Raymond,
Thanks for your reply! I set up the scope and the FET probe as you said, and compensated it at 2 us. The adjustment was a bit fiddly so I followed the manual instructions for removing the tiny PCB inside the attenuator and cleaned + exercised the trimmer, which seems to have settled that down. Now it compensates much better! Thank you!!
Sean
toggle quoted message
Show quoted text
On Thu, Apr 18, 2019 at 05:57 PM, Raymond Domp Frank wrote:
Hi Sean,
There's nothing wrong with your P6201.Two recommendations:
1. Put the 2465B input at 1 MOhm DC or 50 Ohm DC, *not AC*. Use the P6201's
External Terminator setting if the latter and in both cases use the probe's
setting to select AC or DC coupling.
2. Compensate the P6201 at either 20us, 2 us or 100 ns / div, *not 1 ms* / div