Hey Keith,
I calibrated the CRT, timing, vertical and the triggers on my 2467 using a TG501 and a PG506. I didn't touch the transient response calibration, but I don't think the PG506 is fast enough for that. You need a much faster edge, and IIRC Tek specifies a tunnel diode pulser. I think if you have a decent signal generator or ARB, it should be able to produce all the signals the calibration calls for^W^Wneeds. The horizontal (timing) calibration is all done by eye, using the technician as a visual comparator. TG501 is very convenient, as it gives you fairly narrow, fast-rising pulses for timing marks and it's quick and easy to toggle to the requested settings. A square wave of the required frequency would work, though, even a sine at a pinch. There'd be an awful lot of faffing about with controls, though :). The vertical calibration is similar to the horizontal, e.g. the technician is used to adjust levels until traces line up. IIRC you need a max of a 5V peak to peak (ground to peak) signal for the vertical calibration, which is easily done with a digital signal generator or an ARB that has 12bit resolution or better. At the lower end it might be getting hairier, I don't remember offhand what the signal levels are there, but to get good resolution and low noise you might need to use attenuators, depending on how your siggen/ARB works. The trigger calibration is however automated, and is somewhat picky on the signal provided, but again nothing you wouldn't be able to provide with an ARB or a digital siggen. Siggi On Fri, Jan 17, 2025 at 7:00?AM Keith via groups.io <keith= [email protected]> wrote: Thanks Raymond, yes the fast edge rate is a problem, the 8112 is about 5nS |