Re: TDS520 in Brittany: FAIL++Acquisition FAIL++ Attn/Acq interface


 

I repaired a TDS520 with exactly same error screen after repairing a (literally) burned power supply. In my case leaking electrolytic caps damaged a trace to U900 on acquisition board.
Ozan


Hi to all usual suspects here.
You will maybe remember you helped me resurect my 7704A from decades of non
working neglect. Thanks!
Here is for me a new challenge. I did a large leap in time from the 7704A and
just bought a TDS520.
At boot:
All ok except:
FAIL++Acquisition
FAIL++ Attn/Acq interface
As soon as I knew it will be coming I read here all I could about it so I know
that Artek has the documents.
Btw I want to praise Artek a bit here. I bought them the service manual for my
two HP8640B I just bought too (yes, two is gluttony, I know but they were
cheap and defective of course, so how to resist?)
Anyway Artek gives a really good service. The manuals are of upmost quality. I
got them instantly and the charge is very reasonable.
I posted the boot to "TDS520 in Brittany"
/g/TekScopes/album?id=291619
Also the 1993 rom: A11U1331 ver 1.5

There's a 7 segments inside at the back which only blinks alternately his two
right segments.
Apart for the fails the scope is in perfect shape and was only very dusty
inside. I disinserted all flat cables I could find and re-inserted them. I
also removed the dust with hope this will change the fail boot (not really!)
It did not.

All advices will be received gratefully.


Join [email protected] to automatically receive all group messages.