Re: SC502 Trace Issues - Beginning and ending of trace connected.


 

If the intensity control doesn?t work the blanking problem is a consequence of it.

With a high degree of probability one or more diodes of the intensity DC restorer are leaky. Since testing it is quite difficult, I would replace all.

This problem happens in several Tek models and also in scopes of other makers (Who copied the circuits designed by Tek).

Regards,

Ignacio

El 04/03/2024 a las 16:48, Barry Breaux escribió:
Interesting that you mention intensity. The intensity adjustment does nothing. I am troubleshooting that now.

Barry


--