¿ªÔÆÌåÓý

16-pin Connector Part Number for 11713-60044 cable
Hi Guys, Does anyone have infor on the 16-pin Connector Part Number for 11713-60044 cable? Mates with many RF switches and Step Attenuators. Thanks & BR' Ilan 4X6MU
By 4X6MU · #152067 ·
Re: HP 54503A Memory address error
Very similar on a 53310A MDO. Error came and went, only to disappear totally (knock on wood). In my case, I could cause it by switching on the Iwatsu DS6612-Oscilloscope which stood right above the
By Tam Hanna · #152066 ·
Re: Agilent E8357A Network Analyzer Phase Lock Lost Error
Looking over the results from the Serial Bus Test . Most of the Test Point measured voltages on the A-8 Fractional-N Board are wrong except the +5Vdc Supply Voltage, +15 Current Draw and PreTune
By Johnny10 · #152065 ·
Re: HP8566B band amplitude response discrepancy - how much indicates a component failure?
When working with the upper bands which go through the YTX be sure that the <PRESEL PEAK> values aren't way off into the weeds, if they are the measured amplitude will be a lot lower than reality.
By rmb_guru · #152064 ·
Re: Agilent/Keysight AWG 33250A /33220A CAL
Bonjour cher monsieur Yves, mille mercis pour votre r¨¦ponse rapide. Quelle surprise de trouver un Fran?ais avec ce vieil et bel instrument. J'habite ¨¤ Paris, o¨´ habitez-vous ? Nous pouvons
By Froggie the Gremlin · #152063 ·
Re: Agilent/Keysight AWG 33250A /33220A CAL
Bonjour Jon, Les constantes de calibration sont sauvegard¨¦es dans un EEPROM. Je poss¨¨de un 33120A et un 33220A. J'ai d¨¦j¨¤ calibr¨¦ moi-m¨ºme ces deux instruments, ¨¦videmment ?a prend un certain
By Yves Tardif · #152062 ·
Re: HP8566B band amplitude response discrepancy - how much indicates a component failure?
I have only done the adjustment procedure for A20, with the "preferred method" where you enter a bunch of shift-command stuff that gets saved, then you go between recall 1 and recall 2 registers that
By ed breya · #152061 ·
Re: HP8566B band amplitude response discrepancy - how much indicates a component failure?
You¡¯ve gone through the YTO loop calibration, right? If the pretuning controls aren¡¯t right for your replacement YTO, it might end up off-frequency far enough to make certain bands look worse than
By John Miles · #152060 ·
HP8566B band amplitude response discrepancy - how much indicates a component failure?
I'm still looking at this 8566B I just repaired, and pondering what it takes to do more adjustment. Right now it looks like the 2-5.8 GHz band shows over 12.5 dB lower than in the low 0-2.5 GHz band.
By ed breya · #152059 ·
Agilent/Keysight ENA 5071C Receiver Gilbert Cell Mixer 1GC1-4405
I have been fixing some 5071C ENAs.? Many of them have bad receivers.? Upon doing some tests & research, I noted that the Gilbert Cell Mixers were bad.? This is generally on port 1 on roughly 5 of
By Scott Vance · #152058 ·
Re: Agilent E8357A Network Analyzer Phase Lock Lost Error
Useful to be sure the clocking of the frac N counter (5 MHz) reference is ok. Does the manual provide logic swing levels? Scoping the clock at W4-1 should be sufficient. Then, does the manual provide
By alan victor · #152057 ·
HP 8702D OPT 011 conversion to 6GHz
Hello, I bought an 8702D which I didn't receive yet and I am doing some research in the meantime for the Test Set compatibility and also OPT 006 for a 6GHz upgrade. From my understanding if I buy the
By Razvan Popescu · #152056 ·
Re: Agilent E8357A Network Analyzer Phase Lock Lost Error
Great. That at least rules out the divide by 2 counter. Need to take a look again at your notes.
By alan victor · #152055 ·
Re: Agilent E8357A Network Analyzer Phase Lock Lost Error
"The Specification of frequency accuracy is +- 1ppm. However, this test typically adjusts to +-0.01ppm' .
By Johnny10 · #152054 ·
Re: Agilent E8357A Network Analyzer Phase Lock Lost Error
Ratio is 2.0000 But as you said 10MHz is way off. Manual says should be capable of 0.01 hertz ,
By Johnny10 · #152053 ·
HP 54503A Memory address error
Hi All, Has anyone seen and fixed this error before I pull the case? Thanks Chris
By Onewatt · #152052 ·
Re: HP8510C "system bus address error", but then it works
Also check RF Switch address. It must be 31 if you havent switch. Wrong address cause "System bus error" message. [System][F2=HP-IB Addresses][F6=RF Switch]=31
By §¥§Þ§Ú§ä§â§Ú§Û §À§â§î§Ö§Ó · #152051 ·
Agilent/Keysight AWG 33250A /33220A CAL
hello everyone Bonjour a tous... This fine 80 MhZ AWG (circa 2016) needs a full calibration similar to the 33220A 20 MhZ. Do either of these generators, use a back up battery or NV ram that can lose
By Froggie the Gremlin · #152050 ·
Re: Agilent E8357A Network Analyzer Phase Lock Lost Error
And your reported 10 . 000447 MHz? results in error still.... Yes place your counter into ratio mode to properly address counter faults...
By alan victor · #152049 ·
Re: Agilent E8357A Network Analyzer Phase Lock Lost Error
I suggest you place your counter into the RATIO MODE.? What I am addressing is rule out a faulty divide by 2 counter.? ?Since this a significantly robust phase lock synthesizer system, makes sense
By alan victor · #152048 ·