Date
Re: HP AN/USM 281A Oscilliscope needs repair (San Antonio TX)
Great. Thank you! --------------------------------------------
By Andrew Maich <andremajic@...> · #86162 ·
Re: HP AN/USM 281A Oscilliscope needs repair (San Antonio TX)
Ok, I re-read your original post. Thank you for your patience. I used the calibrator sockets on the left side of the screen with 1 piece of wire in the negative and 1 piece of wire in the 10v socket
By Andrew Maich <andremajic@...> · #86161 ·
Re: Fast Rise/Fall Time Pulser
--------------------------------------------
By Reginald Beardsley · #86160 ·
Re: some 89410A/89441A notes
Ok. If I can find any of my 32MB ones, I will try it and let you (and the list) know if it will recognize additional memory. -Dave [hp_agilent_equipment] wrote:
By Dave McGuire · #86159 ·
Re: some 89410A/89441A notes
Unfortunately, I don't have a simm 72 pins more than 16Meg.
By Yves Tardif · #86158 ·
Re: Fast Rise/Fall Time Pulser
SMD parts are too small to have a human readable full part number, so manufacturers either leave them blank (thanks alot!) or put a code on them that uniquely ID's the part within the manufacturer.
By Chuck Harris · #86157 ·
more 89441A: remote sessions
For those of us on UNIX systems (including Linux), a quick and easy way to remotely activate an X11 session from a workstation is: echo "SYST:COMM:LAN:XWIN:STAT ON" | nc vsa 23 > /dev/null On
By Dave McGuire · #86156 ·
Re: help hp5300a / 5303a manual service
As you probably know already, the 5300A manual is readily available online but the 5303A does not appear to be. However, in November 2015 someone on EEVblog said that Dave of Artek Manuals had made a
By Mike Czuhajewski · #86155 ·
Re: Fast Rise/Fall Time Pulser
The Leo Bodnar pulser output level is adjustable but at outputs higher than some 1 v P-P the rise/fall times increase. I reported the ~0.95 v P-P as what was reported on the datasheet for *my*
By Larry McDavid · #86154 ·
Re: HP AN/USM 281A Oscilliscope needs repair (San Antonio TX)
Hi Andrew You seam to be going in circles as you say I don't have a signal generator I posted text below on Sunday 8 oct Basically you do have a signal generator on the scope it is called
By Paul Bicknell · #86153 ·
Re: some 89410A/89441A notes
[hp_agilent_equipment] wrote: This is great news! I think it's safe to conclude that it works; it would likely have failed very early during the memory tests if there were a basic incompatibility
By Dave McGuire · #86152 ·
Re: some 89410A/89441A notes
Send me via private email your original config file with AY9 enabled and I'll try to figure it out. A new checksum needs to be calculated to include this option, in the master config file, plus a few
By Ovidiu Popa [Business] · #86151 ·
Re: some 89410A/89441A notes
Hi Dave, Wow! Good news! I was so curious that I verified wich type of memory is used on USG board : RAM chip was MT4C4001JDJ-6 (FPM memory type). Here, I had an old 72 pin memory SIMM of 16Meg
By Yves Tardif · #86150 ·
Re: some 89410A/89441A notes
I just went and looked at that. I found that same procedure on the eevblog forum, was that your post? I followed the procedure on my unit and it worked. There was one minor issue, though.
By Dave McGuire · #86149 ·
Re: HP AN/USM 281A Oscilliscope needs repair (San Antonio TX)
Woops - forget my reference to the 8558B in my prior response - too many threads. :) The remainder of the post is correct (i.e. look a the non-military manual for the unit). Ken
By Ken Nener · #86148 ·
Re: HP AN/USM 281A Oscilliscope needs repair (San Antonio TX)
Again, this could be related to open resistors in the horizontal deflection amplifier output stage. Specifically I would check R238, R239, and R275. These are the component references from the 8558B
By Ken Nener · #86147 ·
Re: some 89410A/89441A notes
FWIW, could be old news for many, but I uploaded at ko4bb the required files to enable all software options on a 89410A. The archive is still in the upload area.
By Ovidiu Popa [Business] · #86146 ·
Re: Schematics for HP 8558B SA
Thanks for the quick replies. Today I did find a complete military version of the HP 8558B manual with schematics. I'll start off with this and begin work on the SA, first checking and securing the
By Paul · #86145 ·
Re: Frequency calibration?
Right there with ya.. -Dave [hp_agilent_equipment] wrote: -- Dave McGuire, AK4HZ New Kensington, PA
By Dave McGuire · #86144 ·
some 89410A/89441A notes
Here are some random notes on the expansion options for the 89410A/89441A, for the sake of the archives. Second Channel -------------- It's possible to add the second input channel option
By Dave McGuire · #86143 ·