Re: Tricks and thoughts about the 3457A battery
Hello Joel,
I was thinking it is the version with 4 holes. I found the same design
in 2 of my Meinberg GPS167 units. One had a battery holder with 2 pins
and one had a battery with 3 pins.
I looked
By
Razvan Popescu
·
#151138
·
|
Re: 3585B Spectrum Analyzer
This is a link for service manual version A, This might be useful to you.
By
Yves Tardif
·
#151137
·
|
Feet for ESG style instruments. 3D model.
I have modelled a simple foot that I use for the ESG series generators.
It is just tall enough for the instrument chin to hover over the desk surface.
Leo
Here goes nothing:
By
Leo Bodnar
·
#151136
·
|
Re: 3585B Spectrum Analyzer
Hello everyone.
Please help.
I have a 3585B that wont power up.
I need a copy of the service manual, can anyone help??
By
crowbarkev@...
·
#151135
·
|
Re: PN 1855-0064
Thank you both for chiming in.
The unit is a 7562A, I apologize for the typo.
I pulled a couple from another board and measured one side of each package, presumption is the other is similar.
I was
By
John Warren
·
#151134
·
|
Re: Tricks and thoughts about the 3457A battery
My standard approach for replacement of batteries that support NVRAM containing instrument calibration , etc. is to connect either another battery or external power supply **with floating outputs**
By
Greg Muir
·
#151133
·
|
Re: HP6209B component location layout
Just FYI, you can use the (much shorter) www.hpwiki.org hostname in that URL.
-Dave
By
Dave McGuire
·
#151132
·
|
Re: odd HP83651A error
Many thanks for your feedback! What was the problem with yours? A broken data line? Why exactly 100MHz shift?
Or is some YIG lookup table messed up so it locks to the wrong harmonic? How did you get
By
tom_iphi
·
#151131
·
|
Re: odd HP83651A error
Hi Tom,
I got same problem few years ago on my 83630B. The faulty module is the low band module on the RF deck at the far end opposite to the output RF connector.
By
F1EKU
·
#151130
·
|
Re: PCB CAD used and/or sold by HP ?
Here is a link to EGS and some history.
https://www.hewlettpackardhistory.com/item/a-new-dimension-to-design/
By
alan victor
·
#151129
·
|
odd HP83651A error
Hi folks,
my 8510C system recently developped an unlock error, so I have checked the source HP83651A first.
It turns out, that the source output below 2,4GHz is exactly (to the Hz) 100MHz too low,
By
tom_iphi
·
#151128
·
|
Re: PCB CAD used and/or sold by HP ?
Believe it was HP EGS...Engineering Graphics System , ran on series 200 and 300 and recall
a package running on the HP 9816 desktop.
By
alan victor
·
#151127
·
|
Re: Tricks and thoughts about the 3457A battery
David,
everytime I walk past the delivery section in my company I wonder what "sending from and to" of these quite sensitive instruments will do to the calibration. There are always some instruments
By
Martin
·
#151126
·
|
Re: Tricks and thoughts about the 3457A battery
[email protected]> wrote:
Another option is to ignore trying to save the calibration constants,
replace the battery and send it to Keysight and let them calibrate it. Then
it will be as good as you
By
Dr. David Kirkby, Kirkby Microwave Ltd
·
#151125
·
|
Re: Repair HP3455A Multimeter
Hi have a pdf from the flowcharts, group B schematics and the groupB service excerpt.
By
Ralf Rosche
·
#151124
·
|
Re: Tricks and thoughts about the 3457A battery
"Joel Setton via groups.io" <setton@...> writes:
Using an FRAM might be a possible solution. A quick search showed that
there's a FM1608B vom Cypress - 8Kx8 and available in DI
By
Sven Schnelle
·
#151123
·
|
Re: Looking for documentation on 4263A LCR meter
Hi Jared,
I had a check - no info on this instrument
Peter
By
Peter Brown
·
#151122
·
|
Re: Tricks and thoughts about the 3457A battery
Hello Razvan,
You are right, the original battery has 1 pin on the (-) end, and two pins on the (+) end. However, I found that the PC board has three plated holes on the (+) side including one in the
By
Joel Setton
·
#151121
·
|
Re: Tricks and thoughts about the 3457A battery
Hello Martin,
The EPROM idea is interesting, I have explored it but I don't think it could work. The main issue is that the 8Kbyte RAM chip which holds the calibration data (TC5564APL-15) is also
By
Joel Setton
·
#151120
·
|
Re: Tricks and thoughts about the 3457A battery
Congrats!!
Ing. Patricio A. Greco
Taller Aeronáutico de Reparación 1B-349
Organización de Mantenimiento Aeronáutico de la Defensa OMAD-001
Gral. Martín Rodríguez 2159
San Miguel
By
Patricio A. Greco
·
#151119
·
|