Re: one last 7633 question (for now)
One thing that kinda threw me was no trigger light. Well just now looked inside my 7b53a and the previous owner had plugged in the pin header backwards. So the trigger light is now working. Jeesh
By Brendan · #145129 ·
Re: Tek 2465a grid bias and brightness problems
If you have no control range to speak of on the bias pot, then I'd guess Q1980 has gone bad. Looking at the schematic, what the service manual states makes sense. Discounting Q1980s base current, its
By Sigur?ur ?sgeirsson · #145128 ·
Re: one last 7633 question (for now)
Details,details,details. Thanks for helping me not go crazy. I literally looked at that schematic for hours yesterday and today. Finally got to the "well i gotta ask" point. I told my wife I was too
By Brendan · #145127 ·
Re: one last 7633 question (for now)
Note also that some plugins, notably the programmable ones (definitely 7A16P and 7B90P, probably 7A29P as well) use the 5V 'lights' supply to run their internal logic. They don't work at all in
By cmjones01 · #145126 ·
Re: Tek 475 attenuators faulty when two sections in series
Tom, Thanks for the thumbs up and congrats that it's now doing it... One of the atten. blocks showing flaky seems now much more sensible... What wasn't making any sense before was that all
By Fabio Trevisan · #145125 ·
Amazing 453A, nicest old scope I have ever seen
I recently got a 453A off ebay because it was shown working with a bright CRT, and all the BNC connectors were still clean and shiny. It turned out to be amazing in real life, everything worked, (one
By walter shawlee <walter2@...> · #145124 ·
Re: Tek 475 attenuators faulty when two sections in series
Firstly, thank you for taking the time for a lengthy explanation. Secondly, I didn't take any offence in any way. I'm only too aware that I have a lot to learn, especially when I ask a question! I've
By Tom Gardner · #145123 ·
MG Chemical Responds --> Re: Problems with MG Chemicals Super Contact Clearer
They are being proactive. --------------------------------------------------------------------------------- Hello Stan, I have forwarded your concern to production team. They have tried several tubes
By brasscat · #145122 ·
Re: Problems with MG Chemicals Super Contact Clearer
Hello Eric, Yes, it is either the nozzle or the tube. Your other post asked if I tried both ends of the nozzle, you must have been as tired as I was. I think your meant both ends of the tube. I
By brasscat · #145121 ·
Re: Tek 2465a grid bias and brightness problems
On Q1980 collector with grid Bias at min I see 120 V and I have already replaced C1991 and R1994 even if they tested good out of circuit (R1994 was indeed a bit low: 1% out of tolerance on the low
By Francesco · #145120 ·
Re: one last 7633 question (for now)
While I intend to do this mod to my 7623A, it is worth bearing in mind a couple of things. If you check the app note/mod sheet (040-0686-01) you will see that some things should be down-rated
By Colin Herbert · #145119 ·
Re: Tek 2465a grid bias and brightness problems
Both the grid bias pot and U1890A shunt current away from Q1980. By your description it sounds like extra current is sneaking into Q1980s base - maybe C1991 has gone leaky? Page 3-37 has this descript
By Sigur?ur ?sgeirsson · #145118 ·
Re: one last 7633 question (for now)
HAHAH. Thank you. That is good stuff. You are exactly right. Im'm going to wait until I get a donor parts 7633 to do that mod. Make her look pretty. Time to button her up for now.
By Brendan · #145117 ·
Re: one last 7633 question (for now)
I presume you mean the illuminated push switches on the various plug-ins? If so then I'm afraid the answer is nowhere on the 3-slot chassis! I believe the reason was a shortage of power in the
By Adrian Nicol · #145116 ·
one last 7633 question (for now)
I have no lights but I know they work. Im assuming because p10 is empty. I have looked at the schematic for hours but I cant figure out where the input to p10 comes from.
By Brendan · #145115 ·
Re: Tek 2465a grid bias and brightness problems
Tested another U950, from a working scope of a friend, both U950 work on his scope, both doesm't work on mine. It seems like grid bias is not clamping the ac waveform at the correct amplitude. If I
By Francesco · #145114 ·
Re: 7623A + 7A26 + 7B53
Hi Fabio, I have one of these, but it originally came with two 7A18 dual-trace vertical amp plug-ins. I think this, together with the 7B53A, was pretty much the standard complement for the 7623A
By Colin Herbert · #145113 ·
Re: 16700A setting up Xwindows.
Right place is the hp_agilent_equipment on Yahoo! Dave
By David C. Partridge · #145112 ·
Re: 16700A setting up Xwindows.
This will be my last post on this thread. You should do the same, so ppl here can get back to Tek(tronix). Your post on the CC list says nothing about looking for a CD/ROM drive that is 512 byte block
By Pete Lancashire · #145111 ·
Re: A 16700 compatible CD-ROM Drive?
Pete, I'm sorry. I have joined them but I am waiting for Admin approval. Please don't yell at me. It wasn't necessary. I just misunderstood and made a honest mistake. Again I am sorry. Everyone else
By Richard R. Pope · #145110 ·