Is there any audible ticking when that happens? A good start would be to
monitor the various voltages from the PSU section and see if any of them
are dropping out.
toggle quoted message
Show quoted text
On Sat, Apr 30, 2022 at 11:26 AM Harrison <buma7@...> wrote:
Picked up a 2232 scope recently at a Club auction. Pulling it from the
case I found the J9011 connector broken and the eight wires to it
dangling. As was suggested in the Group I hard wired the eight wires
directly to the analog power distribution board. Now I seem to have a
trace on both channels and I can acquire a AC and a DC signal. Plus, probe
compensation can be attained. However, at times the trace spikes and
disappears. After which the power light and the SGL Sweep light blink at a
rate of about once per second and a single digit LED display located at the
front of the analog power distribution board flashes. It seems to have
done this for a little as 30 blinks (seconds) to perhaps 60 blinks
(seconds). After this occurs the trace returns and all seems normal until
the nonoccurence of the cycle. I don't see anything in the operators
manual or the service manual referring to this LED display. Is it a set up
issue, operator error or equipment malfunction? I have pretty much stayed
in the analog (non-store) mode but when I did briefly select the digital
mode I did have the proper displays and selections on the CRT. Can anyone
help me better understand what is going on when the unit acts as I
described. Thank you for your input. 73 de N1FAM