Hi Sean,
toggle quoted message
Show quoted text
The revised March 1988 7104 Operators Manual says the display of the 7D01 Hardware Logic Analyzer and the 7D02 Software Logic Analyzer can cause permanent reduction in CRT microchannel plate gain; consequently, a permanent reduction in writing rate. For more information refer to "Reduction of Display Gain with Display Output Change" in the 7104 Service Manual. Dennis Tillman W7pF -----Original Message-----
From: [email protected] [mailto:[email protected]] On Behalf Of [email protected] Sent: Friday, May 29, 2020 10:09 PM To: [email protected] Subject: Re: [TekScopes] Advice about buying a 7904 Reg, To add to Harvey's excellent response, yes indeed most (all?) 7k plugins will work in any 7k mainframe. HOWEVER, there are at least a couple plugins (I cannot recall which ones right now, perhaps someone else knows off the top of their head) that explicitly must never be used in a 7104 because the way they drive the crt can cause damage to it. You absolutely could simply use the plugins you have for your 7104 in a 7904(A). Sean On Fri, May 29, 2020 at 05:53 PM, Reginald Beardsley wrote: