Re: 59.5keV spectrum cleanup
Final on the aluminum planchet source. 14400secs.
But the source an positively be identified in 3 minutes.
Geo
By
GEOelectronics@...
·
#1595
·
|
Re: 59.5keV spectrum cleanup
By
GEOelectronics@...
·
#1594
·
|
Re: 59.5keV spectrum cleanup
Hi Charles, close but no Am involved. The exciter is natural U-238, we are seeing the decay of the daughter Pa-234m (also known as UX2). We are essentially seeing only the birth of the U-234daughter
By
GEOelectronics@...
·
#1593
·
|
Re: 59.5keV spectrum cleanup
Try a longer PT and threshold setting see if the noise gets better. Also take off the plastic cap
I don¡¯t see any Ar yet.
Dud
Sent: Friday, October 30, 2020 5:37 PM
To: [email protected]
Subject: Re:
By
Dude
·
#1591
·
|
Re: 59.5keV spectrum cleanup
I am not sure exactly what I am looking at here but I like the sharpness of
the peaks and the low noise. Is this a scan with an Am241 exciter? See
attached.
Charles
By
Charles David Young
·
#1590
·
|
Re: 59.5keV spectrum cleanup
25.6 like last run, but auto threshold adjust applied. 8500s update .mca below. Didn't help
"noise" at low end, did noticeably improve definition of peaks above 3 keV.
3.3 yes it is Ka1 of
By
GEOelectronics@...
·
#1589
·
|
Re: 59.5keV spectrum cleanup
That run ended, reset the auto Threshold control, restarted, will report back over next 4 hours, already looks better tho.
Geo
To: [email protected]
Sent: Friday, October 30, 2020 4:58:38 PM
By
GEOelectronics@...
·
#1588
·
|
Re: 59.5keV spectrum cleanup
Click on LINEAR view, looks pretty good to me, the L X-rays are the subject, but you are right I didn't do the threshold routine. Do you think 25.6 is too slow? I'm afraid if something with high count
By
GEOelectronics@...
·
#1587
·
|
Re: 59.5keV spectrum cleanup
It looks like the low energy electronic noise is a problem below 1.8 keV. You¡¯ll need to try resetting the threshold filters or try a higher PT as noise counts are 10 times your signal
Dud
By
Dude
·
#1586
·
|
Re: 59.5keV spectrum cleanup
This is real time an still counting but it accumulated 11,250 secs now. .mca below.
Geo
By
GEOelectronics@...
·
#1585
·
|
Re: 59.5keV spectrum cleanup
Looks like Al at 1.4 keV and P at 2.01 but these need a MUCH longer count. K at 3.3. noise at 4.5? Cr at 5.42 and U Ll at 11.64 and Zn 8.64 and 9.57 keV
Needs way more counts before I put my
By
Dude
·
#1584
·
|
Re: 59.5keV spectrum cleanup
Try this one when you can Dud. It's? a stronger sample on an aluminum planchet this time.
Calibration spot on for U La1 and Lg1
Also was set up to try 25.6 peaking.
By
GEOelectronics@...
·
#1583
·
|
Re: 59.5keV spectrum cleanup
Got it. Thanks.
Geo
To: [email protected]
Sent: Friday, October 30, 2020 1:02:37 PM
Subject: Re: [XRF] 59.5keV spectrum cleanup
It would come off of the ROI. The FWHM will change with energy so
By
GEOelectronics@...
·
#1582
·
|
Re: 59.5keV spectrum cleanup
It would come off of the ROI. The FWHM will change with energy so its specific to a peak.
In DPPmca set an ROI on the peaks using the ^ peak like thing on the tool bar. Use the cursor and left click
By
Dude
·
#1581
·
|
Re: FW: [XRF] 59.5keV spectrum cleanup
Charles,
I don¡¯t see much difference either, Maybe just a very slightly lower count ratio of the 2 low energy peaks of the covered run.
Your MCA files have no calibrates in them, that really has
By
Dude
·
#1580
·
|
Re: 59.5keV spectrum cleanup
' You¡¯ll note that the 19.2 peaking time has a much better 13.6 keV FWHM resolution than the 4.8 (0.192 vs .396 FWHM)."
Dud,THANKS for that update and analysis>
is there a report log to
By
GEOelectronics@...
·
#1579
·
|
Re: Chinese water filter analysis
"
Test everything!
Yes
I am gonna scan my x-ray exciter tube
I have a X ray generating tube with unknown anode material
Sort of like biting the hand that feeds you
Taray"
Easily
By
GEOelectronics@...
·
#1578
·
|
Re: FW: [XRF] 59.5keV spectrum cleanup
Geo,
The 4.8 usec doesn¡¯t have the filter fast thresholds set correctly as a result it has you seeing 3135 valid counts with a noise count of 128,980.While The 19.2 PT has 4534 valid counts with a
By
Dude
·
#1577
·
|
Re: FW: [XRF] 59.5keV spectrum cleanup
From what I know about XRD it requires careful preparation of the sample.
I know grinding is involved. Interpreting the results is also not as
straightforward as XRF.
For now, the XRF is working
By
Charles David Young
·
#1576
·
|
Re: Chinese water filter analysis
Geo
The you tubers are overdoing it to score brownie points?
Should do it for less than a min?
Too long saturates iron residue diluting any other heavy metal present provided it is coming from
By
taray singh <sukhjez@...>
·
#1575
·
|