Chris,
I checked the manual and the 465B uses the same multiplier.?? So I could disassemble my parts unit and remove the multiplier.?? I have no idea whether it is good, and it will probably be a bit of work to remove it.
Let me know what you think.
Bob
--- On Sun, 2/10/13, chrismh_somenumber wrote:
From: chrismh_somenumber
Subject: [TekScopes] WTB or WTS Tektronix 465
To: TekScopes@...
Date: Sunday, February 10, 2013, 3:17 PM
??
I have a 465 that, except for a blown HV multiplier (part 152-0552-00), works fine. The problem started while I was using it, so I know it works fine.
There is no display on the CRT and I've traced the problem to the -2450 supply. Following the troubleshooting procedures, I disconnected the anode lead and the dummy resistor in the HV supply and the -2450 came up. The manual's suggestion at that point is to replace the HV multiplier.
I was planning on selling this scope anyway so I'm hoping that someone on this list:
1. Wants a 465 that only needs a HV multiplier
2. Has the part I need and wants to sell it (part 152-0552-00)
Thanks,
Chris