Re: Delay Time Position Vernier
On Fri, Nov 27, 2020 at 05:08 PM, Dave Peterson wrote:
The dial shows the delay from the start of the A time base to the start of the B time base in time units selected by the number on the A time base selector (if in the cal. position). Usually, this is observed using the B trace. So, with the A time base set at e.g. 1 us/div, when changing the value on the dial from e.g. 3.61 to 4.61, the B trace moves by 1 us (if B is set to "Run after Delay"), *whatever* the setting of the B time base. The fact that the number relates to the A, and not the B time base may seem counter-intuitive at first. Although the "delay time multiplier" knob, as it's called, is almost always used differentially, I guess the "offset" mentioned by some compensates for the absolute minimum delay (for the start of the B time base), achievable when setting the delay time multiplier knob CCW. Raymond |
Re: Tek 3A1 Module
" Ummm, FairRadio (in Lima, OH) has (had?) some of the more common ones in
back... - (No affiliation, just a customer)." Thank you David. I have emailed them, as you suggest. Their web site shows that they do have one, or more. I emailed to ask them if the modules were complete (I did say that I can not ask them if they work, just that they are not stripped of tubes and other parts). I'll await their response. They are only 6.5 hours away (lol). Once all this pandemic BS is over and we have our President Biden manaded inoculations, I would like to take a weekend trip out there and visit them and the big flea markets between Dayton and Cincinnati. Sounds like a nice trip for this spring or next summer (I pray this virus is over by then). In the meantime, I can have them ship stuff to me. Dave On Wed, Nov 25, 2020 at 3:51 PM David Holland <david.w.holland@...> wrote: Today, with that which shall not be named still around? |
Re: TDS784C ROMs
Thank you for trying to help, but none of those links work for me to actually find the commands needed over GPIB to dump the ROMs. Any other hints?
toggle quoted message
Show quoted text
Steve Hendrix At 2020-11-24 01:30 PM, you wrote:
Hey again, |
Re: Delay Time Position Vernier
And the value of the DTP Vernier was irrelevant, right?
So far we have that it appears there was an initial setting by model, but no definition to the reason or significance of that value. Does anyone have a working setup in which the value of the Vernier can be checked? Does it indicate time values? If so, wouldn't it have to be scaled according to the time/div setting? Now that I think about it, isn't that all this dial can do? Indicate decimal values from 0.00 to 9.99? I don't think I've ever tested it turned it fully CW. Does the pot stop at 10 turns? That's kind of a rhetorical question. I'll try myself when I get in the shop. I also need to do my own homework and review the calibration procedure in the manual. Dave |
Re: Delay Time Position Vernier
When I was a pup in the emerging field of digital data storage and field recorders, I was presented with a 465B for my bench. I remember staring at it for a minute or so, until the senior tech reached over and pulled on the power knob. I still hadn't found it...
Learned to love the 465, and still have the refurb one I bought with a loan from Grandma in the previous century, $2K The data recorder (prefer not to say what it recorded...) used dynamic RAM with a 1mS refresh rate, and a 250 nS read cycle. Running the scope at 100 uS / Div on RAM clock the gave you a nice parade of closely spaced pulses at the left side of the screen, and over at the right side you might see the beginning of the next refresh cycle if the clock crystal in the recorder was off a bit. We would use the B-Delayed sweep to slowly scroll through the sequential addresses to the RAM, and thereby be able to see any desired address of the lower 8 bits of the bus as the address incremented. Pretty much a similar use case as with analog TV, the variable delay allowing you to scroll to a particular line, or portion of a line, in the video scan... _Dave KC6UPS |
Re: Tek Glass Tracing
The main feature of the photograph you provide looks more like some sort of residue to me than a crack (if I'm looking at the correct feature in the photo.). It's also not clear why a crack would develop there as it does not appear to be a flexure or stress line. The photo you provide has neither the magnification to see cracking, nor an indication of where you think the cracking has occurred.
