Re: Measure -2450?
I knew probes could be used for a lot of things, but this is a new one on me: <http://www.ebay.com/itm/FLUKE-80K-15-Electronic-Air-Cleaner-Probe-/310418595539> Mark
By Mark Wendt <mark.wendt@...> · #88327 ·
Re: Looking for TDS420(A) Power Supply
Hello RaymondI have a power supply for tds 420A.How much do you want to spend?If you send in your faulty I may be able to repair it or sell you one for parts.Thanks Jan Koria
By m8jpk · #88326 ·
Re: Measure -2450?
Fluke 80K-6 high voltage probes are inexpensive used and good to 6 kilovolts. They are designed to be used with a 10 megohm input resistance multimeter which is standard. They only have 75 megohms of
By David Hess · #88325 ·
Re: Measure -2450?
Can you be more specific?? Are you measuring voltage? If so, an ordinary Simpson 260 or Triplett 630 will do, as long as it doesn't need to be very accurate. If you need precision you can cobble up a
By Bob Albert <bob91343@...> · #88324 ·
Measure -2450?
Is there a way to take this reading without a 3k meter?
By [email protected] · #88323 ·
Re: 549 with compresssed trace on right third of screen
I had a similar problem recently with Aerovox caps in my 564: one of these was defective, causing malfunction when erasing the storage screen. After some checks, the only culprit was a 0.1uF 600V
By unclebanjoman · #88322 ·
free in NYC: 7503, RM564, and TM506, for parts or repair
The 7503 needs work on the plug-in connectors. The side covers of the connectors need to be replaced. (This is a common failure mode and has been discussed in this forum.) I believe the scope is
By sipespresso <sipespresso@...> · #88321 ·
067-0587-01 sold
fixture is sold..thanks Jerry Massengale
By jerry massengale · #88320 ·
Re: 549 with compresssed trace on right third of screen
Hi Keith Areovox capicator has 0,9?F insteat 0,5?F, that is a typical paper-capicator-problem. i think, 1N4148 is better, but i am not sure. dirty contact, i use tuner 600 and q-tips. Best regards.
By mattko87 · #88319 ·
Re: 549 with compresssed trace on right third of screen
Hi Matt- Yes, thank you so much! As you said, I replaced the Aerovox cap (C335) and that particular problem is resolved! The bad cap did not show any leakage. On an ESR meter it measured much _lower_
By keithostertag <keitho@...> · #88318 ·
My new 310A
Bought a 310A for little money, in perfect condition, optically and electrically (what a sharp trace!), lucky me. 2 Questions: It has serial number 013491, rectifier is with silicon diodes. Anybody
By Roland Manser · #88317 ·
Looking for TDS420(A) Power Supply
Hi! My TDS420A has a defective mains power supply. Would any of you have one available from a parts mule? AFAIK, PSU's in TDS420 and TDS420A are identical. True and any more models? N.B. I'm living in
By raymonddompfrank <r.domp.frank@...> · #88316 ·
Re: 549 with compresssed trace on right third of screen
Sorry, my fault. 0,5?F , (C335), that is right. Heerenveen 549 has EROmet capicator, germany capicator. But, Portland 549 has bad black areovox, replaced him and 549 run fine. Best regards Matt
By mattko87 · #88315 ·
Re: Delta Standard for 130 L-C Meter has MINUS 3pF setting [1 Attachment]
================= How did Tek get the Delta Standard for the 130 L-C Meter to have a 0pF and a MINUS 3pF value? Exactly what is a negative capacitance and what is it good for? See attached photo.
By Craig Sawyers <c.sawyers@...> · #88314 ·
Re: 549 with compresssed trace on right third of screen
Hi pleas replaced 0,33?F Areovox capicator in Horz. amplifier. This capicator ist bad. I has repaired many 549, one from Nethersland (Heerenveen) has not this Problem, but Portland-549 has this
By mattko87 · #88313 ·
Re: Delta Standard for 130 L-C Meter has MINUS 3pF setting [1 Attachment]
They do not have 0 and -3pF. The fixture ( I would love to have the original, but made one myself as a substitude) has a starting capacity. I do not remember the values but for instance 3 pF on the
By PA4TIM · #88312 ·
Re: Delta Standard for 130 L-C Meter has MINUS 3pF setting [1 Attachment]
"Only the stray capacitance of the connector and switch assembly is in the circuit in the -3uufd position. The actual capacitance of these strays is approximately 10 to 20uufd. No effort is made to
By David DiGiacomo · #88311 ·
Re: Delta Standard for 130 L-C Meter has MINUS 3pF setting
The service manual is available online if you want to see how it works. "Only the stray capacitance of the connector and switch assembly is in the circuit in the -3pF position. The actual capacitance
By David Hess · #88310 ·
Re: Delta Standard for 130 L-C Meter has MINUS 3pF setting
Pages 6-6 and 6-7 of the 130 manual describe the cal procedure using the S-30 + and - 3pf setting. RP
By ROLYNN PRECHTL K7DFW · #88309 ·
Re: Delta Standard for 130 L-C Meter has MINUS 3pF setting
Since it appears to be a passive device, it is probably and inductance that's equivalent to -3 pF at the test frequency. The zero value is probably an L-C resonant (near the test frequency) circuit
By Ed Breya · #88308 ·