Fluke 80K-6 high voltage probes are inexpensive used and good to 6
kilovolts. They are designed to be used with a 10 megohm input
resistance multimeter which is standard. They only have 75 megohms of
input resistance though which may be a problem measuring many of the
CRT voltages like the grid or focus but will work fine for measuring
the cathode voltage which is regulated.
On Mon, 04 Mar 2013 04:41:16 -0000, "anson_williams@..."
<tractormananson@...> wrote:
Is there a way to take this reading without a 3k meter?