¿ªÔÆÌåÓý

ctrl + shift + ? for shortcuts
© 2025 Groups.io
Date

Re: Yet another 465B restoration question

Don - AC2EV
 

Ripple is within spec.

Why did I replace the capacitors?
1. It's relatively simple and cheap to replace them
2. I've had capacitor issues in other equipment. The biggest offender being the Rifa caps in my other oscilloscope

I did suspect the CRT but given that the scope works in X-Y mode that lead me down that route.

Why the pre-amp and sweep? This is where the manual lead me through the troubleshooting so far. It may be the wrong path, but it's a path I can follow to see where it leads.


Re: Tektronix TDS3034B Boot Challenge - Help Needed

 

To Guy's point, a relatively simple test if you're set up for it would be
to dump the flash a few times over the course of the warmup time and see if
it always reads correctly.

I can take a look at the CLKOUT on another scope if that helps, but it
sounds like you're always seeing 25MHz even when the scope works fine?

Dave Casey

On Sun, Feb 11, 2024 at 10:36?PM Mike <mjs@...> wrote:

Guy,

Thanks for your insights Where I have been focused the last day is to
try and understand why the CLKOUT stays at 25mhz both in the hung and
running fine state. Actually its probably 24mhz due to the 48mhz
oscillator, I think its supposed to be doubled to the log stated 50mhz
(48??) as one of the ending steps of the boot sequence. There may be a
PLL lock problem or the clock setup fails for some reason, I really do not
know for sure at this point that the 50mhz applies to the MPC860 core and
the external CLKOUT remains at 25 as I am reaching out for conformation.
Personally I think the external clock need to be 50mhz as 25 is really slow
for what they have here, all that said it seems to run fine at 25mhz and I
have not seen any issues after a successful boot.

Temperature testing is a bit of a challenge and I have spent a bit of
time on it since the failure mode really mimics a sensitivity to being
warm or cold but nothing I have tried will force it to ether the fail or
pass state. I have been heating or cooling primarily in sections of the
board from larger areas down to the device level. I have done the whole
board and supplies but intentionally limited it in the acquisition section.
I use a large and small heat gun and have used most of a large can of
freeze spray, I track the temperature with a handheld FLlR IR camera that
has resolution down to the SMD passives sizes and shows the temp very
accurately.. Temp wise my upper limit has been ~160f to -40f or so for
some spot areas like the DRAM. I will work up a plan to try to temp cycle
the whole thing as you suggested, definitely worth a try.

When the unit is in the hung state I did see the refresh but did not
measure the period and will put that on the list. After all time spent
with this if I could offer some off the wall perception of the problem that
may or may not make any sense: Its seems like its a 'charge' or maybe a
given 'impedance' somewhere and their value shifts in one direction while
its on and the other direction while it is off -very- slowly. This 'charge'
or 'impedance' change over time is not affected by temperature very much
but when the value crosses some point it will cause it to fail.

Best Regards,
Mike







Re: Method to measure HV

 

I finally thought to look and that "odd" resistor is a 1090M-ohm that's meant for an RCA WG-289 probe. I haven't done anything to check the resistance as nothing I have can do that. I might connect one of these 45M-ohm resistors in series with it and measure the voltage across the 45M-ohm just to see if the 1090M-ohm is still functional.

I finally found a reasonably-priced HK40-6 which should be her midweek and I can finally measure the cathode voltages of my Tek scopes safely and somewhat accurately.

Thanks,
Barry - N4BUQ

If I'm not mistaken, there's an "odd" resistor with something like 90.9 labeled
on it. I'll have to get that back out and see exactly what value it is. I got
these resistors from an estate sale and it looks like they've never been used
for anything.

Yes, it's just to get the HV set on the cathode of the CRT. Technically, it's
supposed to be within 1%. I have a Triplett 630-NA that has a 6000VDC range
that I've used as a rough check but I want to get this much closer than that.

