Good. No problems, I enjoy this 100%.
It's good to have a collleague.
FIRST- Calibration looks fine.
Second- would you like to increase the gain and move the 50.54 peak off to near the right side? This allows lower peaks to spread out better. My top range is ~ 62.5
George?
toggle quoted message
Show quoted text
----- Original Message -----
From: taray singh via groups.io <sukhjez@...>
To:
[email protected]Sent: Tue, 03 Nov 2020 05:52:18 -0500 (EST)
Subject: Re: [XRF] Chinese water filter analysis
Geo
Here is the Am 241 spectrum with the same settings.
Sorry for the delay.
As for scale error above I misquoted being way from my pc,
On starting Dppmca? ,calibrating? and connecting to si pin,the scale box upper? range is about 700 with spectrum about 80.
Not the other way round
That seems odd.
But on actually starting acquisition, everything normalizes.
Like I told you before I am unable to save calibration despite following the usual way
Always have to? recalibrate? playing with lifedata files on starting
Getting used to the routine
Not an issue.
Anyway let me know what you think about the RO sample
Thanks
Taray