"I think the shield from a liquid scint counter I parted out will work for this with some work. " No doubt, we did learn the lining (passivation layer) is nearly as important as the safety shielding. Nothing wrong w LN2 if you lie in/near a city and can get it delivered. I hear Carl W. transports it in the trunk of his car (not recommended). Great on the Kevex, now look at the THERMO power supplies, some as low as $30, and will handle an SDD. If it's a Si-PIN, then the DP-5 should work with it. My SDD is another brand, and won't work with DP-5 (yet- it should when the DP5 is set up as a stand alone MCA). SDD uses high voltage of opposite polarity from Si-PIN. DP-5 stacks marked Si-PIN will drive the SI-PIN with its own preamp nothing else needed. For use with other detectors, the PC5 isn't used, just the DP5 which must be configured as a stand alone MCA, but there is no preamp in in the DP5. Geo ----- Original Message ----- From: Nick Andrews <nickjandrews@...> To: [email protected] Sent: Fri, 07 Feb 2020 10:48:29 -0500 (EST) Subject: Re: [XRF] 2020 Portable X-Ray Tube XRF Outfit.Sensor Port Upgrades Cool.? ?I think the shield from a liquid scint counter I parted out will work for this with some work. While I can foresee moving to a more modern (no LN2 needed) detector than the HPGe, I did buy the thing so I might as well play with it. If it works.? I did pick up that Kevex fan cooled detector for $200?and that should be interesting to try. It arrives Monday.? On the Amptek board stacks.? ?What's the consensus on value?? On Thu, Feb 6, 2020, 1:55 PM <GEOelectronics@...> wrote:
|