¿ªÔÆÌåÓý

ctrl + shift + ? for shortcuts
© 2025 Groups.io

Re: OT (sorry)


 

I have one.

1) get yourself a good copy of the service manual from Artekmanuals.com



2) Check all power supply rails for level and ripple.

3) There is a memory battery - check it and change it.

4) The error e-52 is Shaper IC 401 not able to supply negative output when...
see service blocks 5,6,7.

This may be a power supply issue if you are lucky.


HTH,
Tom

-----Original Message-----
From: lop pol via Groups.Io <the_infinite_penguin@...>
To: TekScopes <[email protected]>
Sent: Mon, Mar 12, 2018 10:15 pm
Subject: [TekScopes] OT (sorry)

Im looking on evil bay at a hp8116A function generator. The one I'm looking at is displaying an e52. I have looked all over for error codes and am not able to find anything. The manual from agilent tells me to look at section 10 but the manual ends at 9-27. I was hoping someone here has that function generator and the manual?

Join [email protected] to automatically receive all group messages.