Thanks for the good info.
Barry - N4BUQ

----- Original Message -----
From: "edbreya=yahoo com" <edbreya@...>
To: "tekscopes" <[email protected]>
Sent: Thursday, February 8, 2024 1:52:17 PM
Subject: Re: [TekScopes] Method to measure HV
Barry wrote "I presume that because the probe uses a 75M and a 75k in series and
the 10M voltmeter is across the 75k, the 10M has minimal effect on the
divider's ratio. Is that correct?"

It depends on what you mean by "minimal effect." In reality, for a commercial
1,000X probe at 75 megs input, the "75 k" termination resistor is likely 75.643
k, which is about right to attain nearly the ideal value of 75.075...k when the
10 megs DMM input R is paralleled. That's why Jean-Paul mentioned being sure of
that 10 meg value.

Your 10X probe deal should work just fine, but be aware of the dependence on the
DMM's input R, and whether it's connected. If you get into a situation where
the 90 megs of R are left dangling out there, it's a current source backed by 3
kV, so can run up to 33 uA into whatever it could contact. It can't hurt you,
but could damage certain parts if contacted. If the DMM is always securely
connected, it's still a current source, but limited to 300 V. The main thing is
to keep it wrapped up to avoid any mishaps. The exact same thing can happen
with any HV probe if it's ground connection is forgotten or inadvertently lost,
so always be sure it's secure.

You can also opt for a higher divide ratio like 100 or 1,000X, which is what
most HV probes do, to get more comfortable measuring voltage, It's quite easy
to adapt any high-meg resistance into a nice HV divider. The higher the divide
ratio, the more flexibility you have in hitting the values right and handling
multiple load values. For instance, you could go with 100X instead of 10X, and
be able to drive a 10 meg or 1 meg input. With the same 90 megs of inlet R,
you'd want the bottom R to be 90 M/99, so 909.09... k. This can be conveniently
made with a 1 M termination paralleled with the DMM's 10 M input R, or, a 10 M
termination with a DMM (or scope) input R of 1 meg. These are all idealized
values, of course. The actual ones have to be worked out. It's most important
to know the tempco, voltco and actual value of the inlet R - everything else
can be worked out to accommodate it.

Anyway, since your application seems simple enough here - just checking a scope
one time (right?), and not needing extreme precision, your 10X thing is already
the way to go. It's good to consider options too, if more of such needs are
expected.

Ed



Re: Tektronix TDS3034B Boot Challenge - Help Needed

 

Guy,

Thanks for your insights Where I have been focused the last day is to try and understand why the CLKOUT stays at 25mhz both in the hung and running fine state. Actually its probably 24mhz due to the 48mhz oscillator, I think its supposed to be doubled to the log stated 50mhz (48??) as one of the ending steps of the boot sequence. There may be a PLL lock problem or the clock setup fails for some reason, I really do not know for sure at this point that the 50mhz applies to the MPC860 core and the external CLKOUT remains at 25 as I am reaching out for conformation. Personally I think the external clock need to be 50mhz as 25 is really slow for what they have here, all that said it seems to run fine at 25mhz and I have not seen any issues after a successful boot.

Temperature testing is a bit of a challenge and I have spent a bit of time on it since the failure mode really mimics a sensitivity to being warm or cold but nothing I have tried will force it to ether the fail or pass state. I have been heating or cooling primarily in sections of the board from larger areas down to the device level. I have done the whole board and supplies but intentionally limited it in the acquisition section. I use a large and small heat gun and have used most of a large can of freeze spray, I track the temperature with a handheld FLlR IR camera that has resolution down to the SMD passives sizes and shows the temp very accurately.. Temp wise my upper limit has been ~160f to -40f or so for some spot areas like the DRAM. I will work up a plan to try to temp cycle the whole thing as you suggested, definitely worth a try.

