TekWiki is down, "Login required" ? 2
Is anyone else having issues with TekWiki today? I attempted to look up an instrument, but received a message stating that a login is required. I don¡¯t have an account, so I can¡¯t access anything past that.
Started by Nick @ · Most recent @
7844 intensified sweep
Team, I have over the past few months acquired 5 7844 scopes, none of which worked. Good thing it was just the cost of picking them up. I have repaired 3 of them so far, one of them still has the power supply on the bench, one was the parts source. On the latest project a R7844, after replacing caps in the power supply, replacing beam 1 vertical amp twice, original was very noisy, first replacement gain was very compressed on the positive side. Working very well now, calibration done. Discovered one last problem. Intensified sweep using a 7B92A. Beam 1 works perfectly, beam 2 the delayed intensified sweep is very dim . Intensity control does control it but at max you can just discern on the screen. Reducing the separation until it corresponds with the main trace it works perfectly. Now for the weird part if I use the older 7B92, intensified Alt trace, it works perfectly Thoughts? Mark W7HPW Real Radios Glow in the Dark
Started by W7HPW Radio @
Manual for 1424 X-Y Monitor 3
Anyone have a manual for the 1424? Wiki Down? Sam
Started by Sam Reaves @ · Most recent @
2465B - Replaced RIFAs and now won't POST 7
2465B s/n 57xxx When I first powered it up, the lights would all come on for a split second, then they would go out, then it would cycle through the self-tests, failing at TEST 05, so I cleaned up the A5 corrosion and replaced a few components there. I also replaced the RIFAs on the power supply. Following that, when I turn it on, the lights all come on, but they don't go out. All J119 points are well within spec. I am a little confused at times when following the service manual, for example the troubleshooting procedure said TP201 should be +10v, but the schematic seems to indicate it should be +5v, which is what I'm getting. I'm not sure how to proceed at this point. I'd like to go as far as possible before I am forced to buy a new oscilloscope to repair this one. Thanks
Started by Bobert29 @ · Most recent @
465 Dead HV Multiplier? 4
My lovely 465 died immediately after I finished some cosmetic repairs. Low voltage test points all look good. No beam, except when I disconnect the ground lead of the HV multiplier as suggested by VintageTek Museum on youtube (https://www.youtube.com/watch?v=hf1qhx3JNRY). (I also hear some crackling sounds with the multiplier disconnected, so I haven't run it in this state for more than a few seconds.) Could the problem be anything else besides the multiplier? I have a digital scope but no high voltage probe. To measure high voltage DC I have been using a 10MOhm resistor and analog ammeter, but of course that's a significant load. If the problem is the multiplier - part number 152-0552-00 - does anyone know if they're still available? I see that a couple members have built replacements, and I would really appreciate a schematic and/or component values. The location of the multiplier within the scope, and the requirement of potting, will make it difficult to troubleshoot a DIY replacement and I have little high voltage experience. So I'd like to get it in 1 shot.
Started by Gabe @ · Most recent @
The Minisert is re-born - at least in concept 2
<https://www.mill-max.com/products/new/zero-profile-press-fit-receptacles> Cheers, David
Started by David C. Partridge @ · Most recent @
2465b graticule lamp question 2
Found several at Mouser that I don't know which to order The specs look the same and Mouser couldn't tell me the difference CM7152 CM7153AS15 The first one is 30% cheaper.. Thoughts?
Started by KL7AM @ · Most recent @
TDS 3054C plays well with Firefox 129.0.2 on Mac Silicon MacBook Pro 6
I recently acquired a TDS3054C that appears to be functional in all regards, except that I've had trouble accessing the built in web-server with modern browsers on my Windows 11 laptop or my Apple silicon (M3 Max) MacBook Pro. I've updated the firmware on the 'scope to v 4.31, the most recent available on the Tektronix website. On the Windows 11 laptop (an i7 Lenovo ideapad) recent versions of the Microsoft 'Edge' browser and the Google 'Chrome' browser simply wouldn't render the display windows served up by 'scope. Same for current Safari or Chrome browsers on my Apple silicon MacBook Pro. As a sanity check, I tried viewing the 'scope pages with Internet Explorer and Chrome browsers running on a virtual Windows XP environment (hosted with VMWare Fusion on an i7 iMac). With that setup, I could view static screen-shots of the 'scope display, and the 'scope status window. Also, I could download and upload waveform data. However, I found that the tek.com 'cgi-bin/connect' service doesn't appear to be available. After confirming that the 'scope was actually serving screen-shots and waveform data, I started looking for other 'legacy' browsers that would let me access the 'scope from the Windows 11 laptop. I found that the mypal browser (https://www.mypal-browser.org) did the trick. For accessing the 'scope from my Apple silicon MacBook Pro, the Firefox browser v129.0.2 does a great job. Now, if there was just some way to download the cgi-bin/connect scripts from www.tek.com, or the tek 'escope' site was working, I'd be a happy camper. Mark