While you may already understand that all you state is a hypothesis that needs to be proven and have made measurements to verify it, your post does not indicate you do or have. Cracking can be tricky in printed traces. You should be able to prove your hypothesis by probing the trace. You'll want to use a rounded probe tip, or a small piece of metals as a pad so as not to damage the trace. Fix one end and drag the probe along the trace. resistance should increase steadily for a healthy trace and increase abruptly at a crack. Most of the time when I see cracking in these kinds of older flex circuits it's at the connector pins. More often than not it's invisible without considerable magnification and careful lighting. I've repaired these kinds of traces with conductive epoxy. If the crack is at a mechanical stress location you'll want to use a small jumper that terminates outside the stress zone and is formed to survive the stress source (like a bend or repeated motion.) I have some 17 strand silver plated 30 AWG teflon wire that I use for this kind of thing. -- Oz (in DFW) N1OZ |
Re: (OT) Where to go for 70s IBM hardware? I'm looking for a terminal.
Just wanted to ask again if anyone has any 70s IBM stuff. System/3, 5xxx,
toggle quoted message
Show quoted text
etc. Old terminals. You can tell them apart because the keyboards look more like a typewriter than a computer, and the gear is usually encased in thick metal plating, and often comes in the form of a full desk. On Fri, Nov 27, 2020 at 4:14 PM n4buq <n4buq@...> wrote:
Hi Steve, |
Re: (OT) Where to go for 70s IBM hardware? I'm looking for a terminal.
Hi Steve,
toggle quoted message
Show quoted text
Looks like your email address and the version that uses "at" and "dot" aren't the same. Is that a typo? Thanks, Barry - N4BUQ On 11/27/2020 9:06 AM, stevenhorii wrote:
Bill, |
Re: (OT) Where to go for 70s IBM hardware? I'm looking for a terminal.
Bill,
Since this is so far off-topic, please feel free to contact me off-list: sonodocsch@... If this gets blocked, it is sonodoc at gmail dot com. My interest in guidance and nav stuff is not professional but I¡¯d guess you would call it ¡°hobby¡±. I have quite a bit of Litton stuff - a lot turned up in the Los Angeles surplus stores since they were in Woodland Hills. I¡¯d be interested to discuss my interests in the G&N hardware. I¡¯m a radiologist by trade, but I have more books on G&N than I have books on radiology. I am also a member of the Institute of Navigation as well as the IEEE (well, in addition to the American College of Radiology and a few other radiology societies). Steve Horii On Fri, Nov 27, 2020 at 09:38 Bill Higdon via groups.io <willard561= [email protected]> wrote: Is the F-111 Nav computer part of the INS system? Also the Navy boxes |
Re: worst condition Tektronix scope?
For decades I have taken equipment that was discarded in disgust or stored horribly and restoring it to operation. Initially, it was because of poverty, now I get great enjoyment from rescuing it from the dumpster and solving the puzzle of it's problems. In 1971 I had returned from duty in southeast Asia to Griffiss Air Force Base in Rome, New York to finish my term. Having worked on some neat stuff, I was interested in having some of the rigs for my own. At that time, a Tektronix 524 would cost $200.00 or more, and a Collins R-390 receiver would be $500.00 . Once I got to the base, I felt certain there would be surplus stores around, and there were, in nearby Utica. Enroute to one I had been told about, I noticed piles of gear out in the open in the heaped ruins of a burned building. Some of it was totally wrecked, but most was a bit smoked up and covered with dirt and ashes. The owner was delighted to sell us anything we wanted from the ruins for very low prices. I got an R-390 for $15.00, quite literally half buried in the ground, and a friend with me got a 535 scope that was just about as bad for the same price. After an initial stop at a self service car wash to get the worst of the dirt and ashes off, we both spent many evenings and a few weekends, and a good amount of penetrating oil and contact cleaner, dismantling and cleaning the rigs. Manuals were easily obtained, I worked in the radio shop and my friend had a connection in the instrumentation lab. We were both very surprised and impressed at how well they turned out physically, indeed good enough that some of the lifers and officers gave us a lot of crap about having "misappropriated" them. Fortunately, we were both returnees from the Vietnam theater, and knew what to expect and do. We had receipts from the surplus store and pre-restoration photos. The lifers then just shrugged in disappoinment and left us alone. We both enjoyed the fruits of our work, I kept the R-390 for years, and my friend used the scope for digital experimenting and very early computer crafting. What frightens me today is so few young people having the interest or ability to do such work, and the priceless learning and skill they could gain from doing so.
toggle quoted message
Show quoted text
Bruce Gentry, KA2IVY On 11/27/20 5:59 AM, Richard Steedman wrote:
Have you read the full description of this guy's scope? |
Re: (OT) Where to go for 70s IBM hardware? I'm looking for a terminal.
