¿ªÔÆÌåÓýIf that is the ripple on the regulated outputs, I would replace the caps.
Especially the one on the +32 regulated supply as it is also the reference for
the +5 supply.
?
It looks like the calibration waveform is being modulated in both amplitude
and period. That most likely is why the xy display is elongated.
?
Take a look at the unregulated 32 volts (around 50 volts) and see how dirty
it is. Confirm that it is 120 Hz (or 100 Hz if you are in one of those weird
countries that doesn't spin their generators fast enough) :).
?
What readings did you see and what are you using to measure ESR?
?
?
Regards,
Tom
?
|