Started by Mark Davis @ · Most recent @
7L5 knob disassembly, and sourcing replacement phototransistor array 2
There has been some discussion over the years as to how to disassemble the knobs for repair/cleaning. I got mine fully apart (thanks to the photos others uploaded when they got stuck) and uploaded photos. For knobs designed like mine, the black cap pops off and there is a screw underneath. Until the screw is removed, the smaller knob cannot be separated from the larger. The clipped off leads on the LED package allow testing each LED individually if desired; mine have a Vf of ~1.5v. Alternately, 7.5v across the array results in them just starting to glow. The center pin on the left of the array (as oriented in close-up) is common. It seems the transistor on the lower left pin in the array is bad, as it doesn't respond to light; the others do. Does anyone know where to find a 1x5 phototransistor array with spacing of ~1.175mm c-c? The schematics don't show much beyond the fact that the array is biased with +/-15V, and nothing shown in the opto section of the 1982 common design parts catalog looks even remotely close. Regards Mark
Started by mark999 @ · Most recent @
Added photo album Tek 7L5 resolution & frequency knob disassembly #photo-notice
mark999 <mpictor@...> added the photo album Tek 7L5 resolution & frequency knob disassembly ( /g/TekScopes/album?id=301996 ) : 7L5 knob encoder disassembly
Started by Group Notification @
Skirt lettering for scope knobs 15
Greetings! I need to re-letter the skirts for some of my scope knobs. Has anyone done that here? I am thinking that black vinyl reversed letters on transfer stock would work. I found a few sites that provide that service, but I am not sure if they can deal with the small size that would be required. Not sure if a design tool is available online. Other suggestions are welcomed! Thanks, Barry
Started by Barry Breaux @ · Most recent @
SC502, SC503, SC504 Broken Knob Shaft
Greetings from Louisiana! I have had several damaged shipments of the SC502, SC503, SC504 scopes. The Slope / Level switch-pot is damaged along with the knob. In a similar way the Volts/Div and Time/Div knobs were broken. In one case the shafts on all 3 knobs were broken. I decided to have shafts built in plastic and 316L Stainless steel. I took the original knob with the printed skirts and drilled out the original shaft and replace with the custom shafts I designed and had manufactured. The plastic shafts are as good as the original, and the 316L stainless steel shafts are far better than the original. I am also working on having the skirts manufactured out of 1/16" acrylic and printed with UV Ink. So now for the question(s)... 1) Does anyone have any SC502, SC503, SC504 knobs with broken shafts that need to be repaired. I can provide the shafts or repair your broken shaft. 2) Does anyone need new skirts for SC502, SC503, SC504. So far I have restored several of these scopes (more than 10). Initially I got plenty of help / suggestions from the group. Many thanks for that. Regards, Barry
Started by Barry Breaux @
Tektronix 547 Lack of Sweep or Intensity Control 13
My ancient Tektronix 547 oscilloscope with 1A1 plugin was working fine until a few weeks ago when the display suddenly went intermittent and bright, and an arcing sound came from the area of the crt high voltage circuit. No arcing could be seen anywhere, but the sound definitely came from the vicinity of the high voltage assembly. At this point, the Intensity control had no effect, and the intermittent sweeps were very bright. The sweeps only occurred when the arcing sounds were heard. Otherwise, a bright spot stayed on the left side of the display. The focus control also had only a slight effect. Suspecting the flyback transformer T801, I sourced an assembly from another 547 that included the transformer and the voltage multiplier comprised of the 5642 diodes. Symptoms remained the same, unfortunately. Replacing one of the regulator tubes in the 350V supply (6CW5) with a brand new one strangely made the arcing go away, but still no display. In fact, no more intermittent sweeps, only the ultra bright spot at the left hand side of the display. Currently, I have no way of measuring the high voltages from this assembly (10kV), so I¡¯m not sure if they¡¯re within spec. I also changed out the four 0.01uF caps in the Intensity circuit in case they were arcing, but no improvement. Scoping the output of the B Sweep Generator section (at tube V91), there was no sign of a sweep waveform (consistent with the stationary bright dot on the left side of the display). I¡¯ve tested every tube in the scope, except for the 6080s (and checked all transistors for opens or shorts). None gave the impression that they would be show stoppers. I also scoped all B+ and B- rails. All were rock steady except for the +350V line, which had some ripple. Even after changing out the 6CW5, the ripple was still there. It¡¯s obviously generating high voltage as evidenced by the bright spot on the crt. I can¡¯t understand the relationship between the inability to control Intensity and Focus and the lack of a sweep signal. It doesn¡¯t matter whether I¡¯m in A, B or ALT timebases. Could the crt have an internal fault? Any advice or best guesses as to where I should look next would be much appreciated. Thanks in advance.