When the unit is in the hung state I did see the refresh but did not measure the period and will put that on the list. After all time spent with this if I could offer some off the wall perception of the problem that may or may not make any sense: Its seems like its a 'charge' or maybe a given 'impedance' somewhere and their value shifts in one direction while its on and the other direction while it is off -very- slowly. This 'charge' or 'impedance' change over time is not affected by temperature very much but when the value crosses some point it will cause it to fail.

Best Regards,
Mike


Re: Extender Cables for TM500/5000 series - Alternatives?

 

Ke-Fong Lin,

Thank you so much for your offer! I mean if it wouldn't be too much trouble for you, that would be incredible! I'd be happy with just one cable extender. Also would be willing to pay for more than just shipping!

-Frank


Tektronix 7904 rear AC mains range selector switch missing

 

Hi, I want to know the connections of the AC mains range selector switch of the 7904. This isn't the A model. I know the AC mains range selector switch is similar to the Tek 465.

I don't have the scope with me as it is at a recycling center. So I can't supply more images. But the guy who owns the shop said I can have it for free if I manage to get it to power up. I've bought from this guy numerous times before so I trust this claim.

The scope is in pretty rough condition, with dirt everywhere inside. I think it might have been rained on. But I think the main thing that is preventing power-up is the missing AC range selector switch.

The AC range selector switch uses two switches. One that determines if it runs on 220v or 110v. The other determines the Hi-Med-Low setting. I'm missing the one for the Hi-Med-Low.

I want to know the wiring inside the little white block that should go there. Are they all shorted? Are the top ones shorted together, and the bottom ones shorted together?

Thanks.


Re: Yet another 465B restoration question

 

It's low likely because of a ton of ripple. More load ---¡· more ripple.


Re: Yet another 465B restoration question

 

Don,

You have bad condensers. That is the most likely problem. Being from '78, that is enough years to warrant replacing them. I have a B version I put new condensers in as well as increasing the wattage of some resistors. The tantalums are know to go bad. A good quality, i.e. Nichicon ULD 47mfd 25V, electrolytic will replace them. I would mine to have Option 5 like yours.

Mark


Re: Extender Cables for TM500/5000 series - Alternatives?

 

@David Partridge and Ke-Fong Lin
The extenders you both DIY¡¯d are incredible! Thank you for sharing. I was
hoping for something that would not require PCB fabrication, but beggars
can¡¯t be choosers!
I have some spare PCBs that you can have for cost of shipping.


Re: Yet another 465B restoration question

 

On Sun, Feb 11, 2024 at 02:40 PM, Don - AC2EV wrote:

Device Under Test:
Tektronix 465B #107437 with option 5 (TV Sync Separator). Boards have date of 1978d
Nice scope, that option is what I'd hoped for, but didn't get, when I found my 465B. Not useful anymore, with digital TV. The rest of the scope is a great classic Tek portable.

First thing I'd suggest would be to turn down the beam Intensity. You don't want to burn the phosphor.

Next, take a close look at the +5 V rail. Yes, it's too low, and it can be difficult to figure out whether it's a power supply problem or too much load from a failed component somewhere down the line. I'd start by looking on that rail for the AC voltage. If things are good there, the AC component will be mV. With your DVM leads connected to the +5 and ground, switch the DVM from DC to AC. If there's lots of AC (ripple) then the PS is the more likely culprit. Look for an electrolytic cap that is no longer a capacitor. One way to test for this is to simply solder a similar capacitor across the Tek one (with the power OFF!) and see if the problem disappears.


Re: Extender Cables for TM500/5000 series - Alternatives?

 

That's very kind of you to offer Eric! I greatly appreciate it. I think I'll shoot an email over to Dan to see if any of those kits are available. I'll keep looking before asking to borrow!

-Frank


Re: Yet another 465B restoration question

 

Always good to check the voltages rails first which I see you did. But I would still be interested in figuring out why that 5V rail is so out of spec (should be 5 +-0.075V). Also, I would check the ripple on those power rails as well. It should be very low (~2mV Vpp).

