Thank you everyone for the feedback. Like I said - have no experience at all with this amazing
instrument and would have probably torn the whole thing apart again chasing an issue that doesn't
exist.
I ran through the adjustment procedure last night and all except the "Polarity Positioning" when
switching back from "Display Invert" to " Display Norm" (Adjustment Procedure, step 15 (f)) could
be
brought within spec.
The only remaining issue is the storage feature not working well - very weak overall and only works
around the center of the screen. Not worrying about that right now, not planning to use storage.
Thank you again, Wolfgang
See the note on page 10 of the operations manual regarding conditioning of the storage CRT if it has
been turned off for two weeks or more.
Craig