Re: SG503 amplitude precision
I'm new here and I'm looking for someone to help me with my Tektronix 2465 DNS (aslo known as 2465 CTS ) I made few videos here: https://www.youtube.com/watch?v=oxZZ7-VjCQ0&t=1s about my sick scope.
By
Tony Fleming
·
#156457
·
|
Re: [Test-Equipment] Suggestions for going through Tek2465DVS/DMM?
I have the same oscilloscope. It is also listed under (at least that is what I found today) as Tektronix 2465CTS (or 2465 CTS) . Mine has some problems, sometimes I loose the text on top and bottom of
By
Tony Fleming
·
#156456
·
|
Re: SG503 amplitude precision
Hello Jared and Rick, Before you make an investment in this enterprise you are considering, you may want to go back to the beginning of these posts from 2015. It was categorically stated the the real
By
Jack Reynolds
·
#156455
·
|
Re: SG503 amplitude precision
Hmmm, If that's the case, I have a crimper and dies for the high quality Canare brand 50ohm BNC. I wonder if it's worth me making a batch of cables to send out for the cost of parts and shipping if
By
Jared Cabot
·
#156454
·
|
Re: SG503 amplitude precision
I may be a little late to the game but I posted this question to a Tek expert on Youtube: https://www.youtube.com/user/w2aew This comment/request might be a little out of scope, to calibrate a Tek
By
Rick
·
#156453
·
|
Re: I wonder if there is much interest in vacuum tube oscilloscopes such as the 500 series
Nearly all of the Tektronix products from the 1960's were a marvel of superb designs. I actually prefer the older scopes as long as high bandwidth is not an issue. These old scopes are generally very
By
Gary Robert Bosworth
·
#156452
·
|
Re: I wonder if there is much interest in vacuum tube oscilloscopes such as the 500 series
John and others, I have a 547 and a 545B that I plan to run for a long time. One of the two late 1A1 plugins I have has low gain in one chancel. Not diagnosed yet. Have had other scopes of that era
By
Roy Morgan <k1lky68@...>
·
#156451
·
|
Re: Tek485 - a quick question on disabling/bypassing power supply protection
In one of your earlier posts you said you had seen spiking behaviour on pin 3 of U1624. This will drive larger spikes from the output pin via Q1614 and C1602 into the protection circuitry and can very
By
Roger Evans
·
#156450
·
|
Re: Tek485 - a quick question on disabling/bypassing power supply protection
Took up Fred's suggestion and connected -5V to the power supply board. Didn't crank the supply up high - tried 0.2A and then 0.4A current limit. Max draw was 333mA with all combs in and P1560/1570
By
Lorenzo
·
#156449
·
|
Re: 11403A question
Actually, nothing wrong with the touch panel, as far as I know it is working. It is a spare I acquired when I bought some other equipment. Thanks all for the help. 73 Bill
By
bill K7WXW
·
#156448
·
|
Re: Tek 1480C waveform monitor FREE
See the link to the discussion on the waveform monitor vs an scope. /g/TekScopes/message/156263 NielsenTelecom
By
[email protected]
·
#156447
·
|
Re: Tek 1480C waveform monitor FREE
Thanks to all that replied on this subject. I've decided to make this Tek 1480C waveform monitor free to anyone in the chicagoland area. The only testing I've done on it is to verify the current draw
By
[email protected]
·
#156446
·
|
Re: 11403A question
614-0727-00 was only used in 11401 11402 used 614-0785-00. CSA803 used 614-0864-00. 11403A used 614-0850-00 which was only used in 11403A. Since this is a complete assy including the front panel maybe
By
zenith5106
·
#156445
·
|
Re: Tek485 - a quick question on disabling/bypassing power supply protection
Squeezed in a bit of time to check for shorts to ground on all the -5V points. There's no sign from my previous scope traces and multimeter probing that any of the other supplies are faulty. Long list
By
Lorenzo
·
#156444
·
|
Re: 11403A question
Hi Bill, In 11401 it was 614-0727-00/01 A9 Touch Panel Circuit Board Assembly (the IR diodes and sensors are on that board) . I feel quite sure that's the same as in my CSA803. No guarantee for your
By
Albert Otten
·
#156443
·
|
Re: CRT Hood for 7000 series 'scopes?
Thanks, Dennis! I will keep my eye open for one of those. Sean
By
Sean Turner
·
#156442
·
|
Re: Pacific Measurements or Wavetek 1038 System
Hello Walter, I removed all pushbutton switches from the frontpanel board. I measured them and all are in an open state. I expected some to be stuck in the closed state, which would explain the not
By
Hans Boon <hans.boon@...>
·
#156441
·
|
Re: Pacific Measurements or Wavetek 1038 System
Good history of Wavetek here. https://www.eetimes.com/document.asp?doc_id=1224975# John
By
John Kolb
·
#156440
·
|
11403A question
If anyone here has the service manual for the 11403A digitizing scope, can you do me a favor and dig up the part number for the faceplate/circuit board that surrounds the display? Much appreciate the
By
bill K7WXW
·
#156439
·
|
Re: Again a 7854 power supply beast to be repaired...
Welcome in the hell of 7854 SMPS repairing GeorgeP! Your contribution is more than welcome! Can you post some relevant images of the signal present at various test points of board A12 and A23 when in
By
unclebanjoman
·
#156438
·
|