentage of that signal your detector will get is directly dependent

on the distance from the target and that solid angle. Close up its 1/r, at a

distance it falls as 1/r2.? Remember the time distance shielding thing

?Do a sensitivity analysis. What gets the best signal? What
changes

things the most? ¨C where do I optimize my effort? Is it Distance, geometric
angle

between beam and detector, collimation, need for attenuator, what size beam

spot do I want on the target, what s that get me in count time etc.

The problem here is not the difference between a SDD and a SI-PIN
¾±³Ù¡¯²õ

the set up geometry and the shield system limiting you to very poor performance.

Dud

KK7IF

?

?

From:
[email protected]

[mailto:[email protected]] On Behalf Of GEOelectronics@...
Sent: Tuesday, February 4, 2020 7:32 PM
To: [email protected]
Subject: Re: [XRF] Stainless Steel XRF

?

?

Yes,

the exciter flux is very low. Only 20 uA and in the beam port right now are 3

ea. 1.6mm Aluminum plates (solid) and then a 4.75mm plate with a 1mm hole in

the center. I'm please with the clean patterns, even tho slow. This is a

Si-PIN, not an SDD.?

?

I

did machine a new part the other day that connects the Al and Pb?

cylinders together, that allows the sensor to be considerably close to the

center of the chamber. Not that? it maters very much, as the beam can be

positioned anywhere on the floor of the chamber.

?

Also

today, a trip to the garage yielded a good thick aluminum part that can be
turned

on the lather to make the lid-liner,

With
luck, the lid will be finished tomorrow and work on the

permanent collimator started.

Geo

?

?
?
?


Re: Stainless Steel XRF

 

¿ªÔÆÌåÓý

Geo,

You use XRF filters to remove tube artifacts, Brem, ?and fine tune low energy response. The filters are used in conjunction with the tube voltage and the filter material is tailored for the absorption edge for the analytes you are interested in. ?Putting a 4.8mm block of Al will reduce the 50 kV intensity by a factor 0.6, at 30 kV the beam is attenuated down to 0.2. At 20 kV its gone at 0.012.? The only thing ¾±³Ù¡¯²õ doing is killing your beam intensity (count time) and shape resulting in excitation only by the highest energies.? For looking at a broad band range of elements don¡¯t use a filter. When looking to reduce noise select a filter and beam kV tailored to that material of interest.? Filters are used to maximize the sensitivity to different groups of elements.? They are not thick, they are thin, a couple of um to mils, and are single or layered composite of Al , Ti, Cu, C, Pd, In, etc.? Filters are used to tailor the excitation ?energy range. The filter is designed to absorb x-rays with energies immediately above the absorption edge of the filter material.? Lower energy x-rays and the much higher energy x-rays will make it ?through.? This reduces the Brem just above the absorption edge.? For an Al filter (not a 4.8 mm! but a mil or so), the absorption edge of Al is 1.56 keV so the Ka of the lower Z elements Mg (1.2 keV) and Na (1.0 keV) will be absorbed by filter while the region just above that will have a very low background allowing better sensitivity to elements from Si (1.7 keV) to Ca (3.7) due to the lower brem. Above that the higher energies are unaffected.? The filter is used to help pull up the sensitivity in energy ranges defined by the element¡¯s Z.

The ?only thing you are doing with the Al block is reducing the total excitation flux rate, lowering the low energy flux rate and killing the low energy element sensitivity while increasing the count time.

Take the Al filter out. ?If you¡¯re getting Sum peaks then attenuate the beam but use the beam current first.

?

Dud

?

?

From: [email protected] [mailto:[email protected]] On Behalf Of GEOelectronics@...
Sent: Wednesday, February 5, 2020 10:24 AM
To: [email protected]
Subject: Re: [XRF] Stainless Steel XRF

?

" Geo,Why is there ?4.8mm of Al attenuator in the beam path? "

?

Two reasons, the first is to selectively block the lowest energies from the tube (to help the detector see less noise at the area of interest below 15keV, then the top layer is to control beam spreading and provide a clear path (1mm hole) for the beam I want to strike the target.

?

