Re: Test input for an SD-32 sampling head
Picosecond pulse labs used to sell NLTL's for pulse sharpening amonst other applications. NLTLs are still commercially available. Bruce
By
Bruce Griffiths
·
#155370
·
|
Re: M48Z35-70PC1 NVRAM for CSA803 & 11801 E5622 error
Reg, There might be a more elegant solution possible. The problem of battery ram was discussed at length in the pinball forums, and people made other solutions for it. Permanent ones. No battery to
By
Maurice Smulders
·
#155369
·
|
Re: 2247A PSU Troubleshooting
To prevent that ground ring short, always practice safe probing; use a probe condom.? One of those plastic SMA connector caps works well, or just tape in a pinch.Jim Ford?Sent from my Verizon,
By
Jim Ford
·
#155368
·
|
Re: 2247A PSU Troubleshooting
Nick, I have a 2246A that I had to get the ripple on the various DC busses knocked down. Replacing the ZS diodes was done, most of the secondary DC busses caps, a number of those were bad, I could not
By
[email protected]
·
#155367
·
|
Re: Test input for an SD-32 sampling head
I have one of Leo's 100 pS impulse units as a source for initial experiments. If I can narrow that to 10 -20 pS with discrete SMD parts I may be able to get the idea to work. There's a close
By
Reginald Beardsley
·
#155366
·
|
Re: 2247A PSU Troubleshooting
Nick, That is the way some of these threads go. You don't need a variac at this stage. A while ago you mentioned you had an isolation transformer, as I recommended and you asked whether to connect the
By
Raymond Domp Frank
·
#155365
·
|
Re: TCP202 replacement body parts?
[email protected]> wrote: I'm happy to report that a kind gentleman member of this list had some parts he shared with me. I've transplanted the probe to a new housing and the patient looks to have
By
Sigur?ur ?sgeirsson
·
#155364
·
|
Re: 2247A PSU Troubleshooting
In my experience, the main transformer either works, or it doesn't... and yours works. Having an internal short to ground is what killed the few T2204's I know about. There are about 17 one amp ultra
By
tom jobe <tomjobe@...>
·
#155363
·
|
Re: 2247A PSU Troubleshooting
It was suggested earlier this this thread that I should find 160-170VDC across C2202. I tested across R2256 as they appear to share the same connections, and got 156VDC. Next I checked for ripple
By
Nicholas Keller
·
#155362
·
|
Re: Test input for an SD-32 sampling head
Chris, I think you refer here to the so-called "nose-to-nose" method of characterizing sampling scopes (Rush e.a.). Albert
By
Albert Otten
·
#155361
·
|
Re: Test input for an SD-32 sampling head
Reginald, To obtain significant NLT compression, you need lots of stages, and a fast and big enough signal to start with - not easily done with conventional electronic and transmission line
By
Ed Breya
·
#155360
·
|
Re: 2247A PSU Troubleshooting
OP here. Variacs weren¡¯t mentioned until today so perhaps you are confusing me with another user. But you are correct that this is starting to feel overwhelming and I am losing confidence :( This is
By
Nicholas Keller
·
#155359
·
|
Free for shipping: Tekelec V.35 interface module
Got it from the Telogy auctions years ago and have no use for it.? No part number, but comes with v.35 cable and internal ribbon cable.? Can provide photos to interested parties (please contact me
By
Dave Seiter
·
#155358
·
|
Re: WTB 012-0648-00 6 pin LEMO M-M 30" cable for Tracking Generator
Hi https://www.ebay.co.uk/itm/LEMO-S-SERIES-6-WAY-PLUG-FFA-1S-306-fd6j28-/121740345972 https://www.mouser.es/datasheet/2/232/unipole_multipole-973098.pdf Is 1s, 12mm, not 2s, 14,8mm Regards Miguel
By
Miguel Work
·
#155357
·
|
Re: WTB 012-0648-00 6 pin LEMO M-M 30" cable for Tracking Generator
https://www.ebay.es/itm/LEMO-Grose-2S-Kabel-Stecker-6-polig-FFA-2S-306/370994224213?hash=item5660fa9c55:g:WLkAAOxycmBS4-MI is this? -----Mensaje original----- De: [email protected]
By
Miguel Work
·
#155356
·
|
Re: Test input for an SD-32 sampling head
LOL With *great* difficulty. That's a large part of the appeal of the project. It's not very expensive and if I succeed serious bragging rights. It would be great fun to take to a Keysight trade show
By
Reginald Beardsley
·
#155355
·
|
Re: 2247A PSU Troubleshooting
I have a Danica SV03 which is an isolated 3A variable source with V & A meters, but when you look inside it is simply a standard variable transformer with a separate isolation transformer so wouldn't
By
Adrian Nicol
·
#155354
·
|
Re: 2247A PSU Troubleshooting
There are variacs with an isolation transformer built in. But very rare. I own one made by Philips a dutch company.
By
Fred Schumacher
·
#155353
·
|
Re: 2247A PSU Troubleshooting
All variacs that I know of are auto-tranformers, so no isolation. David
By
David C. Partridge
·
#155352
·
|
Re: 2247A PSU Troubleshooting
Hi Brian, I found your post very informative but wanted to keep things as simple and clear as possible for the OP. I had recommended a Variac and he wasn't sure what to do. In his case he can either
By
Raymond Domp Frank
·
#155351
·
|