¿ªÔÆÌåÓý

ctrl + shift + ? for shortcuts
© 2025 Groups.io
Re: 59.5keV spectrum cleanup
Final on the aluminum planchet source. 14400secs. But the source an positively be identified in 3 minutes. Geo
By GEOelectronics@... · #1595 ·
Re: 59.5keV spectrum cleanup
By GEOelectronics@... · #1594 ·
Re: 59.5keV spectrum cleanup
Hi Charles, close but no Am involved. The exciter is natural U-238, we are seeing the decay of the daughter Pa-234m (also known as UX2). We are essentially seeing only the birth of the U-234daughter
By GEOelectronics@... · #1593 ·
Re: 59.5keV spectrum cleanup
Try a longer PT and threshold setting see if the noise gets better. Also take off the plastic cap I don¡¯t see any Ar yet. Dud Sent: Friday, October 30, 2020 5:37 PM To: [email protected] Subject: Re:
By Dude · #1591 ·
Re: 59.5keV spectrum cleanup
I am not sure exactly what I am looking at here but I like the sharpness of the peaks and the low noise. Is this a scan with an Am241 exciter? See attached. Charles
By Charles David Young · #1590 ·
Re: 59.5keV spectrum cleanup
25.6 like last run, but auto threshold adjust applied. 8500s update .mca below. Didn't help "noise" at low end, did noticeably improve definition of peaks above 3 keV. 3.3 yes it is Ka1 of
By GEOelectronics@... · #1589 ·
Re: 59.5keV spectrum cleanup
That run ended, reset the auto Threshold control, restarted, will report back over next 4 hours, already looks better tho. Geo To: [email protected] Sent: Friday, October 30, 2020 4:58:38 PM
By GEOelectronics@... · #1588 ·
Re: 59.5keV spectrum cleanup
Click on LINEAR view, looks pretty good to me, the L X-rays are the subject, but you are right I didn't do the threshold routine. Do you think 25.6 is too slow? I'm afraid if something with high count
By GEOelectronics@... · #1587 ·
Re: 59.5keV spectrum cleanup
It looks like the low energy electronic noise is a problem below 1.8 keV. You¡¯ll need to try resetting the threshold filters or try a higher PT as noise counts are 10 times your signal Dud
By Dude · #1586 ·
Re: 59.5keV spectrum cleanup
This is real time an still counting but it accumulated 11,250 secs now. .mca below. Geo
By GEOelectronics@... · #1585 ·
Re: 59.5keV spectrum cleanup
Looks like Al at 1.4 keV and P at 2.01 but these need a MUCH longer count. K at 3.3. noise at 4.5? Cr at 5.42 and U Ll at 11.64 and Zn 8.64 and 9.57 keV Needs way more counts before I put my
By Dude · #1584 ·
Re: 59.5keV spectrum cleanup
Try this one when you can Dud. It's? a stronger sample on an aluminum planchet this time. Calibration spot on for U La1 and Lg1 Also was set up to try 25.6 peaking.
By GEOelectronics@... · #1583 ·
Re: 59.5keV spectrum cleanup
Got it. Thanks. Geo To: [email protected] Sent: Friday, October 30, 2020 1:02:37 PM Subject: Re: [XRF] 59.5keV spectrum cleanup It would come off of the ROI. The FWHM will change with energy so
By GEOelectronics@... · #1582 ·
Re: 59.5keV spectrum cleanup
It would come off of the ROI. The FWHM will change with energy so its specific to a peak. In DPPmca set an ROI on the peaks using the ^ peak like thing on the tool bar. Use the cursor and left click
By Dude · #1581 ·
Re: FW: [XRF] 59.5keV spectrum cleanup
Charles, I don¡¯t see much difference either, Maybe just a very slightly lower count ratio of the 2 low energy peaks of the covered run. Your MCA files have no calibrates in them, that really has
By Dude · #1580 ·
Re: 59.5keV spectrum cleanup
' You¡¯ll note that the 19.2 peaking time has a much better 13.6 keV FWHM resolution than the 4.8 (0.192 vs .396 FWHM)." Dud,THANKS for that update and analysis> is there a report log to
By GEOelectronics@... · #1579 ·
Re: Chinese water filter analysis
" Test everything! Yes I am gonna scan my x-ray exciter tube I have a X ray generating tube with unknown anode material Sort of like biting the hand that feeds you Taray" Easily
By GEOelectronics@... · #1578 ·
Re: FW: [XRF] 59.5keV spectrum cleanup
Geo, The 4.8 usec doesn¡¯t have the filter fast thresholds set correctly as a result it has you seeing 3135 valid counts with a noise count of 128,980.While The 19.2 PT has 4534 valid counts with a
By Dude · #1577 ·
Re: FW: [XRF] 59.5keV spectrum cleanup
From what I know about XRD it requires careful preparation of the sample. I know grinding is involved. Interpreting the results is also not as straightforward as XRF. For now, the XRF is working
By Charles David Young · #1576 ·
Re: Chinese water filter analysis
Geo The you tubers are overdoing it to score brownie points? Should do it for less than a min? Too long saturates iron residue diluting any other heavy metal present provided it is coming from
By taray singh <sukhjez@...> · #1575 ·