With curves like this
/g/TekScopes/photo/59361/1?p=Name,,,20,1,0,0 I wouldn't even bother to calculate a rise time. Obviously the main problem is roll-off. I would search the manual for Adjustments which influence that nanoseconds part of the curve and try these first.
toggle quoted message
Show quoted text
On Sun, Jun 17, 2018 at 09:59 pm, tmillermdems wrote:
So BW = 0.35/risetime in seconds or 0.35/0.000000006 = ~~60 MHz.
The 465 should go around 100 MHz bandwidth. Around 3.5 ns.
Is the bandwidth limit turned on ( maybe a fault with the switching? ).
Do you have any other scope with a higher bandwidth available? Some way to
verify the setup.
----- Original Message -----
From: "lop pol via Groups.Io" <the_infinite_penguin@...>
To: <[email protected]>
Sent: Monday, June 18, 2018 12:22 AM
Subject: Re: [TekScopes] 465 bad rise time
Time div is set to .05us but I have 10x mag on so that should be 5ns.
On Sun, Jun 17, 2018 at 09:10 pm, tmillermdems wrote:
Ok, it looked like a probe at first. What is the H sweep rate set to?
Can't
see it in the pictures.
Looks like 1 1/4 divisions on the horizontal 10-90% points.
Regards
----- Original Message -----
From: "lop pol via Groups.Io" <the_infinite_penguin@...>
To: <[email protected]>
Sent: Monday, June 18, 2018 12:02 AM
Subject: Re: [TekScopes] 465 bad rise time
Im using a 50ohm cable with a 50ohm through terminator.
/g/TekScopes/photo/59361/2?p=Name,,,20,1,0,0
On Sun, Jun 17, 2018 at 08:50 pm, tmillermdems wrote:
You are looking at the waveform with a scope probe that presents a 10
meg
ohm load to the pulse generator. What you need to do is use a 50 ohm
cable
to go from the generator to the scope input and include a 50 ohm through
termination right at the scope. A real simple solution would be to use a
BNC
T with the center loaded with a 49.9 ohm resistor. Or just look for a 50
ohm
through termination on ebay.
Regards
----- Original Message -----
From: "lop pol via Groups.Io" <the_infinite_penguin@...>
To: <[email protected]>
Sent: Sunday, June 17, 2018 9:47 PM
Subject: [TekScopes] 465 bad rise time
Hey guys. I have done a lot of work to this 465. I had initially planned
on
selling it to help me buy some plugins for my 7633. I decided I don't
want
to sell it and am keeping it. Today I checked its rise time with a pg506
fast rise. Both channel 1 and 2 closely resemble each other and they are
not
close to right. Before I start trying to calibrate this thing, does it
look
like there is some type of component issue left to fix? I am thinking
since
both channels are so alike there is something wrong besides just out of
calibration. Thanks guys.
here are some pictures
/g/TekScopes/photo/59361/0?p=Name,,,20,1,0,0
/g/TekScopes/photo/59361/1?p=Name,,,20,1,0,0