Started by Gordon Wood @ · Most recent @
453 scope, MOD 703K and MOD 703H 6
have a trio of Tek 453's. Two are MOD 703K and the other is MOD 703H. I have had one of the 703K's on my bench for many years. Of late, it has developed trigger issues. I can only get a stable waveform if feed it a square wave. It will not trigger on a sine wave. The other 703K triggers fine, but has a few other issues, chief among them it is missing the center control knob for the sweep control. Not just the knob, but the center shaft. Not clear what is actually missing internally since that control is fairly well buried (as is everything else on these). The 703H has many other issues, stability wise, so it may become a donor for missing knobs/controls, feet etc. Thoughts on the trigger issue?
Started by Ray M @ · Most recent @
TM5006 Power Supply broken¡­ 3
Hi Group Members! Got a broken TM5006 Mainframe and found a few damaged parts showing a short on my Multimeter. Here the original part #, and now I¡¯m looking for x-Types for those parts. I¡¯d be nice if someone could give me a hint or confirm about the parts I found are fitting! 151-0632-00 some people use MJE13007, or some the MJE13009? 152-0655-00 A115AX39 ??? Can not find this type of rectifiers at all¡­ 152-0400-00 1N4936 152-0686-00 MR821 Any help welcome! Best, Joe
Started by HM8001 @ · Most recent @
How to clean the finger contacts on an SG503 13
I've got an SG503 on my bench that seems to be working fine on all ranges with nice leveled output, etc. However, the finger contacts under the cam switch need to be cleaned as I have to fiddle with the switch a bit to make contact at some settings. It is my understanding that the best way to clean the finger contacts is to get a piece of paper/cardstock with a bit a 99% isopropyl under the contacts and gently pull it across the contacts. For those who have done this before, my question is what is the best approach to do this with a SG503? Is there a way to do this without a major disassembly? I'm not adverse to disassembling the module but I don't wan't to overcomplicate things if I don't have to. Thanks! Jay
Started by Jay Czaja @ · Most recent @
Tektronix 2247A problem. Keep or return? 31
Hi folks. I recently purchased a Tek 2247A scope and upon testing it for the first time, it locked up when testing the B sweep with a msg that said 'INVALID INTERRUPT FROM 0993:F082' When I rebooted, the screen has an indecipherable error message that appears to read something along the lines of 'Part Number 100-xxxx-00' with a bunch of exclamation marks. I couldn't clearly read what the part number as it was out of focus and could get it to focus well enough to read the numbers. The 'Auto Level' and 'Auto' LEDs were flashing which indicates both a RAM and ROM problem according to the service manual. The SGL SEQ LED was also flashing but there's nothing in the service manual that mentions that LED, or at least nothing that I've found. I let the scope sit for 10 minutes and it booted back up in service mode and gave a failed battery warning, and ran through self calibration. I left the unit on and it locked up after about 10 minutes with the same error as before. After the reboot I went into the service menu and ran the self test 'until fail' and the test ran for 30 minutes without issue. I also ran through the self calibration once more. It's now been powered on for a few hours without any problems or locking up and all functions 'seem' to be working as expected. Thoughts? Would something like this be caused by a battery issue? That seems like an easy fix but if it's something deeper, I'm not sure if I'd want to chance it. Thanks! Jay
Started by Jay Czaja @ · Most recent @
MSO4104 Problems 5
Hi. I've got a faulty MSO4104 here (not the B version) and it has a bit of a history. Someone at work blew up channel 1 on the 1Meg input many years ago but the scope was still useful on the other three channels (and on 50R input on channel 1). So it has been in use like this for several years. However, someone at work recently upgraded the firmware on it to V2.68 and this caused it to report a failed SPC self calibration. So it halted the SPC after a few seconds. This left the scope in a bit of a mess with lots of sampling artefacts on the traces and some trace distortion on all four channels. So it became fairly useless. I've had a look at channel 1 and restored normal operation on the 1 Meg input (a resistor was blown open circuit) and so I was hoping that it would now pass the SPC. However, it still fails although it gets a lot further into the SPC before failing. I think it might be failing on channel 3. Does anyone know what the format is for the error log file? It gives a date and some words and numbers on each line. I think there may be an issue with channel 3. I also have my own MSO4104 here to compare against. It is running V2.68 FW but it is a newer instrument. Is it possible to go backwards with the firmware to an earlier version of FW? Any other suggestions to get any meaningful debug info from it during the SPC process? Can it be logged remotely whilst doing it? I've looked through the user and service manuals for stuff about the error log but can't find anything.
Started by jmr @ · Most recent @
Tektronix 2445B for sale
no personal connection https://sacramento.craigslist.org/ele/d/sacramento-tektronix-2445b-oscilloscope/7843398888.html
Started by Robert Simpson @
Tektronix MDO3024 3
Hi, I have a Tektronix MDO3024 scope. Working great ! I would like to have on the screen the analog CH1 screen and the RF signal. Is it possible on the MDO 3000 series or it is only with the MDO 4000 series ? Thank you Matthieu
Started by MattSNE @ · Most recent @