Some of the same thought went into the Am_X8 to reduce the 13.9 +17.74 but leave the 59.5 pretty much unaffected flux-wise.

?

Geo

?

----- Original Message -----
From: Dude <dfemer@...>
To: [email protected]
Sent: Wed, 05 Feb 2020 00:25:05 -0500 (EST)
Subject: Re: [XRF] Stainless Steel XRF

?

Geo,

Why is there ?4.8mm of Al attenuator in the beam path? ?What is the
no collimated diameter of the micro-focus beam on the target? Is it well
aligned with the 1mm collimator, if ¾±³Ù¡¯²õ not you¡¯ll have ?a lot of attenuation.?
Collimation usually uses 2 ?separated collimators and sometimes one on the
detector. ?Start simple, no collimator, no attenuators. Look at the beam
diameter at the target plane. Set up the collimation to get the beam diameter
you want at the target plane. ?Then move the Detector close to the target say
about 1 ¨C 2cm. ?Take a shot using the lowest beam current and look at Dead Time.
You should be able to count at 40Kcps and have good resolution and count time.?
Play with the beam current. Play with Detector distance to target.

Your comment ¡°¡­allows the sensor to be considerably close to the center of the
chamber. Not that? it maters very much¡±
?What is considerably
closer? It matters very much. The beam will expose a small circle on the target
that will fluoresce over a solid angle. The flux in is proportional to the flux
out. The percentage of that signal your detector will get is directly dependent
on the distance from the target and that solid angle. Close up its 1/r, at a
distance it falls as 1/r2.? Remember the time distance shielding thing

?Do a sensitivity analysis. What gets the best signal? What changes
things the most? ¨C where do I optimize my effort? Is it Distance, geometric angle
between beam and detector, collimation, need for attenuator, what size beam
spot do I want on the target, what s that get me in count time etc.

The problem here is not the difference between a SDD and a SI-PIN ¾±³Ù¡¯²õ
the set up geometry and the shield system limiting you to very poor performance.

Dud

KK7IF

?

?

From: [email protected]
[mailto:[email protected]] On Behalf Of GEOelectronics@...
Sent: Tuesday, February 4, 2020 7:32 PM
To: [email protected]
Subject: Re: [XRF] Stainless Steel XRF

?

?

Yes,
the exciter flux is very low. Only 20 uA and in the beam port right now are 3
ea. 1.6mm Aluminum plates (solid) and then a 4.75mm plate with a 1mm hole in
the center. I'm please with the clean patterns, even tho slow. This is a
Si-PIN, not an SDD.?

?

I
did machine a new part the other day that connects the Al and Pb?
cylinders together, that allows the sensor to be considerably close to the
center of the chamber. Not that? it maters very much, as the beam can be
positioned anywhere on the floor of the chamber.

?

Also
today, a trip to the garage yielded a good thick aluminum part that can be turned
on the lather to make the lid-liner,

With luck, the lid will be finished tomorrow and work on the
permanent collimator started.

Geo

_._,_._,_


Re: Stainless Steel XRF

 

" Geo,Why is there ?4.8mm of Al attenuator in the beam path? "

Two reasons, the first is to selectively block the lowest energies from the tube (to help the detector see less noise at the area of interest below 15keV, then the top layer is to control beam spreading and provide a clear path (1mm hole) for the beam I want to strike the target.

Some of the same thought went into the Am_X8 to reduce the 13.9 +17.74 but leave the 59.5 pretty much unaffected flux-wise.

Geo

----- Original Message -----
From: Dude <dfemer@...>
To: [email protected]
Sent: Wed, 05 Feb 2020 00:25:05 -0500 (EST)
Subject: Re: [XRF] Stainless Steel XRF

Geo,

Why is there ?4.8mm of Al attenuator in the beam path? ?What is the
no collimated diameter of the micro-focus beam on the target? Is it well
aligned with the 1mm collimator, if ¾±³Ù¡¯²õ not you¡¯ll have ?a lot of attenuation.?
Collimation usually uses 2 ?separated collimators and sometimes one on the
detector. ?Start simple, no collimator, no attenuators. Look at the beam
diameter at the target plane. Set up the collimation to get the beam diameter
you want at the target plane. ?Then move the Detector close to the target say
about 1 ¨C 2cm. ?Take a shot using the lowest beam current and look at Dead Time.
You should be able to count at 40Kcps and have good resolution and count time.?
Play with the beam current. Play with Detector distance to target.