I really don't understand why people feel the need to replace all caps just because the unit is old. I'm glad it didn't cause any further issues because that can be one of the downsides to doing that.

Why do you suspect the issue being in the preamp section? Does the anomally appear in signals for both CH1 and CH2? Both channels have their own preamp circuitry so if the anomally is appearing on both channels, I would suspect the issue being further downstream of the inputs. Possibly the vertical output amplifer.

I would also be looking at the CRT circuitry. In some of the photos you show, the trace rotation is clearly off. It's possible you might have some issues with geometry of the display.

You have the right idea by injecting signal and following it thru the vertical amp circuitry to the CRT. If the issue is there, with the help of the service manual, you'll be likely to find it.

I've never owned or worked on a 465B, but I'm sure there are plenty of people in this group who have that will be able to give you more specific suggestions. Also, might be helpful to search through the forum to see if this issue has popped up before. Use the search box on the top right.

Good luck!

-Frank


FS: Tek 453, and a few other items, near Madison, WI, USA

 

An acquaintance near here says he has some Tektronix gear that used to belong to his deceased father, who formerly worked for Burroughs. He sent me some photos of what he could get to easily, but that only included a Tek 453, a DeVRY Technical Institute scope (sounds like a kit built as part of an electronics correspondence course) and two old analog VOMs, which I can't identify.

I gather that there are some Tek plugins of some kind buried farther out of reach, but I don't know what might be there. I'll keep on him for more info.

The condition is dirty, and otherwise unknown, so the price is likely to be cheap, or less. He would just like to find them a good home.

I will continue to bug him for more info on the other parts, but in the meantime if there's anyone in this area that's interested, let me know. I can send photos and put you in contact with the owner.


Yet another 465B restoration question

Don - AC2EV
 

First Post! Not seeing much by the way of editing tools for posts so I'll do my best.

Test equipment:
Hantek DSO2C15 oscilloscope
Tektronix 2465A oscilloscope
Fluke 87 V multimeter
DE-5000 LCR meter
Siglent Technologies SDG1032X - set to arrive later this week

Operator Level:
BSEET "engineer" trained in soldering and de soldering correctly. Familiar with schematics, P&IDs, pinning charts, flow diagrams, etc.
Have restored several pieces of vintage equipment (Tube amps, tube test equipment, Epson HX-20 laptops,etc)

Device Under Test:
Tektronix 465B #107437 with option 5 (TV Sync Separator). Boards have date of 1978

Last Calibrated:
6-30-1994
Last Used:
~20 years ago

Synopsis:
X-Y mode mostly works with extra signal anomalies present (rolling of signal on screen), Channel modes have some odd interactions with time base and cause lockup. Can sometimes get a signal to display properly.
+5VDC rail is low at 4.11V. All other supplies are within spec to the service manual including the high voltage.

So far:

Applied DeOxit D100L to push button switches only
Checked vertical attenuators with no signs of corrosion. Leaf switches look clean and functional, I didn't touch them as they appear very fragile
Gave all switches, knobs, buttons, levers a good workout.
Oscilloscope appears to be in good physical shape. Circuit boards appear to be visually OK, no signs of damage or leakage
Replaced all tantalum capacitors with aluminum electrolytic

Troubleshooting via ther service manual indicates possible faults in the Vertical Pre-Amp and A+B Sweep generator. I've tried getting a trace to appear but I've found many times the scope appears to "lock up". The trigger LED will light up and then basically everything stops functioning and I have to turn the scope off and back on.

Vertical Pre-Amp
+4.14V connected
+4.36 disconnected

Once I get the signal generator I'll start tracing the signal through the channel A vertical amp. In the meantime I pulled each transistor on the vertical pre-amp board and checked them for "go/no-go" with a arduino based component tester. They all tested "go", I did not pull the datasheets and compare the specs.

