Yes. If the instrument you are using does SCPI, then it's not to hard to write some code in Python to do some basic measurements. AFAIK, if you have the working GPIB hardware. Keysight 82357B USB/GPIB
By Roy Thistle · #147093 ·
Re: Artek Manuals and HP 3325B
As I have a couple of EIP 588s in the queue at my bench, I've been looking for the SM out there. Did a quick search, and bam!, one of the free libraries to the rescue!... And as I look closer, in the
By Radu Bogdan Dicher · #147092 ·
Re: Agilent E4406 fails to turn on - CULPRITS IDENTIFIED
Hi all, Thanks to Milan's pointer I de-soldered the two electrolytic capacitors on the little PSU daughter board and they were faulty. The 47uF@C5 was only 22uF and the 33uF@C6 thinks its a 100ohm
By Tim Tuck · #147091 ·
Re: Thermal Couple attached to heat sink
John, That pad is thermally conductive and is called a TIM pad. TIM? meansThermal Interface Material. Both Mouser & Digikey among others sell this stuff. Bergquist is one manufacturer but there are
By 2notlit · #147090 ·
Re: 8566B Attenuator Service
Mark - will do. Chuck - these contacts were not gold plated.
By Jinxie · #147089 ·
Re: 8566B Attenuator Service
Hi Jinxie - Please let us know if the calibration works now. Mark ________________________________ Sent: Monday, September 16, 2024 1:02 PM To: [email protected]
By Mark Bielman · #147088 ·
Re: 8566B Attenuator Service
If you were working for the telephone company, applying abrasive of any sort to relay contacts would have gotten you fired on the spot. A single technician, helpfully using such a technique could put
By Chuck Harris · #147087 ·
Re: 8566B Attenuator Service
Thanks, Vladan. Whilst I'd like to do a thorough job on this attenuator, I really need to clear my bench. This whole thing has gone on far too long and I have a backlog of 'stuff' to deal with.
By Jinxie · #147086 ·
Pb with HP16535A multi probes expansion card for HP16500B logic analyser.
Hi, Yesterday, while using my logic analyser (HP16500B), a component suddenly burned on my HP16535A multi probes expansion card. The PCB is a little damage, but this is fixable. I tested the board and
By @eikasus · #147085 ·
Re: 8566B Attenuator Service
It's probably residue from a silicone lubricant which was used on attenuators built before mid-1985. The stuff was causing problems so it was eventually abandoned. Unfortunately, a lot of attenuators
By pianovt · #147084 ·
Re: GPIB scripting software?
I have submitted my e-mail address to Harke Smits. In a project some years ago, I used python routines written by others but never had the time to dive into this language. Looking forward to what I
By Ulf Kylenfall · #147083 ·
Re: Artek Manuals and HP 3325B
Tim, The Artek office mgr was out due to medical and is back now, filling orders once or twice a week. I'll let them know to expedite if possible. Regards, Ron -
By Ron · #147082 ·
Twelve(+/-) bins of original HP/Agilent manuals free for pickup in San Diego
I have approximately twelve plastic storage bins (22" x 15" x 12") of test and measurement related manuals that are surplus to my needs. The vast majority of them are HP/Agilent manuals (e.g. scopes,
By Stan · #147081 ·
Re: Artek Manuals and HP 3325B
The spam filter explanation is extremely likely - I fell prey to that recently. It's been the only issue in years of using their service, other than the unfortunate times when things were in flux due
By Radu Bogdan Dicher · #147080 ·
Re: Artek Manuals and HP 3325B
No problems at all as late as about 10 days ago.? Pretty sure the email address used will be the same Manuals@... ( manuals@... ) used to contact them so be sure it is not being
By Harold Foster · #147079 ·
Re: 8566B Attenuator Service
YAY! ________________________________ Sent: Monday, September 16, 2024 9:04 AM To: [email protected] <[email protected]> Subject: Re:
By Mark Bielman · #147078 ·
Re: Artek Manuals and HP 3325B
That is pretty uncharacteristic for her. At least in my experience.? She has done a great job in keeping up with my manual orders.? I ordered several earlier this year.? I made sure to thank her
By Frank Mashockie · #147077 ·
Re: Artek Manuals and HP 3325B
Based on my experience what most likely what happened, is she sent the email with the link, and a spam filter somewhere ate it. I contacted her with a different email address a few days after my order
By Jim n8qoh · #147076 ·
Re: 8566B Attenuator Service
Progress! There was nothing wrong with the logic and the solenoid was *not* energized long-term after all. That 20dB pad was operated by a solenoid which had some kind of contaminant on its lower Y
By Jinxie · #147075 ·
Re: Artek Manuals and HP 3325B
Dave died, and his widow is trying to keep things running by herself. [email protected]> wrote:
By Michael A. Terrell · #147074 ·