Thanks Mark & Morris.

I like the idea of replacing the 5642¡¯s with silicon, so I¡¯ll order the 2CL2FLs from Amazon. One less thing to go wrong. Strangely, Digikey doesn¡¯t seem to carry them.

It may take a week or so to source the 6.3V xfmr. Not sure I have one in my junk box, but I know someone who does. I was hoping to avoid wrestling that brute of a ¡®scope on its side to get at its power supply. Drat! Oh well. The xfmr may well be the culprit, since I can remember hearing some low ¡®corona¡¯ sounds from that area, although the fan makes it hard to tell where the sound is coming from. I believe you are referring to only terminals 5&6 on the transformer¡­correct?

My notes say that I replaced C858 already, so we¡¯ll assume that¡¯s ok. I will obtain a C802 replacement and try that out too.

I don¡¯t have a ¡®megger¡¯ or anything similar, so measuring leakage from the focus pot is not possible, other than with my DMM. Instead, I could get a 5 Meg pot and isolate it from the chassis, or a couple of 2.2 Meg resistors in series with centre tap between the two.

Will report back when we get that all configured.

Your expertise and experience with this type of problem is much appreciated!