I'd have had to wait for my ST replacement NVRAM without the assist from Tom Miller.
Wherever did you find a DS1213? Digikey and Mouser don't have them. I doubt that any of the ones on eBay have good batteries. But you can see the outline of the tiny lithium cells potted on the underside, so a bit of Dremel work would allow fitting a larger pair of cells in holders in the unused DIP footprints in the left front corner of the board.
The next issue is what data is stored in these? It's obviously not the RTC or anything else I can determine such as operating hours, serial number, etc.
The only downside is now I need an SD-24 and an SD-32 ;-)
--------------------------------------------
toggle quoted message
Show quoted text
On Sun, 3/24/19, Jack2015 <tolkien@...> wrote:
Subject: Re: [TekScopes] 11801 diagnostic help
To:
[email protected] Date: Sunday, March 24, 2019, 8:02 PM
Congrats!nice job.
After replacing the DS1213
with a new NVRAM in my CSA803 (I? replaced one DS1213 only
)
I rebooted it 3 times to get rid of the
self test failure.
1st
time:same err E5322
2nd time:Battery
error,but can exit diagnostic and enter operational
screen.
subsys
--------------block------------area------------routine
Timebase? -----TBC control------ Static RAM
-----Battery
3rd time:passed self test.