Re: sick hp 5421A Digital Processor
It's true that the interests on the GammaSpectroscopy list are mostly about gamma spectroscopy measurements and materials, and less about the instrumentation. The common focus on using PC sound cards
By Dave McGuire · #103577 ·
Re: sick hp 5421A Digital Processor
Yup, I've been nonfunctional due to painkillers following a small oral surgical procedure. I should have that info for you by tomorrow. -Dave -- Dave McGuire, AK4HZ New Kensington, PA
By Dave McGuire · #103576 ·
Re: sick hp 5421A Digital Processor
Hello Dave, just in case my e-mail reply to your kind offer ended in your spam folder: you got mail ;-) Kind regards Chris
By ChrisBeee · #103575 ·
HP 5347A with 8482A Power Sensor?
Hello group. Wondering if anyone has tried using the 8482A power sensor with the 5347A. It is the only sensor I have. Power ranges are the same but not the frequency range so clearly cannot measure
By Mark Bielman · #103574 ·
WTB HP 8660C Module Extension Cables 11672
I wish to buy the Service Kit p/n 11672A, or the two extender cables, p/n 11672-60001 and 11672-60002 for the 8660x signal generator series. Please contact me off list if you have something for sale.
By Bob Koller <testtech@...> · #103573 ·
Re: HP3336b
To quote Wikipedia, as just one example.... "An RCA connector, sometimes called a phono connector or (in other languages) Cinch connector, is a type of electrical connector commonly used to carry
By Nigel · #103572 ·
Re: A dirty way to get more frequency resolution out of a signal generator
This will probably work but you would need a counter to verify the output signal is what you are expecting. There is no guaranty that a low resolution generator will be able to hold steady on digits
By victor.silva · #103571 ·
Re: [WANTED] HP 85700A 32kB RAM card
Hi Steve, On the screen issue, try the display centering adjustments on the second page of the CAL menu. If that doesn't help. over-scan and centering problems can also be caused by the wrong sync.?
By Mark Litwack · #103570 ·
Re: HP3325 - no output, but sync works
Simple check is to take the cover off, locate the attenuator board and find the input cable. Put a scope on that and it should have a signal ¨C this would prove the rest of the generator is
By nigel adams · #103569 ·
Re: HP3325 - no output, but sync works
Odds are the attenuators is shot. If it has the flat black aeromat relays you won¡¯t hear them even if the work. They are latching type so checking the drive signal will only give you a plus. You
By Stephen Hanselman · #103568 ·
Re: LO port on HP 8591A
Thanks, your comment made me laugh. Glad to hear its not just about saving $20 anymore and we now actually have another brand/series of SA that it can be used for.
By Marcus Bannerman · #103567 ·
Re: Hi all..have a question on a source for micro-grabber clips (smd size)
Something like HP 5090-4356? Plenty around on ebay at a $ or so each. wrote:
By Orin Eman · #103566 ·
Re: Hi all..have a question on a source for micro-grabber clips (smd size)
I don't know if something like this is what you're talking about, but I thought I'd offer it on the chance that it was what you're looking for.
By G Hopper, KB7WSD · #103565 ·
Re: Finding feet for older HP gear
Hi Sean you are asking for components but you forgot to give your location Paul located England _____ From: [email protected] [mailto:[email protected]] On
By Paul Bicknell · #103564 ·
Re: Hi all..have a question on a source for micro-grabber clips (smd size)
Are you able to put up an image of what you are looking for, might help folk understand more clearly what you are after. Regards Nigel ________________________________ Sent: 02 March 2020 04:07:19 To:
By nigel adams · #103563 ·
Re: HP3325 - no output, but sync works
Joe and Szabolcs: thanks for the ideas! I have not yet opened the box (after years...). Since this unit came from my employer, who has a metrology, and calibration group, I¡®m betting on the bad
By Randy Newman · #103562 ·
Hi all..have a question on a source for micro-grabber clips (smd size)
Hi all: I am in the process of troubleshooting a Tek AM503 current probe amplifier and am looking for the small SMD sized type of grabber clips. I have some of the larger Pomona types with the
By Randy Newman · #103561 ·
Finding feet for older HP gear
Hi all, I have a number of examples of the older style case HP gear (8640B, 8405A, and now a 3440A) that all lost their feet at some point before I got them. I know these feet are rather scarce, but I
By Sean Turner · #103560 ·
Re: [WANTED] HP 85700A 32kB RAM card
The logical step in any situation is to replace a failed component.? But when one relies 100% upon a mission critical part, without knowing that it failed, they can be subtly surprised if it had
By Greg Muir · #103559 ·
Re: Fuses for 5350B input 2 [lower frequency input]
Thankyou Askild, and Greg. I have an inquiry in to one of the suppliers on eBay. Shipping showed as about $350 - I think this is a little too high, and might be an error ;). I have a 5350B on its way
By donald collie · #103558 ·