I did all the testing with the tantalum installed and then with the electrolytic. Replacing the capacitors did not fix the issue or introduce any new issues.

Are there any common failures that I should be checking for before going into the in-depth signal trace?

I've put some pictures up on my Google photos. These images show some of the anomalies when applying a sine wave and square wave using a cheap signal generator. The anomaly is not from the signal generator (verified with other scopes)


Tek575 Serial No: 100082 Guernsey Ltd HV Transformer ?

 

I took a shoot in the dark an bought a non working 576 from a local classified section (something fell on top of it and cracked the aluminum bezel and sheered of a couple of the knobs on top.
Apparently when you switch it on only the screen illumination works based on what the seller told me.
I¡®m not able to pick it up for a couple of weeks but got some pictures.
Interestingly both high voltage covers are metal compared to the ones I worked with so far which were Metal (flyback) and brown ?plastic¡°.
It also has the little solder holder in place.
Know the big question is what Version the flyback transformer will be brown or black¡­
Do you have maybe any details what I can expect?
The seller also confirmed that the CRT looks ok¡­we will see¡­
Thanks for your help


Re: Dennis Tillman - Update

 

I am glad to read that Dennis is doing well. I hope he has a great and speedy recovery.

It would have been nice if there'd been a Tektronix 576 in the operating room. Dennis could've given them some pointers.

Barry - N4BUQ

From John with permission, personal friend of Dennis...

Just got a call from Dennis. He sounds very good.

On Saturday, Feb 10, they put in only one stent, the second being unnecessary.
The one stent went in high up in the aorta rather than one in each in branch.
I remind you that I still am 4 years away from my MD, so please excuse me from
any medical errors.

Dennis was very impressed with not only the size of the room used in the
procedure, but the number of machines there. Especially impressive was the
computer controlled (of course) X-ray machine that took pictures from different
angles and, of course, gave immediate results.

The doctors told him he was the ideal patient, totally immobile while under
total sedation during the actual stent installation. (Not everyone is immobile
during anesthesia.) This is important because the stent is inserted in an
artery in the groin and any motion there could result in major bleeding. They
told him to call 911 if that happened. Medical humor.

He will remain in Massachusetts for a while so the doctors can monitor his
recovery. He expects to be on a plane back to Bellevue on about February 25.

It was a long Thanksgiving vacation.



Dennis Tillman - Update

 

From John with permission, personal friend of Dennis...

Just got a call from Dennis. He sounds very good.

On Saturday, Feb 10, they put in only one stent, the second being unnecessary. The one stent went in high up in the aorta rather than one in each in branch. I remind you that I still am 4 years away from my MD, so please excuse me from any medical errors.

Dennis was very impressed with not only the size of the room used in the procedure, but the number of machines there. Especially impressive was the computer controlled (of course) X-ray machine that took pictures from different angles and, of course, gave immediate results.

The doctors told him he was the ideal patient, totally immobile while under total sedation during the actual stent installation. (Not everyone is immobile during anesthesia.) This is important because the stent is inserted in an artery in the groin and any motion there could result in major bleeding. They told him to call 911 if that happened. Medical humor.

He will remain in Massachusetts for a while so the doctors can monitor his recovery. He expects to be on a plane back to Bellevue on about February 25.

It was a long Thanksgiving vacation.


Re: Walter Shawlee obituary in the NY Times

 

It was amusing that the movie shows NASA engineers checking /sums of numbers /with their slide rules. The old-timers in the theater groaned quietly but audibly.

--
Prof. Thomas H. Lee
Allen Ctr., Rm. 205
420 Via Palou Mall
Stanford University
Stanford, CA 94305-4070

On 2/11/2024 8:37 AM, stevenhorii wrote:
The movie ¡°Apollo 13¡± famously showed the astronauts
doing a calculation on a slide rule. It was the ¡°pocket¡± model and was
identical to the consumer one


Re: 2465B Probe Compensation Output

 

The probe compensation signal (referred to as the calibrator in the service manual) is different than most other scopes as Jon notes. It will produce a 5 cycle square wave on any sweep rate set between 100ms and 100ns/div. There is a 200ns shift added to the chop signal at the end of some sweeps to desynchronize the chop signal from the sweep rate. The frequency of the calibrator is controlled by the same divider chain that controls the vertical chop rate. So the 200ns can show up in the calibrator signal as irregular pulse width. Apparently it only shows up when measuring the calibrator signal with a separate instrument. But maybe try measuring the signal with trigger set to Single sequence to shut off the sweeps. see if it disappears? Also, the comp signal has output impedance of 50ohms. Maybe try measuring with output set to 50ohm to match impedance?


Re: Walter Shawlee obituary in the NY Times

 

Jim,

There actually were slide rules designed for particular specialties and
occupations. A company called Hemmi (Japanese, and still around) made all
manner of specialist slide rules. I believe they did make an ¡°electronics¡±
one as well as ¡°chemistry¡± and quite a few others. I think they would even
make custom ones if you could put enough orders together.

Materials: I think the wood slide rules were mostly mahogany but bamboo was
used a lot - apparently fewer problems with humidity and bowing. Pickett is
famous for their metal rules. The early ones were magnesium but later
production switched to aluminum (likely because of magnesium¡¯s oxidation).
Pickett is also famous for their yellow color. They claimed it to be easier
on the eyes and the models were the same as the white background ones but
had the designation ¡°ES¡± appended to the model number. Pickett is also
famous for supplying the slide rules carried on the Apollo missions (they
may have also had them on earlier NASA missions, but I know for sure about
the Apollo ones). The movie ¡°Apollo 13¡± famously showed the astronauts
doing a calculation on a slide rule. It was the ¡°pocket¡± model and was
identical to the consumer one - no special scales - but NASA had a beta
cloth (the fire-resistant cloth that was used for various bags for things
as well as the outer layer of the space suits) case rather than leather.
After NASA selected Pickett, they even sold that particular model in a box
with some text on it about it being used for space missions. If I recall
correctly, Walt actually had some of these new in that box. They are highly
collectible now - if you have the original box in good condition. Most of
the manufacturers made plastic rules in addition to wood ones. K&E is well
known for this as is Faber Castell (also made a lot of drafting instruments
and pencils). A highly collectible slide rule is one with an electronic
calculator on the back! An earlier one had an addiator-type calculator on
the back (useful since you really can¡¯t do addition and subtraction on a
slide rule).

This site describes a collector¡¯s information about a huge variety of slide
rules. Among them are the electronics rules:



and this Web site is specifically about electronics slide rules and how
they are used:



It¡¯s the section from ¡°Steve¡¯s slide rules¡± It also has instructions on how
to use some of the specialty scales.

There¡¯s a slide rule group on groups.io. There¡¯s also the Oughtred Society (
www.oughtred.org) another very helpful slide rule group.

Circular slide rules? One - you don¡¯t ¡°run off the edge¡±. Two - because of
circumference/diameter, with a 4¡± diameter circular rule, the outermost
scale would be over 12¡± in length - longer than most regular rules. There
are some circular rules that are 10¡± diameter - that would be equivalent to
a huge straight rule.

I just checked - Walt¡¯s ¡°Slide Rule Universe¡± had (may still have?) a
Pickett N515 Electronics slide rule as well as a 535T electronics rule. I
found them on the ¡°old¡± site but it may be worth a call or email to Susan
to see if any of these are available:



Good luck and a warning - you acquire one slide rule and it¡¯s all too easy
to become a collector. Very much like buying one Tek scope and assuring
yourself (and your spouse or partner) that you really don¡¯t need any more!
If you do manage to get one, think of it as part of Walt¡¯s legacy.

