Hello Zenith5106,
thanks a lot for your help, the DC balance is as expected now...
best regards
Peter/the Netherlands
zenith5106 wrote:
toggle quoted message
Show quoted text
--- In TekScopes@..., Peter de Vroome <p.devroome@a...>
wrote:
Dear Tektronix experts,
I have a Tek TDS350 with a DC balance problem in channel A, when I
turn
the attenuator the trace is hopping app. 1cm. Is this easy to
adjust by
a potmeter, or is it not possible for customers to adjust?
Channel B is ok..
Thanks for your advice
Peter
---------
Do an SPC (= self cal). Press Utility, on bottom/left menu select
Cal, select Signal Path. To start, select OK Compensate Signal Path.
It will take a couple of minutes and should take care of your balance
problem unless there is a hardware problem. Make sure the scope is
warmed up.
/Zenith, Sweden