Is your scope working properly, or is it having some sort
of problem?
There are frequently problems with the CA3046 transistor
arrays, which will cause glitching in the CH1 and 2 traces
when they fail in the little addition around CH5 of the
channel hybrid.
-Chuck Harris
Panos wrote:
toggle quoted message
Show quoted text
Hello to all.
Taking some measurements on my TEK 2465 using another oscilloscope, I see something strange.
I read for CHN5B signal -1,2 Volt versus -3,6 Volt. This only with BW LIMIT off.
Also for CHN 5A I read signal 600mV p-p versus -6 Volt as look at waveforms on service manual.
Am I doing something wrong, or I have problem somewhere in my TEK scope?
Waveforms 28 and 46 in pages 238 and 239 in service manual. These appears in diagram No 5 in page 244.