I also did a check - there are quite a few Pickett N515 electronics slide
rules on eBay. Since they were used by the Cleveland Institute of
Electronics, it¡¯s not surprising that so many were sold.

Steve H (despite being a collector, not the ¡°Steve¡± of that slide rule Web
site)

On Sat, Feb 10, 2024 at 23:37 ¸é±ð²Ô¨¦±ð <k6fsb.1@...> wrote:

I too thought the article was quite wonderful....and re slide rules.
Walter and I talked about them a bit. great guy and sorely missed..i
still use my 6 in slide rules ....my 18 in is bamboo and belonged to my
dad. Dad gave me a 6 in pickett aluminum while in high school....I used
it in College along with my 6 in circular ( it was tossed by friend
after getting "the NEW hp35") for my engineering classes...most of my
colleagues had the HP calculators (( out of my budget!)...so on the top
on exams I had to write "slide rule" since my numbers were not
"EXACT"...but good enough....I Still use them all....but I do like my
HP35, 25 and 11C all of which I use...but the slide rule is always
around if the batteries die..especially in the 25 and 35...battery packs
need rebuilding....a 6 inch is always in arms reach even in the car!
oh well fond memories, Walter will not be forgotten.
¸é±ð²Ô¨¦±ð

On 2/10/24 7:50 PM, Jim Ford wrote:
Yes, I had only bought once or twice from Sphere, but Walt and Susan did
include some useful goodies. Say, that NYT article mentioned slide rules
made out of bamboo, and I'd never thought about it until today, but my
dad's old slipstick from the 1960s had the open grain at the ends and was
almost certainly made of bamboo! I'm too young (born in 1965, engineering
school 1983-1991 including grad school) to have used one, but I'm starting
to think I missed out. Say, what make and model would you folks recommend
for an EE with a bent toward RF, microwaves, and fiber optics? Steve,
interestingly enough, my dad is a chemist turned software engineer who used
to teach p-chem! Definitely not my thing, and I jumped to EE after starting
freshman year as a biochemistry major; those chem labs blew it for me.
Rest In Peace, Walter. Jim Ford Sent from my T-Mobile 4G LTE Device
-------- Original message --------From: stevenhorii <
sonodocsch@...> Date: 2/10/24 6:08 PM (GMT-08:00) To:
[email protected] Subject: Re: [TekScopes] Walter Shawlee obituary in
the NY Times I miss Walt. I first got acquainted with Sphere Research
through sliderules. I had saved mine from my university days. I still have
the DietzgenN1725 Vector Log Log. My father was an architect and bought it
for me at adiscount since the firm he worked for had a contract with
Dietzgen fordrafting instruments. That slide rule got me through p-chem. I
only usedthe hyperbolic scales to see how they worked. I did use most of
the othersfor homework problems.A lot of my collection came from Sphere
Research - new-in-the-box Pickettrules including their famous N4, also a
vector log log. I have one of thosenuclear bomb effects circular slide
rules and got it in the book ¡°TheEffects of Nuclear Weapons¡± but on eBay,
not from Walt. Since then, I alsofound a Soviet version. Being a
radiologist, I thought that understandingthe dangers of the radiation from
these weapons was something I should knowabout.It was only later that I
started buying electronics from Walt. He wasalways generous and would
include a bonus bag of parts. Most were farbetter than the usual ¡°floor
sweeping¡± sorts of stuff; a lot of usableparts in those bags.I¡¯m sad that
he¡¯s gone but glad Susan carries on. The effects of his yearsof catering to
many of those in this group are probably well known sinceI¡¯m sure many of
you would turn your browser to his site the minute his¡°Stuff Day¡±
announcements appeared.Steve H.On Sat, Feb 10, 2024 at 16:37 John Griessen
via groups.io <john@...> wrote:> That was a nicely
done bio.>>> >>>