Re: 468 horizontal jitter


 

On Sat, Oct 13, 2018 at 09:51 AM, Jeff Urban wrote:


In case you haven't, it is important to check if it is a centering problem or
a triggering problem. Even in auto or free run it is still triggered, so if it
is the problem could be all the way back at the ramp generator.

If the trace itself is moving and not the sweep (I hope you got that...) then
I would more look for hash on a PS line somewhere. Of course that applies to
the ramp generator as well but different PS legs of course.
Thanks Jeff.
There is jitter on the A gate out with no input just a free running trace. I had to put my 465 in 10x to see it but it does match the jitter on the 468. So... This looks like its going to be a fun one to figure out.