Your comment ¡°¡­allows the sensor to be considerably close to the center of the
chamber. Not that? it maters very much¡±
?What is considerably
closer? It matters very much. The beam will expose a small circle on the target
that will fluoresce over a solid angle. The flux in is proportional to the flux
out. The percentage of that signal your detector will get is directly dependent
on the distance from the target and that solid angle. Close up its 1/r, at a
distance it falls as 1/r2.? Remember the time distance shielding thing

?Do a sensitivity analysis. What gets the best signal? What changes
things the most? ¨C where do I optimize my effort? Is it Distance, geometric angle
between beam and detector, collimation, need for attenuator, what size beam
spot do I want on the target, what s that get me in count time etc.

The problem here is not the difference between a SDD and a SI-PIN ¾±³Ù¡¯²õ
the set up geometry and the shield system limiting you to very poor performance.

Dud

KK7IF

?

?

From: [email protected]
[mailto:[email protected]] On Behalf Of GEOelectronics@...
Sent: Tuesday, February 4, 2020 7:32 PM
To: [email protected]
Subject: Re: [XRF] Stainless Steel XRF

?

?

Yes,
the exciter flux is very low. Only 20 uA and in the beam port right now are 3
ea. 1.6mm Aluminum plates (solid) and then a 4.75mm plate with a 1mm hole in
the center. I'm please with the clean patterns, even tho slow. This is a
Si-PIN, not an SDD.?

?

I
did machine a new part the other day that connects the Al and Pb?
cylinders together, that allows the sensor to be considerably close to the
center of the chamber. Not that? it maters very much, as the beam can be
positioned anywhere on the floor of the chamber.

?

Also
today, a trip to the garage yielded a good thick aluminum part that can be turned
on the lather to make the lid-liner,

With luck, the lid will be finished tomorrow and work on the
permanent collimator started.

Geo

-----
Original Message -----

From: Dude <dfemer@...>

To: [email protected]

Sent: Tue, 04 Feb 2020 19:52:13 -0500 (EST)

Subject: Re: [XRF] Stainless Steel XRF

?



margin: 0.0in;

font-size: 12.0pt;

font-family: "Times New Roman" , serif;

}

a:link, span.MsoHyperlink {

color: blue;

text-decoration: underline;

}

a:visited, span.MsoHyperlinkFollowed {

color: purple;

text-decoration: underline;

}

p {

margin-right: 0.0in;

margin-left: 0.0in;

font-size: 12.0pt;

font-family: "Times New Roman" , serif;

}

span.EmailStyle18 {

font-family: Calibri , sans-serif;

color: black;

}

*.MsoChpDefault {

font-size: 10.0pt;

}

div.Section1 {

page: Section1;

}

/*]]>*/

Geo,

Getting
only 1800 counts/sec ?in 600 sec is saying something is

not quite right with the setup. Not separating the Fe and Ni Kb and Ka is also

not right but may? be related to the low count rate. Before you go any
further

find out why you¡¯re getting the low count rate. This spectra should come in in
?30

sec count. ?I suspect the beam flux is too low (what¡¯s the collimator
diameter

and where is the micro-focus point?) and if you haven¡¯t moved the detector

closer to the target ?that may be the problem as well.? Get rid of
the lead shield

setup and set up the geometry by hand without it (but keeping ?shielding
between

you and it). Play around getting an optimal geometry and beam current. Start by

getting the detector closer to the target.? I¡¯d think running ?a test
with and

without the plastic cap on the low energy end would be interesting test to see

as well. Use a Si target.

Dud

Sent:
Tuesday, February

4, 2020 2:02 PM
To: [email protected]
Subject: [XRF] Stainless Steel XRF

?

Baby steps- a 304 stainless bolt from the shop.



Geo

?

?

?

?

?