Re: Manual for HP 466A amplifier
I don't mind getting it scanned at the museum, but does it still make sense to do that after the revelation that Artek has it available? I hadn't checked before; I had assumed the OP had already
By Dave McGuire · #150604 ·
Re: Manual for HP 466A amplifier
I've got one of those amps and would like the schematic also. -- John Griessen
By John Griessen · #150603 ·
Re: Service manual scan post processing
While we seem to all agree that JPEG compression is not ideal, there ARE lossless options for TIFF and thus PDF formats.? While yes, storage and networking and processing and scanning time is all
By Bert · #150602 ·
Re: HP 3403C True RMS Voltmeter display fault
Yes, that's why I wanted to see a picture. I've never seen the ball-of-molten-tantalum failure on anything other than a dipped unit. Seconded. -Dave -- Dave McGuire, AK4HZ New
By Dave McGuire · #150601 ·
Re: HP 3403C True RMS Voltmeter display fault
The manual on the web gives type 150D for C1 & C2 on A7, these are hermetic sealed solid tants, of course they might have been replaced with a bead or the disguised bead in an axial or radial block
By factory · #150600 ·
Re: E5052A Power On Test (3.3V Bus Supply)
I don't know if this helps, but I dug out the most recent Keysight calibration certificate for this E5052A and there is a noise floor test done at 70 MHz with correlation set to 1. I'm not sure what
By jmr · #150599 ·
Re: HP Agilent 8753ES VNA repair [Help]
For reference list of the component I have on the A12 board: - 10H116 : Triple line
By MattSNE · #150597 ·
Re: Agilent MSO6014A Not Booting
This is a long shot (and I know nothings about this particular piece of equiptment), but here goes: I have an Agilent N5030A which was not booting.? I changed the battery on the mother board (and
By Scott Vance · #150596 ·
Re: E5052A Power On Test (3.3V Bus Supply)
-120dBc/Hz at 1 Hz is way better than anything I've experienced. I had a look at a plot of my homebrew oscillator and it managed -116dBc/Hz at 10 Hz.? Pretty lame by comparison... I guess you are
By jmr · #150595 ·
Re: HP Agilent 8753ES VNA repair [Help]
On TP20 I have a RF signal is at 39.280 MHz, -1.3 dBm ?for frequency <16 MHz instead of 39,996 MHz ?--> Delta is 716 KHz On TP14 for a 10MHz setting on the VNA I have a 2sd LO frequency at 10.720
By MattSNE · #150594 ·
Re: E5052A Power On Test (3.3V Bus Supply)
Hi Leo This E5052A is quite old now and it does have a few internal spurious terms. Also the noise floor performance isn't that great even down at 10 MHz. I don't have any oscillators here that can
By jmr · #150593 ·
Re: HP Agilent 8753ES VNA repair [Help]
Hello Yves, Thank you for your last post. My VCO frequencies in TP20 *are not correct*. I have : - a RF signal is at 39.100 MHz, -1.3 dBm instead of 39.840 on your board for frequency >16 MHz - a RF
By MattSNE · #150592 ·
Re: HP Agilent 8753ES VNA repair [Help]
There are contradictions in the test results: If you say that the VCO frequencies are correct (39.996MHz for LOW and 39.84Mz for HIGH), then this necessarily means that the VCXO is at precisely 40
By Yves Tardif · #150590 ·
Re: HP Agilent 8753ES VNA repair [Help]
De : [email protected] <[email protected]> De la part de MattSNE via groups.io Envoyé : 2 mars 2025 03:07 ? :
By Yves Tardif · #150589 ·
8510C/8530A Surplus units (UK)
Hi, I have accumulated quite a bit of 8530A/8510C stuff over the years. the 8530A stuff? I was going to use to experiment with an antenna range. Realistically it's never going to happen, and all the
By vee-dub565 · #150588 ·
Re: E5052A Power On Test (3.3V Bus Supply)
Hi Jeremy, This is a great win. What is your overall impression of the E5052A? I have R& FSPN8 and 53100A but I would like something I can try to push to lower phase noise levels. I need below
By Leo Bodnar · #150587 ·
Re: HP Agilent 8753ES VNA repair [Help]
Here is a more precise TP16 measurement. VCXO is Ok at 40.000 MHz
By MattSNE · #150586 ·
Re: HP Agilent 8753ES VNA repair [Help]
Is the test point on the picture correct ?
By MattSNE · #150585 ·
Re: HP Agilent 8753ES VNA repair [Help]
Hi Yves, Thank you for the details. I have some more questions. " For high band VCO must be 39.84Mz and VCO at 39.996MHz for lowband <16MHz." --> ?TP20 VCO. This part seems to work in my
By MattSNE · #150584 ·
Agilent MSO6014A Not Booting