Is the F-111 Nav computer part of the INS system? Also the Navy boxes might be part of an INS system. I worked at Litton in SLC back in the 70's & worked on the Analog INS as well as the digital. The analog systems had electro-mechanical inetegrators for all 3 axis. I left when I was told I was going to be put on the Astro INS, in retrospect I should have stayed, oh well.
|
Re: Choosing the right replacement cap
On 2020-11-27 1:37 a.m., Thomas Dodge wrote:
I am not sure if I am doing this the correct way. I have three TektronixThis might be a redundant suggestion but have the basic power supply checks been done? Do you have a service manual? --Toby and the best one I have. Any help with this is appreciated. I did read |
Re: worst condition Tektronix scope?
I did hear a story a few years ago about a 500 series scope that was partially melted in one of the big north bay fires (2014 or 2016?), but no pictures.? (Also, it may not even have been a Tek; it was described as something like "one of those big old tube oscilloscopes with the round screen")
toggle quoted message
Show quoted text
-Dave On Thursday, November 26, 2020, 11:33:21 PM PST, Jeff Dutky <jeff.dutky@...> wrote:
I'd like to try to drag this thread back on topic, however: I can't believe that my dirty 475 is the worst scope that anybody has seen. Somebody has to have better stories, at least, whether or not they have the pictures to prove it. -- Jeff Dutky |
Re: Choosing the right replacement cap
I am not sure if I am doing this the correct way. I have three Tektronix
scopes. I have a Tektronix 533 scope that there is no trace. I have a Tektronix 7704A, and the intensity is very very bright, and the Time/Division switch has no effect. I also have a Tektronix 684A that I bought from a surplus place, and had been on the shelf for a long time. When I first brought it home and plugged it in, it didn't do anything, so I turned it off and on again. After a while, it decided to work, and went through the self test, and it was fine. I turned it off again, and then on again, and it was still ok. I just turned it on again some months later, and now it has no display on the screen at all. Usually when it was powered on, the front panel lights will come on and start to initialize, but it is not doing that now. Can anyone help with this? This is really a great scope and the best one I have. Any help with this is appreciated. I did read somewhere that it might need a chip replacement, but not sure. Thanks very much. Thanks, Tom Dodge On Tue, Nov 24, 2020 at 5:58 PM Dave Peterson via groups.io <davidpinsf= [email protected]> wrote: What parameters besides capacitance, voltage, and temperature are |
Re: Choosing the right replacement cap
Dave,
That this is the amount of ripple current the cap can take to make the lifetime spec? YES A spec with twice the ripple for a given lifetime can take more power supply noise? YES A UCY2C470MPD1TD with 375mA ripple and 10000 hr lifetime can take more power supply noise than a ULD1H470MED with 190mA ripple and 10000 hr lifetime? YES The UCY is cruising in the 465 HV oscillator supply while the ULD might suffer degradation despite being well within voltage and temperature specs? MAYBE, and MAYBE not. It depends on the actual ripple current. If say, there's only 100mA of ripple current, then both the UCY and ULD are cruising. If there's 250mA of ripple current, then that removes the ULD from the equation. You can measure ripple current on a(nother) scope. You can FEEL how well the cap is handling ripple current by its temperature. All other factors being equal, a hot cap is battling to handle the ripple current, and a cool cap is handling the ripple current just fine. And if the cap is running hot, then the 10,000hr lifetime is meaningless. ESR is the "internal resistance" of the cap. The lower the ESR, the more efficiently the cap is passing the current onwards. It's all about circuit efficiency, translated into MTBF, and signal cleanliness down the line. SMPS's require absolutely the lowest ESR caps possible that the bean-counters will authorize (and will stop working if a cap's ESR rises above a certain threshold), but a Linear PSU is far more tolerant of higher ESR caps. Menahem |