Jim,
Great News!? One never knows when one small component will completely disable an instrument.? Proving that every part is important, even critical to proper operation.?

Michael Lynch?479-226-0126 Home Phone479-477-1115 Cell Phonemlynch001@[email protected]@...

On Saturday, August 31, 2019, 10:30:21 AM CDT, James R. Bartlett <james.r.bart@...> wrote:

Hi Michael.
Latest update.
Scope is working again.
Changed C963, replaced R963,.
Had to readjust the? pot,for the + 50 voltsupply, as +15 was showing +18,
Trace seems reasonable and all controls appear to work.
Probably needs a good going over but the main thing is it is working.
Many thanks to yourself and others who came up with various suggestions.
Hope you have a nice weekend and enjoy Labor Day.
Again thanks for all
Regards
Jim
In the land of the Leprechauns ?